使用常规儿童胸部X线摄影协议对辐射剂量与低对比度细节可探测性的比较分析
该研究使用CDRAD 2.0模体与PMMA板模拟四个儿童年龄组胸部,在八家医院17台X线机上按各自现有儿童胸部X线协议采集,比较低对比度细节可探测性(IQFinv)与入射空气比释动能(IAK)。结果显示医院间及医院内IQFinv与IAK范围均存在明显差异,作者认为四个年龄组尤其低IQFinv、高IAK的机房需进一步辐射剂量优化。
要点
- 儿童胸部X线协议在医院间与医院内IQFinv和IAK差异明显
- 新生儿组医院间IQFinv范围1.40-4.44,医院内1.52-2.18
- 新生儿组医院间IAK范围8.56-52.62 μGy,医院内21.79-52.62 μGy
核心结论
- 医院间与医院内低对比度细节可探测性和入射空气比释动能存在相当大差异,四个儿童年龄组尤其低可探测性、高剂量机房需进一步辐射剂量优化。
研究设计与样本
- 使用CDRAD 2.0模体与医用级聚甲基丙烯酸甲酯板模拟四个儿童年龄组胸部,在八家医院17台X线机上按各院现有胸部X线协议采集。
诊断标准与关键数值
- 以图像质量因子倒数(IQFinv)物理测量低对比度细节可探测性,以固态剂量计测量入射空气比释动能(IAK)。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: To compare low contrast detail (LCD) detectability and radiation dose for routine paediatric chest X-ray (CXR) imaging protocols among various hospitals.
METHODS: CDRAD 2.0 phantom and medical grade polymethyl methacrylate (PMMA) slabs were used to simulate the chest region of four different paediatric age groups. Radiographic acquisitions were undertaken on 17 X-ray machines located in eight hospitals using their existing CXR protocols. LCD detectability represented by image quality figure inverse (IQFinv) was measured physically using the CDRAD analyser software. Incident air kerma (IAK) measurements were obtained using a solid-state dosimeter.
RESULTS: The range of IQFinv, between and within the hospitals, was 1.40-4.44 and 1.52-2.18, respectively for neonates; 0.96-4.73 and 2.33-4.73 for a 1-year old; 0.87-1.81 and 0.98-1.46 for a 5-year old and 0.90-2.39 and 1.27-2.39 for a 10-year old. The range of IAK, between and within the hospitals, was 8.56-52.62 μGy and 21.79-52.62 μGy, respectively for neonates; 5.44-82.82 μGy and 36.78-82.82 μGy for a 1-year old; 10.97-59.22 μGy and 11.75-52.94 μGy for a 5-year old and 13.97-100.77 μGy and 35.72-100.77 μGy for a 10-year old.
CONCLUSIONS: Results show considerable variation, between and within hospitals, in the LCD detectability and IAK. Further radiation dose optimisation for the four paediatric age groups, especially in hospitals /X-ray rooms with low LCD detectability and high IAK, are required.