文献速递乳腺MR

更高场强能否带来更好的诊断准确性?3 T与1.5 T乳腺MRI的前瞻性比较

Does higher field strength translate into better diagnostic accuracy? A prospective comparison of breast MRI at 3 and 1.5 Tesla.
PubMed 文献 · Eur J Radiol · 2019-02-28

本前瞻性单中心研究将982例按标准适应证接受乳腺MRI的患者随机分配至3 T或1.5 T扫描仪,评估BI-RADS分类的诊断准确性。结果显示总体敏感度94.7%、特异度91.4%、准确率91.9%、阴性预测值99.0%;1.5 T与3 T在敏感度、特异度、准确率及阴性预测值上均无显著差异(P=0.29-0.74)。结论为两种场强乳腺MRI诊断准确性相当,同样适用于乳腺成像。

要点

  • 1.5 T与3 T乳腺MRI的敏感度、特异度、准确率及阴性预测值均无显著差异。
  • 总体敏感度94.7%、特异度91.4%、准确率91.9%、阴性预测值99.0%,处于文献报道范围内。
  • 两场强组患者年龄、参考标准结果及MRI适应证均无显著差异。

核心结论

  • 在这项前瞻性单中心研究中,1.5 T与3 T乳腺MRI均表现出相当高的诊断准确性。
  • 1.5 T与3 T同样适用于乳腺成像。

研究设计与样本

  • 前瞻性单中心研究,连续12个月纳入982例按标准适应证(问题解决、术前分期)接受乳腺MRI的患者。
  • 患者被随机分配至3 T或1.5 T扫描仪,两者均配备专用16通道乳腺线圈。
  • 因时间段可用性受限,随机化比例为1/10(3 T/1.5 T)。
  • 1.5 T与3 T协议经优化并符合国际推荐。
  • 所有检查由两位乳腺MRI专家(经验分别>25年和8年)阅片,按乳腺分配BI-RADS分类。

诊断标准与关键数值

  • 总体敏感度94.7%、特异度91.4%、准确率91.9%、阴性预测值99.0%。
  • 1.5 T与3 T配对比较:敏感度94.1% vs 97.9%,特异度91.6% vs 89.3%,准确率92.0% vs 91.2%,阴性预测值98.9% vs 99.3%,均无显著差异(P=0.29-0.74)。
  • 共纳入1961个乳腺,其中1746个(89%)在1.5 T成像,215个(11%)在3 T成像。
  • 1.5 T与3 T队列在患者年龄(P=0.71)、参考标准结果(P=0.09)及MRI适应证(P=0.53)方面无显著差异。

局限性

  • 因时间段可用性受限,随机化比例为1/10(3 T/1.5 T),导致3 T组样本量明显少于1.5 T组(215个乳腺 vs 1746个乳腺)。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: We performed a prospective monocentric study to inter-individually compare the diagnostic accuracy of breast MRI at 1.5 T and 3 T.

METHODS: During a consecutive period of 12 months all 982 patients receiving breast MRI according to standard indications (problem solving, preoperative staging) were randomized to one 3 T or 1.5 T scanner both equipped with dedicated 16-channel breast coils. Protocols at 1.5 T and 3 T were optimized and in line with international recommendations. Due to restricted time slot availabilities, the randomization-key was defined as 1/10 (3 T/1.5 T). All examinations were read by two experts in breast MRI (>25 and 8 years of experience) who assigned a BI-RADS category per breast. Histopathological verification or long-term MRI follow-up (>24 months) served as standard of reference. Results were analyzed using cross tabulations, standard estimates of diagnostic accuracy, Chi-square test and Mann-Whitney U test (alpha = 5%).

RESULTS: 1961 breasts were included. 1746 (89%) were imaged at 1.5 T and 215 at 3 T (11%). The 1.5 T and 3 T study cohort did not show significant differences regarding patient age (P = 0.71), results of the reference standard (P = 0.09) and indication for MRI (P = 0.53). Overall Sensitivity (94.7%), Specificity (91.4%), Accuracy (91.9%) and Negative Predictive Value (99.0%) were within the range of the literature. Pairwise comparison of Sensitivity (1,5 T/3 T: 94.1/97.9%), Specificity (91.6/%89.3%), Accuracy (92.0%/91.2%) and Negative Predictive Value (98.9%/99.3%) were without significant differences (P = 0.29-0.74).

CONCLUSION: In this prospective monocentric study, we identified comparably high diagnostic accuracy for both 1.5 T and 3 T breast MRI. Both 1.5 T and 3 T are equally suited for breast imaging.

来源:Eur J Radiol,查看原文。版权归原作者所有。