文献速递乳腺MR

同时多层回波平面成像用于乳腺良恶性病变的加速扩散加权成像

Simultaneous Multislice Echo Planar Imaging for Accelerated Diffusion-Weighted Imaging of Malignant and Benign Breast Lesions.
PubMed 文献 · Invest Radiol · 2019-01-01

研究比较同时多层(SMS)加速扩散加权EPI与参考EPI对乳腺良恶性病变的诊断价值,68例患者均行两种序列检查。SMS EPI扫描时间从3分27秒缩短至1分53秒,脂肪抑制评分显著更优(3.28 vs 2.97,P<0.001),ADC值及曲线下面积无显著差异,作者认为SMS加速可用于临床乳腺MRI扩散成像。

要点

  • SMS EPI将扫描时间从3分27秒缩短至1分53秒,诊断效能与参考EPI相当。
  • SMS EPI脂肪抑制评分显著优于参考EPI(3.28 vs 2.97,P<0.001,Pearson r=0.49)。
  • 鉴别良恶性病变的ADC阈值:SMS EPI为1.42×10⁻³ mm²/s,参考EPI为1.23×10⁻³ mm²/s。

核心结论

  • SMS加速可用于临床乳腺MRI扩散成像,SMS EPI在显著缩短扫描时间的同时达到与参考EPI相同的诊断准确度。

研究设计与样本

  • 经机构伦理委员会批准,68例患者行参考EPI(TE=54毫秒,TR=9000毫秒,TA 3:27分钟)和SMS EPI(加速因子2,TE=58毫秒,TR=4300毫秒,TA 1:53分钟)及标准MRI方案。

关键影像表现/征象

  • 两种序列伪影和信号缺失均轻微,参考EPI评分4.68 vs SMS EPI 4.65(P=0.52),信号缺失评分4.85 vs 4.77(P=0.14)。
  • SMS EPI脂肪抑制显著优于参考EPI(评分3.28 vs 2.97,P<0.001,Pearson r=0.49)。
  • 良性病变两种EPI序列平均ADC均为1.86×10⁻³ mm²/s;恶性病变参考EPI为0.90×10⁻³ mm²/s,SMS EPI为0.89×10⁻³ mm²/s。
  • 两种EPI序列ADC值(P=0.75)和曲线下面积(SMS 0.985 vs 无SMS 0.975)无显著差异。

诊断标准与关键数值

  • SMS EPI鉴别良恶性病变ADC阈值为1.42×10⁻³ mm²/s(敏感度1,特异度0.88)。
  • 参考EPI鉴别良恶性病变ADC阈值为1.23×10⁻³ mm²/s(敏感度1,特异度0.92)。
  • 排除囊肿后SMS EPI阈值为1.11×10⁻³ mm²/s(敏感度0.89,特异度0.93),参考EPI为1.23×10⁻³ mm²/s(敏感度1,特异度0.87)。

临床应用与阅片要点

  • SMS EPI可在1分53秒完成乳腺扩散加权成像,图像质量与参考EPI相当且脂肪抑制更优,作者认为可用于临床实践。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: Comparison of the diagnostic value of simultaneous multislice (SMS) accelerated diffusion-weighted echo planar imaging (EPI) of malignant and benign lesions of the breast compared with a reference EPI sequence.

MATERIALS AND METHODS: The study was approved by the institutional ethics committee. Sixty-eight patients were examined with a diffusion-weighted EPI (reference EPI; TE = 54 milliseconds; TR = 9000 milliseconds; TA, 3:27 minutes) and a diffusion-weighted SMS accelerated EPI (SMS EPI; acceleration factor 2; TE = 58 milliseconds; TR = 4300 milliseconds; TA, 1:53 minutes) in addition to the standard magnetic resonance imaging (MRI) protocol. Further acquisition parameters were as follows: 3 T MAGNETOM Skyra (Siemens Healthcare, Erlangen, Germany), 2.5-mm isotropic resolution, field of view = 185 to 190 × 350 mm, 62 slices, b = 50 and 800 s/mm with 1 and 4 averages, respectively. A dedicated 16-channel bilateral breast coil was used for imaging. Image quality was evaluated with respect to the presence of artifacts, signal voids, and quality of fat suppression. These parameters were rated using a 5-point Likert scale (1 = very strong to 5 = negligible). The apparent diffusion coefficient (ADC) was measured in 72 focal lesions (46 breast carcinomas and 26 benign lesions), and the diagnostic value of the 2 datasets was statistically evaluated and compared. The evaluation was performed a second time excluding cysts.

RESULTS: Artifacts and signal voids were negligible in both sequences (mean on Likert scale for reference EPI 4.68 vs SMS EPI 4.65, P = 0.52, and mean on Likert scale for reference EPI 4.85 vs SMS EPI 4.77, P = 0.14). Fat suppression was significantly better in SMS EPI (mean on Likert scale 3.28 vs 2.97, P < 0.001, Pearson r = 0.49). For benign lesions, the mean ADC in both EPI sequences was 1.86 · 10 mm/s. For malignant lesions, a mean ADC of 0.90 · 10 mm/s for the reference EPI and 0.89 · 10 mm/s for the SMS EPI was found. No significant difference between the EPI sequences was observed for ADC values (P = 0.75) and for the area under the curve (SMS, 0.985; no SMS, 0.975). The cutoff for differentiation of benign and malignant lesions was at ADC = 1.42 · 10 mm/s for SMS EPI (sensitivity, 1; specificity, 0.88) and at 1.23 · 10 mm/s for the reference EPI (sensitivity, 1; specificity, 0.92). Excluding the cysts, the cutoff for differentiation of benign and malignant lesions was at ADC = 1.11 · 10 mm/s for SMS EPI (sensitivity, 0.89; specificity, 0.93) and at 1.23 · 10 mm/s for the reference EPI (sensitivity, 1; specificity, 0.87).

CONCLUSIONS: Our data indicate that SMS acceleration can be used for diffusion imaging in breast MRI in clinical practice. Simultaneous multislice EPI achieved the same diagnostic accuracy in breast MRI, but in a substantially reduced scan time.

来源:Invest Radiol,查看原文。版权归原作者所有。