文献速递乳腺US

基于定量超声的多参数分类器用于乳腺肿块鉴别

A Quantitative Ultrasound-Based Multi-Parameter Classifier for Breast Masses.
PubMed 文献 · Ultrasound Med Biol · 2019-04-26

研究纳入43例拟行空心针活检的可疑乳腺肿块患者,用临床乳腺超声系统采集射频回波信号,以参考体模法获得系统无关的定量超声参数,并用最小马氏距离分类器区分癌与纤维腺瘤。两参数(有效散射体直径及其异质性指数)分类器ROC曲线下面积为0.98(95%CI 0.95-1.00),三参数(特异性衰减、有效散射体直径及其异质性指数)分类器ROC曲线下面积为0.999(95%CI 0.995-1.000)。作者认为系统无关的定量超声参数组合分类器有希望用于乳腺肿瘤表征,但需更大规模研究进一步验证。

要点

  • 三参数(特异性衰减、有效散射体直径及其异质性指数)分类器ROC曲线下面积达0.999(95%CI 0.995-1.000)。
  • 两参数(有效散射体直径及其异质性指数)分类器ROC曲线下面积为0.98(95%CI 0.95-1.00)。
  • 定量超声参数经参考体模法获得系统无关估计,以最小马氏距离分类器区分乳腺 carcinoma 与纤维腺瘤。

核心结论

  • 系统无关的定量超声参数组合的无模型分类器在区分乳腺 carcinoma 与纤维腺瘤上表现出很高的ROC曲线下面积,作者认为是有希望的工具,但需更大规模研究进一步检验。

研究设计与样本

  • 研究为初步结果报告,纳入43例拟行可疑乳腺肿块空心针活检的患者。
  • 使用配备线阵探头的临床乳腺超声系统采集射频回波信号,以参考体模法获得系统无关的定量超声参数估计。
  • 分类采用最小马氏距离至训练类质心,并以活检结果作为参考标准进行检验,分类性能用ROC曲线下面积评估。

关键影像表现/征象

  • 在乳腺肿块感兴趣区内计算了特异性衰减、平均背散射系数、有效散射体直径、有效散射体直径异质性指数、包络幅度信噪比、Nakagami形状参数m和广义谱最大折叠平均等定量超声参数。

诊断标准与关键数值

  • 两参数分类器使用有效散射体直径与有效散射体直径异质性指数,ROC曲线下面积为0.98(95%CI 0.95-1.00)。
  • 三参数分类器使用特异性衰减、有效散射体直径与有效散射体直径异质性指数,ROC曲线下面积为0.999(95%CI 0.995-1.000)。

临床应用与阅片要点

  • 原文未据此给出临床推荐。

局限性

  • 原文报告为初步结果,并指出需要更大规模研究进一步检验该思路。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

This manuscript reports preliminary results obtained by combining estimates of two or three (among seven) quantitative ultrasound (QUS) parameters in a model-free, multi-parameter classifier to differentiate breast carcinomas from fibroadenomas (the most common benign solid tumor). Forty-three patients scheduled for core biopsy of a suspicious breast mass were recruited. Radiofrequency echo signal data were acquired using clinical breast ultrasound systems equipped with linear array transducers. The reference phantom method was used to obtain system-independent estimates of the specific attenuation (ATT), the average backscatter coefficients, the effective scatterer diameter (ESD) and an effective scatterer diameter heterogeneity index (ESDHI) over regions of interest within each mass. In addition, the envelope amplitude signal-to-noise ratio (SNR), the Nakagami shape parameter, m, and the maximum collapsed average (maxCA) of the generalized spectrum were also computed. Classification was performed using the minimum Mahalanobis distance to the centroids of the training classes and tested against biopsy results. Classification performance was evaluated with the area under the receiver operating characteristic (ROC) curve. The best performance with a two-parameter classifier used the ESD and ESDHI and resulted in an area under the ROC curve of 0.98 (95% confidence interval [CI]: 0.95-1.00). Classification performance improved with three parameters (ATT, ESD and ESDHI) yielding an area under the ROC curve of 0.999 (0.995-1.000). These results suggest that system-independent QUS parameters, when combined in a model-free classifier, are a promising tool to characterize breast tumors. A larger study is needed to further test this idea.

来源:Ultrasound Med Biol,查看原文。版权归原作者所有。

引用

  1. Nasief HG, Rosado-Mendez IM, Zagzebski JA, et al. A Quantitative Ultrasound-Based Multi-Parameter Classifier for Breast Masses. Ultrasound Med Biol. 2019. PMID: 31031035. doi:10.1016/j.ultrasmedbio.2019.02.025

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。