剪切波弹性成像与常规超声评估肾功能的比较:以51Cr-乙二胺四乙酸和99Tc-二巯基丁二酸测定的肾功能为参照
本研究纳入57例接受51Cr-乙二胺四乙酸GFR测定和99mTc-二巯基丁二酸肾闪烁显像的患者,比较剪切波弹性成像(SWE)测得的杨氏模量与常规超声指标对肾功能的评估价值。SWE与左右肾GFR均呈负相关,预测优势功能肾的准确率为94.7%,其AUC(0.800)优于常规超声(0.252–0.415), cutoff值≥5.52 kPa提示肾功能≤60 mL/min/1.73 m2,敏感度82.4%、特异度76.2%。
要点
- SWE预测优势功能肾的准确率为94.7%,AUC 0.800优于常规超声的0.252–0.415。
- SWE cutoff值≥5.52 kPa提示肾功能≤60 mL/min/1.73 m2,敏感度82.4%、特异度76.2%。
- SWE与左右肾GFR均呈负相关(右肾r=-0.635,左肾r=-0.817)。
核心结论
- SWE在评估肾功能和区分优势功能肾方面优于常规超声。
研究设计与样本
- 纳入57例转诊行51Cr-乙二胺四乙酸GFR测定和99mTc-二巯基丁二酸肾闪烁显像的患者,以放射性核素GFR为参考标准,比较SWE杨氏模量与常规超声指标(肾长、体积、皮质厚度、实质回声)与GFR及肾闪烁显像结果的相关性。
诊断标准与关键数值
- SWE与右肾GFR的Spearman相关系数r=-0.635(p<0.0001),与左肾GFR的r=-0.817(p<0.0001);左肾皮质厚度与GFR的r=0.381(p=0.04),左肾体积与GFR的r=0.356(p=0.019);SWE的AUC为0.800,常规超声的AUC为0.252–0.415;SWE预测优势功能肾的准确率为94.7%;SWE cutoff值≥5.52 kPa提示肾功能≤60 mL/min/1.73 m2,敏感度82.4%、特异度76.2%。
临床应用与阅片要点
- 作者认为SWE在评估肾功能和区分优势功能肾方面优于常规超声。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
The purpose of this study was to assess the potential of shear wave elastography (SWE) as an indicator of abnormal kidney function defined by radiolabeled glomerular filtration rate (GFR). Fifty-seven patients referred for 51Cr-ethylenediaminetetraacetic acid GFR and 99mTc-dimercaptosuccinic acid renal scintigraphy were included. Young's modulus (YM) measured with SWE and kidney length, volume, cortical thickness and parenchymal echogenicity measured with conventional ultrasound were correlated with patients' GFR and renal scintigraphy results. Spearman correlation coefficients between SWE and GFR were negative for the right (r = -0.635, p < 0.0001) and left (r = -0.817, p < 0.0001) kidneys. Positive correlations between left renal cortical thickness (r = 0.381, p = 0.04) and left kidney volume (r = 0.356, p = 0.019) with GFR were reported. SWE correctly predicted the dominant functioning kidney in 94.7% of cases. The area under the receiver operating characteristic curve for SWE (0.800) was superior to that for conventional ultrasound (0.252-0.415). The cutoff value of ≥5.52 kPa suggested a kidney function ≤60 mL/min/1.73 m2 (82.4% sensitivity and 76.2% specificity). SWE has advantages over conventional ultrasound in assessing kidney function and distinguishing the dominant functioning kidney.
引用
- Leong SS, Wong JHD, Md Shah MN, et al. Comparison of Shear Wave Elastography and Conventional Ultrasound in Assessing Kidney Function as Measured Using 51Cr-ethylenediaminetetraacetic Acid and 99Tc-Dimercaptosuccinic Acid. Ultrasound Med Biol. 2019. PMID: 30962016. doi:10.1016/j.ultrasmedbio.2019.01.024
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。