文献速递腹部技术与安全PETCT

基于头戴显示器视觉反馈的相位式呼吸门控PET成像的临床效用

The clinical utility of phase-based respiratory gated PET imaging based on visual feedback with a head-mounted display system.
PubMed 文献 · Br J Radiol · 2019-04-30

研究纳入41例肝胆或胰腺癌患者,20例接受RGV-PET(相位式呼吸门控PET结合头戴显示器引导的视觉反馈),21例接受RG-PET(相位式呼吸门控PET)。RGV-PET的SUVpeak变化率变异系数为7.01±4.43,显著低于RG-PET的10.72±5.74(p<0.05);RGV-PET在bins 1和2的SUVpeak变化率较RG-PET显著改善(p<0.05);RGV-PET的四维CT伪影视觉评分显著低于RG-PET(p<0.05)。结论认为RGV-PET有助于呼吸稳定并改善呼吸门控成像质量。

要点

  • RGV-PET的SUVpeak变化率变异系数为7.01±4.43,显著低于RG-PET的10.72±5.74(p<0.05)。
  • RGV-PET在bins 1和2的SUVpeak变化率较RG-PET显著改善(p<0.05)。
  • RGV-PET的四维CT伪影视觉评分显著低于RG-PET(p<0.05)。

核心结论

  • RGV-PET有助于呼吸稳定并改善呼吸门控成像质量。
  • RGV-PET适用于多种PET/CT呼吸门控成像,可能改善PET图像的定量性。

研究设计与样本

  • 研究纳入41例肝胆或胰腺癌患者,其中20例接受RGV-PET,其余21例接受RG-PET。
  • RGV-PET为相位式呼吸门控PET成像与头戴显示器引导的视觉反馈相结合的新方法。
  • 测量原发病灶在RG-PET和RGV-PET各五个bins中获得的SUVpeak,并基于非门控PET成像计算SUVpeak变化率。
  • 通过比较RG-PET与RGV-PET的SUVpeak变化率变异系数来评估定量变异,并对四维CT伪影发生率进行视觉评分定性评估。

诊断标准与关键数值

  • RGV-PET的SUVpeak变化率平均值的变异系数为7.01±4.43,显著低于RG-PET的10.72±5.74(p<0.05)。
  • RGV-PET在bins 1和2的SUVpeak变化率较RG-PET显著改善(p<0.05)。
  • RGV-PET的四维CT伪影视觉评分显著低于RG-PET(p<0.05)。

临床应用与阅片要点

  • RGV-PET将相位式呼吸门控PET与头戴显示器引导的视觉反馈结合,可能有助于呼吸稳定。
  • RGV-PET可能改善呼吸门控PET成像的定量性。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: We developed a new respiratory-gated positron emission tomography (PET) imaging method (RGV-PET) that phase-based respiratory gated PET imaging (RG-PET) combine with head-mounted display (HMD)-guided "visual feedback." The purpose of this study was to investigate whether RGV-PET is effective at improving the quantitative measurement of tracer uptake in tumors using the phase-based respiratory gating method.

METHODS: Of the 41 enrolled patients with hepatobiliary or pancreatic cancer, 20 patients underwent RGV-PET and the remaining 21 patients underwent RG-PET. We measured the peak standardized uptake value (SUVpeak) of the primary lesion in each five bins obtained from both RG-PET and RGV-PET. The SUVpeak change rate calculated based on the ungated PET imaging. To evaluate the quantitative variation, the coefficient of variation of the SUVpeak change rate was compared between RG-PET and RGV-PET. In addition, we performed qualitative evaluation using visual score for the incidence of artifacts on four-dimensional-CT.

RESULTS: The coefficient of variation of the average SUVpeak change rate in RGV-PET was 7.01 ± 4.43, which was significantly lower than the values in RG-PET (10.72 ± 5.74, p < 0.05). A significant improvement in the SUVpeak change rate of RGV-PET was obtained in bins 1 and 2 compared to RG-PET ( p < 0.05). The visual score of RGV-PET was significantly lower than RG-PET ( p < 0.05).

CONCLUSION: RGV-PET was effective for respiratory stabilization and improved respiratory gating imaging quality.

ADVANCES IN KNOWLEDGE: The RGV-PET is applicable to various PET/CT respiratory gating imaging and may improve the quantitativeness of PET images.