文献速递心血管MR

疾病状态下两种电影衍生应变软件程序的可重复性比较

A comparison of the reproducibility of two cine-derived strain software programmes in disease states.
PubMed 文献 · Eur J Radiol · 2019-01-23

本研究为前瞻性研究,纳入主动脉瓣狭窄、血液透析及糖尿病患者各10例,于1.5T和3T下比较特征追踪(FT)与组织追踪(TT)软件衍生应变参数的重测可重复性。结果显示FT与TT的GCS和GLS可重复性均优良,而TT对纵向和圆周PEDSR的重测可重复性优于FT。

要点

  • FT与TT软件测得的GCS和GLS在疾病人群中重测可重复性均优良。
  • TT软件对纵向和圆周PEDSR的重测可重复性优于FT软件。
  • Bland-Altman分析显示TT应变测量无系统偏倚且一致性界限更窄。

核心结论

  • FT与TT软件在多种疾病中测量GCS和GLS的可重复性均优良,TT对纵向和圆周PEDSR的量化重测可重复性优于FT。

研究设计与样本

  • 前瞻性研究,纳入主动脉瓣狭窄、血液透析及糖尿病患者各10例,共30例,于1.5T和3T下进行重测可重复性扫描,比较TT与FT软件计算的GCS、GLS、圆周PEDSR和纵向PEDSR。

诊断标准与关键数值

  • FT衍生GCS的CoV在主动脉瓣狭窄、糖尿病、血液透析患者中分别为8.1%、5%、7.9%,TT衍生GCS分别为3.3%、9.2%、5.4%。
  • FT衍生GLS的CoV在主动脉瓣狭窄、糖尿病、血液透析患者中分别为8%、6.4%、8.2%,TT衍生GLS分别为5.3%、4.8%、7%。
  • FT衍生圆周PEDSR的CoV在主动脉瓣狭窄、糖尿病、血液透析患者中分别为18.2%、18%、17.4%,TT衍生圆周PEDSR分别为6.1%、11.7%、11%。
  • FT衍生纵向PEDSR的CoV在主动脉瓣狭窄、糖尿病、血液透析患者中分别为18.2%、18.9%、18.3%,TT衍生纵向PEDSR分别为8.9%、9.1%、11.4%。
  • Bland-Altman分析显示TT衍生应变测量无系统偏倚且一致性界限更窄。

临床应用与阅片要点

  • 在疾病状态下测量GCS和GLS时,FT与TT软件均可提供优良的重测可重复性。
  • 量化纵向和圆周PEDSR时,TT软件的重测可重复性优于FT软件。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND: Systolic strain and peak-early diastolic strain rate (PEDSR) measure subclinical cardiac dysfunction. These parameters can be derived from cardiovascular magnetic resonance (CMR) cine images using new software packages, but the comparative test-retest reproducibility of these software in disease states is unknown. This study compared the test-retest reproducibility of strain measures derived from two software packages (feature-tracking software (FT) and tissue-tracking (TT)) in disease populations with preserved ejection fractions.

METHODS: This was a prospective study of 10 patients with aortic stenosis (AS), 10 haemodialysis patients and 10 diabetic patients at 1.5 and 3-Tesla. 30 subjects underwent test-retest reproducibility scans of global circumferential strain (GCS), global longitudinal strain (GLS), circumferential-PEDSR and longitudinal-PEDSR calculated using TT and FT software.

RESULTS: Test-retest reproducibility of GCS and GLS were similar for FT and TT across patient groups. Coefficient of variability (CoV) for FT-derived GCS 8.1%, 5% and 7.9% for AS, diabetic and haemodialysis patients, compared to 3.3%, 9.2% and 5.4% for TT-derived GCS, with CoV for FT-derived GLS 8%, 6.4% and 8.2% for AS, diabetic and haemodialysis patients, compared to 5.3%, 4.8% and 7% for TT-derived GLS). Reproducibility of FT-derived circumferential and longitudinal-PEDSR was worse than TT-derived circumferential and longitudinal-PEDSR (CoV for FT-derived circumferential-PEDSR 18.2%, 18% and 17.4% for AS, diabetic and haemodialysis patients, compared to 6.1%, 11.7% and 11% for TT-derived circumferential-PEDSR with CoV for FT-derived longitudinal PEDSR 18.2%, 18.9%, 18.3% for AS, diabetic and haemodialysis patients, compared to 8.9%, 9.1% and 11.4% for TT-derived longitudinal-PEDSR). Bland-Altman analysis revealed no systematic bias with tighter limits of agreement for TT-derived strain measures.

CONCLUSIONS: Reproducibility of GCS and GLS are excellent with FT and TT software across diseases. TT had superior test-retest reproducibility for quantification of longitudinal and circumferential-PEDSR than FT-derived PEDSR across diseases.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. Graham-Brown MP, Gulsin GS, Parke K, et al. A comparison of the reproducibility of two cine-derived strain software programmes in disease states. Eur J Radiol. 2019. PMID: 30927959. doi:10.1016/j.ejrad.2019.01.026

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。