双层光谱探测器双能心脏CT评估心肌晚期碘增强及细胞外容积定量
单中心回顾性研究纳入40例患者行双层光谱探测器CT晚期碘增强(LIE)检查,其中21例同期行心脏MRI。虚拟单能50 keV图像的对比噪声比和主观图像质量优于常规120 kVp及碘密度图,LIE检出与心脏MRI的一致性极好(κ=0.90),CT与MRI计算的细胞外容积(ECV)相关性极好(r²=0.94)。双能心脏CT可评估心肌LIE并定量ECV,结果与心脏MRI相当。
要点
- 虚拟单能50 keV图像上LIE观察者间一致性κ=0.87,优于常规120 kVp(κ=0.70)和碘密度图(κ=0.83)。
- 虚拟单能50 keV检出晚期强化病灶与心脏MRI一致性极好(κ=0.90),常规120 kVp仅为好(κ=0.66)。
- CT与心脏MRI计算的ECV相关性极好(r²=0.94),支持双能CT定量心肌ECV。
核心结论
- 双能心脏CT可评估心肌晚期碘增强并定量细胞外容积,结果与心脏MRI相当。
研究设计与样本
- 单中心回顾性研究,共40例患者使用双层光谱探测器CT行LIE检查,其中21例同期行心脏MRI,比较常规120 kVp、虚拟单能50 keV和碘密度图三组图像。
关键影像表现/征象
- 虚拟单能50 keV图像的对比噪声比显著高于常规120 kVp图像。
- 虚拟单能50 keV图像的平均主观图像质量评分显著高于常规120 kVp和碘密度图。
诊断标准与关键数值
- LIE观察者间一致性:虚拟单能50 keV κ=0.87,碘密度图 κ=0.83,常规120 kVp κ=0.70。
- 与心脏MRI检出晚期强化病灶的一致性:虚拟单能50 keV κ=0.90,碘密度图 κ=0.87,常规120 kVp κ=0.66。
- CT与心脏MRI计算ECV的相关性 r²=0.94。
临床应用与阅片要点
- 双能心脏CT的虚拟单能50 keV图像可用于评估心肌晚期碘增强,观察者间一致性及与心脏MRI的一致性均优于常规120 kVp图像。
- 双能心脏CT可定量细胞外容积,与心脏MRI计算结果相关性极好(r²=0.94)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To explore the usefulness of myocardial late iodine enhancement (LIE) and extracellular volume (ECV) quantification by using dual-energy cardiac CT.
MATERIALS AND METHODS: In this single-center retrospective study, a total of 40 patients were evaluated with LIE CT by using a dual-layer spectral detector CT system. Among these, 21 also underwent cardiac MRI. Paired image sets were created by using standard imaging at 120 kVp, virtual monochromatic imaging (VMI) at 50 keV, and iodine density imaging. The contrast-to-noise ratio and image quality were then compared. Two observers assessed the presence of LIE and calculated the interobserver agreements. Agreement between CT and cardiac MRI when detecting late-enhancing lesions and calculating the ECV was also assessed.
RESULTS: The contrast-to-noise ratio was significantly higher by using VMI than by using standard 120-kVp imaging, and the mean visual image quality score was significantly higher by using VMI than by using either standard or iodine density imaging. For interobserver agreement of visual detection of LIE, the agreement for VMI was excellent and the κ value (κ, 0.87) was higher than that for the standard 120-kVp (κ, 0.70) and iodine density (κ, 0.83) imaging. For detecting late-enhancing lesions, agreement with cardiac MRI was excellent by using VMI (κ, 0.90) and iodine density imaging (κ, 0.87) but was only good by using standard 120-kVp imaging (κ, 0.66). Quantitative comparisons of the ECV calculations by using CT and cardiac MRI showed excellent correlation (r 2 = 0.94).
CONCLUSION: Dual-energy cardiac CT can assess myocardial LIE and quantify ECV, with results comparable to those obtained by using cardiac MRI.© RSNA, 2019See also the commentary by Litt in this issue.
引用
- Oda S, Emoto T, Nakaura T, et al. Myocardial Late Iodine Enhancement and Extracellular Volume Quantification with Dual-Layer Spectral Detector Dual-Energy Cardiac CT. Radiol Cardiothorac Imaging. 2019. PMID: 33778497. doi:10.1148/ryct.2019180003
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。