基于病灶大小评估和优化肺癌评估中的新型PET重建算法——贝叶斯惩罚似然重建
本研究对46例肺癌患者的FDG PET/CT扫描分别用OSEM和β值为200、300、400、500、1000的贝叶斯惩罚似然(BPL)重建,比较肝脏信噪比、肝脏SUVmean及肿瘤SUVmax/SUVmean,并以视觉评分评估图像质量。BPL在β值500时视觉评分最高,肿瘤SUVmax随β值增大而降低,小病灶受BPL影响更大;结论认为BPL可改善图像质量而不影响定量,β值500在本研究中似乎最佳。
要点
- BPL重建的肝脏信噪比除β值200外均显著高于OSEM,肝脏SUVmean各算法间无统计学差异。
- 肿瘤SUVmax和SUVmean随β值增大而降低,β值500时视觉评分最高、β值200时最低。
- 小尺寸组中OSEM与β值200 BPL之间及β值200与1000 BPL之间的SUVmax百分比差异显著大于其他尺寸组(p<0.01)。
核心结论
- BPL算法在不影响图像定量的情况下改善图像质量,本研究中β值500似乎最佳,较小肿瘤受BPL影响更大。
研究设计与样本
- 本研究为影像重建算法比较研究,纳入46例肺癌患者的FDG PET/CT扫描。
- 分别用OSEM和β值为200、300、400、500、1000的BPL进行重建。
- 测量肝脏信噪比、肝脏SUVmean及肿瘤SUVmax和SUVmean,并按肿瘤大小分为三组计算SUVmax百分比差异,同时以视觉评分评估图像质量。
关键影像表现/征象
- BPL的肝脏信噪比显著高于OSEM,但β值200的BPL除外。
- 各算法间肝脏SUVmean无统计学差异。
- 肿瘤SUVmax和SUVmean随β值增大而降低。
- β值200的BPL肿瘤SUVmax显著高于OSEM(p<0.01),β值400、500、1000的BPL肿瘤SUVmax显著低于OSEM(p<0.01)。
- 视觉分析显示β值500的BPL评分最高、β值200的BPL评分最低。
诊断标准与关键数值
- BPL的β值取200、300、400、500、1000;肝脏信噪比除β值200外均显著高于OSEM。
- β值200的BPL肿瘤SUVmax显著高于OSEM(p<0.01),β值400、500、1000的BPL肿瘤SUVmax显著低于OSEM(p<0.01)。
- 小尺寸组中OSEM与β值200 BPL之间及β值200与β值1000 BPL之间的SUVmax百分比差异显著大于其他尺寸组(p<0.01)。
临床应用与阅片要点
- 本研究中β值500的BPL视觉评分最高,作者认为其似乎为最佳β值。
- 较小肿瘤的SUVmax受BPL重建影响更大,阅片和报告时需结合病灶大小解读SUVmax变化。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE. The purpose of this study was to characterize the Bayesian penalized likelihood (BPL) reconstruction algorithm in comparison with an ordered subset expectation maximization (OSEM) reconstruction algorithm and to determine its optimal penalization factor (expressed as a beta value) for clinical use. MATERIALS AND METHODS. FDG PET/CT scans of 46 patients with lung cancer were reconstructed using OSEM and BPL with beta values of 200, 300, 400, 500, and 1000. The liver signal-to-noise ratio, mean standardized uptake value (SUVmean) of the liver, and maximum standardized uptake value (SUVmax) and SUVmean of the cancers were measured. Tumors were categorized into three size groups, and the percentage difference in the tumor SUVmax between OSEM and BPL with a beta value of 200 as well as the percentage difference in the SUVmax between BPL with a beta value of 200 and BPL with a beta value of 1000 were calculated. Image quality was assessed by visual scoring. RESULTS. BPL showed a significantly higher liver signal-to-noise ratio than OSEM, except for BPL with a beta value of 200. The liver SUVmean showed no statistical difference among all algorithms. The SUVmax and SUVmean of tumors decreased as the beta value increased. BPL with a beta value of 200 produced a significantly higher tumor SUVmax than did OSEM (p < 0.01), and BPL with a beta value of 400, 500, or 1000 produced a significantly lower tumor SUVmax than did OSEM (p < 0.01). Visual analysis showed the highest and lowest scores for BPL with beta values of 500 and 200, respectively. In the small size group, the percentage difference in the SUVmax between OSEM and BPL with a beta value of 200 and the percentage difference in the SUVmax between BPL with a beta value of 200 and BPL with a beta value of 1000 were significantly larger than that in the other size groups (p < 0.01). CONCLUSION. The BPL algorithm improves image quality without compromising image quantification. A beta value of 500 appeared to be optimal in this study. Smaller tumors were more influenced by BPL.