儿童患者多排CT与第三代双源双能CT的辐射剂量与图像质量趋势
该研究回顾性比较了儿科肿瘤分期(颈部、胸部、腹部)和低剂量胸部CT患者,分别使用16排LightSpeed CT(2017年)与第三代Somatom Force双源CT(2018年)扫描,按年龄≤10岁与>10岁分组。结果显示双源CT在肿瘤患者中总DLP降低5-6倍、图像质量更优,低剂量胸部CT中≤10岁组总DLP降低3倍,图像质量更佳且运动伪影减少。
要点
- 双源CT在儿童肿瘤分期中总DLP较16排CT降低5-6倍,图像质量更优。
- 低剂量胸部CT中,双源CT在≤10岁患儿总DLP降低3倍,图像质量更好。
- 双源CT的虚拟平扫图像有助于降低肿瘤患儿的辐射剂量。
核心结论
- 第三代双源CT在更低辐射剂量下提供高质量图像并减少运动伪影。
研究设计与样本
- 回顾性研究,纳入<20岁儿童患者,分为肿瘤分期组(颈部、胸部、腹部)和低剂量胸部CT组,每组再按年龄≤10岁与>10岁分组,分别于2017年使用16排LightSpeed CT(GE Healthcare)或2018年使用Somatom Force双源CT(Siemens Healthineers)扫描;图像质量由两名放射科医师共识评估五级半定量量表,采用非参数检验比较。
关键影像表现/征象
- 双源CT在肿瘤患者中实现显著更低的kVp和管电流,且图像质量更优。
- 双源CT在低剂量胸部CT中图像质量更优。
- 双源CT减少运动伪影。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
AIM: To provide an overview on dose reduction and image quality after the installation of a third-generation dual-source CT (dsCT) in a Pediatric Radiology Department.
MATERIALS AND METHODS: We included pediatric patients (< 20 years old) undergoing CT for oncological staging (neck, chest and abdomen) or low-dose chest CT for lung diseases. Each of these two groups were further divided in two age groups (≤ or > 10 years old) including patients scanned in the same period of two consecutive years, in 2017 with a 16-row LightSpeed CT (GE Healthcare) or in 2018 with a Somatom Force dsCT (Siemens Healthineers). Technical parameters such as kVp, mAs, slice thickness, exposure times and dose indicators were retrieved and compared. Image quality was evaluated in consensus by two radiologists on a five-point semiquantitative scale. Nonparametric tests were used.
RESULTS: In oncological patients, significantly lower kVp and tube current with better image quality were achieved with the dsCT. Radiation dose (total DLP) was 5-6 times lower with dsCT, thanks also to virtual non-contrast images. In low-dose chest CT, the frequent use of tin filter required higher tube current; a total DLP 3 times lower was achieved with dsCT in patients ≤ 10 years old. The image quality was better with the dsCT in low-dose chest CT protocols.
CONCLUSION: The third-generation dsCT provides high-quality images with reduced motion artifacts at lower dose.