文献速递胸部技术与安全CT

超低剂量胸部计算机断层扫描:迭代重建水平对图像质量的影响

Ultra-low dose chest computed tomography: Effect of iterative reconstruction levels on image quality.
PubMed 文献 · Eur J Radiol · 2019-02-18

本研究在体模上测试固定超低剂量(CTDIvol 0.25、0.49、0.74 mGy)与不同探测器准直(40、80 mm)条件下,比较滤波反投影与不同水平ASIR-V迭代重建的图像质量。结果显示ASIR-V较FBP降低图像噪声5%–55%,提高SNR与CNR,但高比例ASIR-V使噪声纹理更粗糙、出现斑块状伪影,提示ASIR-V可能有助于改善超低剂量胸部CT图像质量。

要点

  • ASIR-V较FBP降低图像噪声5%–55%,SNR随ASIR-V水平升高而升高。
  • NI与ASIR-V水平呈正相关,100% ASIR-V噪声纹理最粗糙、斑块感最明显。
  • FBP与ASIR-V≤20%的NTD无显著差异(p≤0.087)。

核心结论

  • 与FBP相比,ASIR-V在所有测试的超低剂量下均改善了定量图像质量参数,且在不同体模大小和准直条件下表现一致,ASIR-V可能有助于改善超低剂量胸部CT的图像质量。

研究设计与样本

  • 体模研究,使用Kyoto Kagaku Lungman胸部体模(加或不加扩展环),以固定超低剂量0.25、0.49、0.74 mGy CTDIvol及40、80 mm准直扫描,采用肺重建核、FBP及不同ASIR-V水平(10%–100%)重建。

关键影像表现/征象

  • 与FBP相比,ASIR-V使图像噪声降低5%–55%,且不受剂量和准直影响。
  • 在0.74 mGy、40 mm准直下,ASIR-V使最低对比度病灶的CNR从12.7(FBP)升至42.1(100% ASIR-V)。
  • 小体模测得的SNR和CNR高于大体模。
  • 在所有剂量、准直和体模大小下,FBP与各ASIR-V水平的HU测量值一致。
  • 随ASIR-V水平升高,NPS曲线向低频左移,NTD与ASIR-V水平呈正相关(r≥0.946,p≥0.001),提示噪声纹理更斑块化、更粗糙,100% ASIR-V时最明显。

诊断标准与关键数值

  • NTD以偏离3个标准差以外的像素百分比计算,用于评估迭代重建特有伪影。
  • FBP与ASIR-V≤20%的NTD差异无统计学显著(p≤0.087)。
  • SNR与ASIR-V水平呈正相关(r≥0.925,p≤0.001)。

临床应用与阅片要点

  • 超低剂量胸部CT采用ASIR-V重建时,提高ASIR-V水平可降低噪声、提高SNR和CNR,但需留意高比例ASIR-V所致噪声纹理粗糙和斑块状伪影。
  • FBP与ASIR-V≤20%的NTD无显著差异,提示低比例ASIR-V时噪声纹理改变较小。
  • ASIR-V改善定量图像质量参数的表现与体模大小和准直条件一致。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To optimize image quality and radiation dose of chest CT with respect to various iterative reconstruction levels, detector collimations and body sizes.

METHOD: A Kyoto Kagaku Lungman with and without extensions was scanned using fixed ultra-low doses of 0.25, 0.49 and 0.74 mGy CTDIvol, and collimations of 40 and 80 mm. Images were reconstructed with the lung kernel, filtered back projection (FBP) and different ASIR-V levels (10-100%). Contrast-to-noise ratios (CNR) were calculated for 12 mm simulated lesions of different densities in the lung. Image noise, signal-to-noise ratios (SNR), variations in Hounsfield units (HU), noise power spectrum (NPS) and noise texture deviations (NTD) were evaluated for all reconstructions. NTD was calculated as percentage of pixels outside 3 standard deviations to evaluate IR-specific artefacts.

RESULTS: Compared to the FBP, image noise reduced (5-55%) with ASIR-V levels irrespective of dose or collimation. SNR correlated positively (r ≥ 0.925, p ≤ 0.001) with ASIR-V levels at all doses, collimations, and phantom sizes. ASIR-V enhanced the CNR of the lesion with the lowest contrast from 12.7-42.1 (0-100% ASIR-V) at 0.74 mGy with 40 mm collimation. As expected, higher SNR and CNR were measured in the smaller phantom than the bigger phantom. Uniform HU were observed between FBP and ASIR-V levels at all doses, collimations, and phantom sizes. NPS curves left-shifted towards lower frequencies at increasing levels of ASIR-V irrespective of collimation. A positive correlation (r ≥ 0.946, p ≥ 0.001) was observed between NTD and ASIR-V levels. NTD of the FBP was not significantly (p ≤ 0.087) different from NTD of ASIR-V ≤ 20%. The data from the NPS and NTD indicates a blotchier and coarser noise texture at higher levels of ASIR-V, especially at 100% ASIR-V.

CONCLUSION: In comparison with the FBP technique, ASIR-V enhanced quantitative image quality parameters at all ultra-low doses tested. Moreover, the use of ASIR-V showed consistency with body size and collimation. Hence, ASIR-V may be useful for improving image quality of chest CT at ultra-low doses.

来源:Eur J Radiol,查看原文。版权归原作者所有。