文献速递技术与安全胸部US

简易实验模型产生的超声B线样伪影为破解B线关键问题提供线索

Ultrasonic B-Line-Like Artifacts Generated with Simple Experimental Models Provide Clues to Solve Key Issues in B-Lines.
PubMed 文献 · Ultrasound Med Biol · 2019-04-26

研究评估超声机器设置对健康模型单条B线形态的影响并分析频谱,同时用简易实验模型产生B线样伪影并评估设置影响。B线显示受空间复合成像、聚焦区和频率影响,B线与非B线区域频谱中心频率和带宽相同,B线样伪影可由橘子的纺锤形汁囊、食用线状魔芋凝胶、玻璃珠和玻璃板产生。研究提示部分B线的物理基础是多次混响。

要点

  • 空间复合成像、聚焦区和频率可影响B线显示。
  • B线与非B线区域同深度频谱的中心频率和带宽相同。
  • 部分B线的物理基础为多次混响。

核心结论

  • 研究提示部分B线的物理基础是多次混响。
  • B线样伪影为B线物理基础、超声-病理相关性和机器设置影响等关键问题提供线索。

研究设计与样本

  • 研究评估超声机器设置对健康模型单条B线形态的影响并分析频谱。
  • 研究设计简易实验模型产生B线样伪影并评估机器设置对其形态的影响。

关键影像表现/征象

  • B线显示受空间复合成像、聚焦区和频率影响。
  • B线与非B线区域同深度频谱的中心频率和带宽相同。
  • 橘子的纺锤形汁囊、食用线状魔芋凝胶、玻璃珠和玻璃板可产生B线样伪影。
  • B线样伪影的显示同样受空间复合成像、聚焦区和频率影响。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

We evaluated the influence of settings on an ultrasound machine on the configuration of a single B-line in a healthy model and analyzed the frequency spectrum. We also devised simple experimental models that generated B-line-like artifacts and evaluated the influence of the machine settings on the configuration. Visualization of B-lines was affected by the spatial compound imaging, the focal zone and the frequency. The spectra of both the B-line and non-B-line region at the same depth had the same center frequency and bandwidth. B-line-like artifact was generated by a spindle-shaped juice sac of a mandarin orange, an edible string-shaped glucomannan gel, glass beads and glass plates. Visualization of B-line-like artifacts was also affected by these machine settings. Our study indicated that the physical basis of some B-lines is multiple reverberations. B-line-like artifacts provide clues for solving key issues, such as the physical basis of B-lines, the sonographic-pathologic correlation in B-lines and the effects of machine settings.

来源:Ultrasound Med Biol,查看原文。版权归原作者所有。

引用

  1. Kameda T, Kamiyama N, Kobayashi H, et al. Ultrasonic B-Line-Like Artifacts Generated with Simple Experimental Models Provide Clues to Solve Key Issues in B-Lines. Ultrasound Med Biol. 2019. PMID: 31031034. doi:10.1016/j.ultrasmedbio.2019.03.003

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。