自动乳腺超声在无症状女性中鉴别乳腺肿块良恶性的诊断性能:与手持超声的对比研究
本研究从门诊女性中随机按比例选取398例、共599个肿块,同时行手持超声(HHUS)与自动乳腺超声(ABUS)检查,以病理或至少1年随访为参考标准。ABUS与HHUS在准确率(80.1% vs 80.6%)、特异度(77.62% vs 80.24%)、阳性预测值(46.12% vs 46.46%)和阴性预测值(97.96% vs 95.67%)上无显著差异;ABUS敏感度(92.23% vs 82.52%,P<.01)和AUC(0.85 vs 0.81,P<.05)显著高于HHUS。结论认为ABUS在鉴别乳腺肿块良恶性方面优于HHUS,尤其在特异度方面。
要点
- ABUS敏感度92.23%高于HHUS的82.52%(P<.01),AUC 0.85高于0.81(P<.05)。
- ABUS与HHUS的准确率、特异度、阳性预测值、阴性预测值均无显著差异。
- ABUS与病理最大径相关系数r=0.885,HHUS为0.855,差异无统计学意义(P>.05)。
核心结论
- 自动乳腺超声在鉴别乳腺肿块良恶性方面优于手持超声,尤其在特异度方面。
研究设计与样本
- 对比研究,从门诊随机按比例选取398例女性共599个肿块,同时行HHUS与ABUS检查,以病理结果或至少1年随访为参考标准。
诊断标准与关键数值
- 最终BI-RADS 2类和3类肿块判为良性,BI-RADS 4类和5类判为恶性,诊断由病理或至少1年随访确认。
- ABUS敏感度92.23% vs HHUS 82.52%(P<.01),AUC 0.85 vs 0.81(P<.05)。
- ABUS准确率80.1% vs HHUS 80.6%,特异度77.62% vs 80.24%,阳性预测值46.12% vs 46.46%,阴性预测值97.96% vs 95.67%,均无显著差异。
- ABUS与病理最大径相关系数r=0.885,HHUS与病理r=0.855,差异无统计学意义(P>.05)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: Our aim was to investigate the diagnostic potential of an automated breast ultrasound (ABUS) system in differentiating benign and malignant breast masses compared with handheld ultrasound (HHUS).
METHODS: Women were randomly and proportionally selected from outpatients and underwent both HHUS and ABUS examinations. Masses with final American College of Radiology Breast Imaging Reporting and Data System categories 2 and 3 were considered benign. Masses with final Breast Imaging Reporting and Data System categories 4 and 5 were considered malignant. The diagnosis was confirmed by pathologic results or at least a 1-year follow-up. Automated breast US and HHUS were compared on the basis of their sensitivity, specificity, positive predictive value, negative predictive value, and accuracy. Diagnostic consistency and areas under the receiver operating characteristic curves were analyzed. The maximum diameters of masses were compared among HHUS, ABUS, and pathologic results.
RESULTS: A total of 599 masses in 398 women were confirmed by pathologic results or at least a 1-year follow-up; 103 of 599 masses were malignant, and 496 were benign. There were no significant differences between ABUS and HHUS in terms of diagnostic accuracy (80.1% versus 80.6%), specificity (77.62% versus 80.24%), positive predictive value (46.12% versus 46.46%), and negative predictive value (97.96% versus 95.67%). There were significant differences in sensitivity (92.23% versus 82.52%; P < .01) and areas under the curve (0.85 versus 0.81; P < .05) between ABUS and HHUS. The correlation of the maximum diameter was slightly higher between ABUS and pathologic results (r = 0.885) than between HHUS and pathologic results (r = 0.855), but the difference was not significant (P > .05).
CONCLUSIONS: Automated breast US is better than HHUS in differentiating benign and malignant breast masses, especially with respect to specificity.
引用
- Niu L, Bao L, Zhu L, et al. Diagnostic Performance of Automated Breast Ultrasound in Differentiating Benign and Malignant Breast Masses in Asymptomatic Women: A Comparison Study With Handheld Ultrasound. J Ultrasound Med. 2019. PMID: 30912178. doi:10.1002/jum.14991