文献速递乳腺技术与安全PET

评估小视野混合伽马相机用于围手术期碘-125粒子定位

Assessing a small field of view hybrid gamma camera for perioperative iodine-125 seed localisation.
PubMed 文献 · Br J Radiol · 2019-04-24

本临床前离体研究定量和定性评估了小视野混合伽马相机(HGC)对碘-125粒子的成像性能,系统灵敏度为0.41 cps/MBq(源到准直器距离48 mm),空间分辨率为1.53 ± 0.23 mm(源到准直器距离10 mm)。相机可在1分钟采集时间内于源到准直器距离63 mm处探测到碘-125粒子,但在散射材料下探测深度限于皮肤表面下20 mm以内(源到准直器距离63 mm、粒子活度2.43 MBq)。结论认为HGC能够探测碘-125粒子,可能成为放射性粒子定位的有用工具,并额外提供混合光学-γ图像用于引导保乳手术。

要点

  • HGC对碘-125的系统灵敏度为0.41 cps/MBq(源到准直器距离48 mm),空间分辨率为1.53 ± 0.23 mm(源到准直器距离10 mm)。
  • 无散射材料时,HGC可在1分钟采集时间内于源到准直器距离63 mm处探测到碘-125粒子。
  • 有散射材料模拟深部肿瘤时,探测深度限于皮肤表面下20 mm以内(源到准直器距离63 mm、粒子活度2.43 MBq)。

核心结论

  • 这项初步离体研究表明,HGC能够探测碘-125粒子,可能成为放射性粒子定位的有用工具,并额外提供混合光学-γ图像用于引导保乳手术。

研究设计与样本

  • 初步离体研究,使用定制体模在不同源深度、源到准直器距离、活度水平、采集时间和源间距条件下对相机系统进行定量和定性评估。

关键影像表现/征象

  • HGC对碘-125的系统灵敏度为0.41 cps/MBq(源到准直器距离48 mm)。
  • HGC对碘-125的空间分辨率为1.53 ± 0.23 mm(源到准直器距离10 mm)。
  • 无散射材料时,相机可在源到准直器距离63 mm处、1分钟采集时间内探测到碘-125粒子。
  • 有散射材料模拟深部肿瘤时,探测限于皮肤表面下20 mm以内(源到准直器距离63 mm、粒子活度2.43 MBq)。
  • 对混合图像的主观评估显示HGC具有定位碘-125粒子的能力。

临床应用与阅片要点

  • 作者认为HGC可能成为放射性粒子定位的有用工具,并额外提供混合光学-γ图像用于引导保乳手术。
  • 作者指出小视野HGC可提供碘-125放射性粒子的高分辨率融合光学-γ图像,提示其可能在放射性粒子定位等术中外科操作中有潜在用途。

局限性

  • 原文结论称此为初步离体研究。
  • 有散射材料时探测深度限于皮肤表面下20 mm以内(源到准直器距离63 mm、粒子活度2.43 MBq)。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To examine the imaging capability of a novel small field of view hybrid gamma camera (HGC) using 125I seeds prior to surgical use.

METHODS: The imaging performance of the camera system was assessed quantitatively and qualitatively at different source depths, source to collimator distances (SCD), activity levels, acquisition times and source separations, utilising bespoke phantoms.

RESULTS: The system sensitivity and spatial resolution of the HGC for 125I were 0.41 cps/MBq (at SCD 48 mm) and 1.53 ± 0.23 mm (at SCD 10 mm) respectively. The camera was able to detect the 125I seed at a SCD of 63 mm (with no scattering material in place) in images recorded within a 1-min acquisition time. The detection of the seeds beneath scattering material (simulating deep-seated tumours) was limited to depths of less than 20 mm beneath the skin surface with a SCD of 63 mm and seed activity of 2.43 MBq. Subjective assessments of the hybrid images acquired showed the capability of the HGC for localising the 125I seeds.

CONCLUSION: This preliminary ex vivo study demonstrates that the HGC is capable of detecting 125I seeds and could be a useful tool in radioactive seed localisation with the added benefit of providing hybrid optical γ images for guiding breast conserving surgery.

ADVANCES IN KNOWLEDGE: The SFOV HGC could provide high resolution fused optical-gamma images of 125I radioactive seeds indicating the potential use in intraoperative surgical procedure such as RSL.

来源:Br J Radiol,查看原文。版权归原作者所有。

引用

  1. Ng AH, Alqahtani MS, Jambi LK, et al. Assessing a small field of view hybrid gamma camera for perioperative iodine-125 seed localisation. Br J Radiol. 2019. PMID: 30864832. doi:10.1259/bjr.20190020

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。