文献速递泌尿生殖MR

前列腺癌扩散加权MRI中复数平均高b值图像的图像质量与诊断准确性

Image quality and diagnostic accuracy of complex-averaged high b value images in diffusion-weighted MRI of prostate cancer.
PubMed 文献 · Abdom Radiol (NY) · 2019-01-01

回顾性纳入84例3T多参数前列腺MRI(无直肠内线圈)患者,比较常规幅度平均与复数平均对采集和计算高b值图像质量及诊断准确性的影响。复数平均提高图像质量和置信度且不影响ADC值;计算高b值图像与采集高b值图像相当,可替代后者以缩短扫描时间。

要点

  • 复数平均可提高采集和计算高b值图像的图像质量(P<0.0001),且不显著影响病灶ADC值(P=0.08)。
  • 计算高b值图像质量与采集高b值图像无显著差异,联合复数平均与计算高b值诊断准确性最高但未达统计学显著(79.55% vs 72.73%,P=0.317)。
  • 采集b=2000 s/mm²序列平均耗时6分30秒(±1分16秒),约占全协议28分31秒(±4分26秒)的25%。

核心结论

  • 复数平均可提供更好的图像质量和置信度,对显著前列腺病灶检测的ADC值和诊断准确性无显著影响;计算高b值图像与采集高b值图像相当,可替代后者以帮助减少成像时间。

研究设计与样本

  • 回顾性研究,纳入84例在3特斯拉下接受多参数前列腺MRI(无直肠内线圈)的患者;采集b=50、900、2000 s/mm²,经后处理生成常规幅度平均和复数平均的 trace 加权图像,并衍生计算高b值(b=2000 s/mm²)图像和ADC图;由两名放射科医生在不同阅片会话中评估图像质量和PIRADS,通过PIRADS与Gleason评分相关性评估诊断准确性。

诊断标准与关键数值

  • 复数平均与幅度平均相比,病灶ADC值无显著差异(P=0.08);复数平均改善采集高b值图像质量(P<0.0001)和计算高b值图像质量(P<0.0001);复数平均改善两名阅片者对采集高b值的置信度(P<0.0001,P<0.05),仅改善阅片者A对计算高b值的置信度(P<0.0001);计算高b值图像质量与采集高b值图像无显著差异;联合复数平均与计算高b值检测显著前列腺病灶的诊断准确性为79.55%,幅度平均采集高b值为72.73%(P=0.317,未达统计学显著)。

临床应用与阅片要点

  • 复数平均可用于改善采集和计算高b值图像的图像质量与置信度;计算高b值图像与采集高b值图像相当并可替代后者,有助于减少成像时间。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To evaluate the impact of complex-averaging on image quality (IQ) and diagnostic accuracy of acquired and calculated high b value (aHBV, cHBV) images in diffusion-weighted prostate MRI.

MATERIALS AND METHODS: This retrospective study included 84 patients who underwent multiparametric prostate MRI at 3 Tesla without endorectal coil. DWIs were acquired at three different b values which included two lower b values (b = 50,900 s/mm2) and one higher b value (aHBV at 2000 s/mm2). The acquired data were postprocessed to generate two different types of trace-weighted images-using conventional magnitude-averaging and complex-averaging. Using lower b values (b = 50,900 s/mm2) from both conventional and complex-averaged image sets, cHBV images (b = 2000 s/mm2) and ADC maps were derived. All image sets were reviewed by two radiologists in different reading sessions to assess image quality and PIRADS. The diagnostic accuracy of different image sets for the detection of prostate lesions was performed by correlating PIRADS and Gleason scores.

RESULTS: Complex-averaging did not impact ADC values of the prostate lesions compared to magnitude-averaging (P = 0.08). Complex-averaging improved image quality of acquired high b value and calculated high b value images (P < 0.0001). Complex-averaging also improved the level of confidence (LOC) of the acquired high b value for both readers (P < 0.0001, P < 0.05), but only for reader A in calculated high b value (P < 0.0001). The image quality of calculated high b value images was not significantly different than acquired high b value images. The dataset combining complex-averaging and calculated high b value provided the highest diagnostic accuracy (but not statistically significant) for detection of the significant prostate lesion compared to the magnitude-averaged acquired high b value (79.55% vs. 72.73%; P = 0.317). The mean acquisition time for b = 2000 s/mm2 sequence (aHBV) was 6 min 30 s (± 1 min 16 s) out of a total of 28 min 31 s (± 4 min 26 s) for the entire mp-MRI protocol (approximately 25% of total scan time).

CONCLUSION: Complex-averaging provides better image quality and level of confidence without significant impact on ADC values and diagnostic accuracy for detection of the significant prostate lesions . The calculated high b value images are also comparable to (and can substitute) the acquired high b value images which can help in reducing the imaging time.

来源:Abdom Radiol (NY),查看原文。版权归原作者所有。