文献速递腹部儿科影像MR

儿童和青年肝脏多回波定量Dixon成像中导航门控与屏气采集技术的比较

Comparison of navigator-gated and breath-held image acquisition techniques for multi-echo quantitative dixon imaging of the liver in children and young adults.
PubMed 文献 · Abdom Radiol (NY) · 2019-01-01

回顾性纳入42例2017年1月至2018年7月间行肝脏MRI的儿童和青年,比较屏气与导航门控多回波Dixon序列的PDFF和R2*值。PDFF几乎完全一致(rc=0.997,平均偏差0.3%),R2*相关强但一致性差(rc=0.832);导航门控图像临床限制性呼吸运动频率显著更高(88% vs 48%,p=0.0001)。

要点

  • 导航门控与屏气mDixon的PDFF几乎完全一致(rc=0.997,平均偏差0.3%),可替代屏气序列定量肝脏脂肪。
  • R2*值相关强但一致性差(rc=0.832),两序列R2*不可直接互换。
  • 导航门控图像临床限制性呼吸运动频率显著更高(88% vs 48%,p=0.0001)。

核心结论

  • 尽管呼吸运动伪影更多,自由呼吸导航门控mDixon序列产生的PDFF值与屏气序列几乎完全一致。

研究设计与样本

  • 回顾性研究,经机构审查委员会批准并豁免知情同意。
  • 纳入2017年1月至2018年7月间行肝脏MRI屏气和导航门控mDixon序列的患者42例(女15例,男27例;平均年龄15.7±4.6岁)。
  • 一名对序列不知情的阅片者测量每个序列四幅图像的PDFF和R2*,另一名不知情阅片者以5分Likert量表评估呼吸运动。
  • 采用Pearson相关(r)、Lin一致性系数(rc)和Bland-Altman分析评估技术间一致性;临床限制性运动(评分≥3)频率采用Fisher精确检验比较。

诊断标准与关键数值

  • PDFF在序列间几乎完全一致:rc=0.997(95%CI 0.994-0.998),平均偏差0.3%,95%一致性界限-2.4%至+1.7%。
  • R2*值相关很强但一致性差:r=0.837,rc=0.832(95%CI 0.716-0.910)。
  • 屏气序列平均PDFF 16.6±13.1%、R2* 29.3±4.7 s-1;导航门控序列平均PDFF 17.0±13.2%、R2* 29.6±5.2 s-1。
  • 临床限制性呼吸运动(评分≥3)频率导航门控88% vs 屏气48%,p=0.0001。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: Acquired over a breath hold, multi-echo Dixon (mDixon) magnetic resonance imaging (MRI) of the liver can be used to quantify proton density fat fraction (PDFF) and iron-related signal decay. However, young, obese, and co-morbid patients may have limited breath holding capacity and could benefit from a motion-robust mDixon acquisition. The purpose of this study was to compare hepatic PDFF and R2* values between navigator-gated and breath-held mDixon MRI acquisition techniques in children and young adults with suspected liver disease.

MATERIALS AND METHODS: This retrospective study was institutional review board-approved with a waiver of informed consent. Patients who underwent liver MRI with breath-held and navigator-gated mDixon sequences between January 2017 and July 2018 were included. One reviewer, blinded to sequence, measured PDFF and R2* on four images from each sequence. Another blinded reviewer graded respiratory motion (5-point Likert scale). Pearson correlation (r), Lin's concordance coefficients (rc), and Bland-Altman analyses were used to assess agreement between techniques. Frequency of clinically limiting motion (score ≥ 3) was compared with Fisher's exact test.

RESULTS: Forty-two patients were included (15 female, 27 male; mean age: 15.7 ± 4.6 years). Mean PDFF and R2* were 16.6 ± 13.1% and 29.3 ± 4.7 s-1 (breath-held) versus 17.0 ± 13.2% and 29.6 ± 5.2 s-1 (navigator-gated). PDFF agreed almost perfectly between sequences (rc = 0.997, 95% CI 0.994-0.998; mean bias: 0.3%; 95% limits of agreement: - 2.4 to +1.7%), while R2* values correlated very strongly but with poor agreement (r = 0.837, rc = 0.832, 95% CI 0.716-0.910). Navigator-gated images exhibited significantly higher frequency of clinically limiting respiratory motion (88% vs. 48%, p = 0.0001).

CONCLUSION: Despite greater respiratory motion artifact, a free-breathing navigator-gated mDixon sequence produces PDFF values with almost perfect agreement to a breath-held sequence.

来源:Abdom Radiol (NY),查看原文。版权归原作者所有。

引用

  1. Gilligan LA, Dillman JR, Tkach JA, et al. Comparison of navigator-gated and breath-held image acquisition techniques for multi-echo quantitative dixon imaging of the liver in children and young adults. Abdom Radiol (NY). 2019. PMID: 30815713. doi:10.1007/s00261-019-01960-1

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。