能量与入路方式对肝外肿瘤播散及消融区的影响:基于包膜下肿瘤模型的离体实验
在牛肝块包膜下创建42个肿瘤模拟物,比较低功率与高功率射频消融及平行或垂直电极入路对肝外肿瘤播散和消融区的影响。平行入路时低功率组包膜造影剂渗漏1/10(10%),高功率组9/10(90%)(p<0.001);垂直入路两组各6/11(54.5%)。低功率组首次roll-off和爆裂声显著延迟(p=0.002),两组消融区大小无统计学差异(p=0.247)。
要点
- 平行入路时低功率组包膜造影剂渗漏1/10(10%),高功率组9/10(90%)(p<0.001)。
- 垂直入路时两组包膜造影剂渗漏均为6/11(54.5%)。
- 低功率组首次roll-off和爆裂声显著延迟(p=0.002),与入路方式无关。
核心结论
- 低功率射频消融联合平行入路在包膜下实性肝肿瘤射频消融中可能有效且安全,可减少肝外肿瘤播散;电极垂直放置于包膜时无论功率高低均存在肝外肿瘤播散风险。
研究设计与样本
- 离体实验:在牛肝块包膜下注射碘对比剂混合物创建42个肿瘤模拟物,行射频消融,电极平行或垂直于肝表面穿过肿瘤模拟物,采用低功率和高功率方案(分别为1组和2组),消融前后行CT扫描。
关键影像表现/征象
- 平行入路时,消融后CT显示1组包膜造影剂渗漏1/10(10%),2组9/10(90%)(p<0.001)。
- 垂直入路时,两组各发现6/11(54.5%)包膜造影剂渗漏。
- 1组首次roll-off和爆裂声显著延迟,与入路方式无关(p=0.002)。
- 两组肝标本消融区大小无统计学差异(p=0.247)。
临床应用与阅片要点
- 作者提出,包膜下实性肝肿瘤射频消融时低功率联合平行入路可能有效且安全,可减少肝外肿瘤播散。
- 作者指出,电极垂直于包膜放置时无论功率高低均存在肝外肿瘤播散风险。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: To evaluate the impact of energy and access methods on extrahepatic tumor spreading and the ablation zone in an ex vivo subcapsular tumor mimic model with a risk of extrahepatic tumor spreading.
MATERIALS AND METHODS: Forty-two tumor-mimics were created in bovine liver blocks by injecting a mixture of iodine contrast material just below the liver capsule. Radiofrequency (RF) ablations were performed using an electrode placed parallel or perpendicular to hepatic surface through the tumor mimic with low- and high-power protocols (groups 1 and 2, respectively). Computed tomography (CT) scans were performed before and after ablation. The presence of contrast leak on the hepatic surface on CT, size of ablation zone, and timing of the first roll-off and popping sound were compared between the groups.
RESULTS: With parallel access, one contrast leak in group 1 (1/10, 10%) and nine in group 2 (9/10, 90%) (p < 0.001) were identified on post-ablation CT. With perpendicular access, six contrast leaks were identified in each group (6/11, 54.5%). The first roll-off and popping sound were significantly delayed in group 1 irrespective of the access method (p = 0.002). No statistical difference in the size of the ablation zone of the liver specimen was observed between the two groups (p = 0.247).
CONCLUSION: Low-power RF ablation with parallel access is proposed to be effective and safe from extrahepatic tumor spreading in RF ablation of a solid hepatic tumor in the subcapsular location. Perpendicular placement of an electrode to the capsule is associated with a risk of extrahepatic tumor spreading regardless of the power applied.
引用
- Kim JS, Ko Y, Kwon H, et al. Impact of Energy and Access Methods on Extrahepatic Tumor Spreading and the Ablation Zone: An Ex vivo Experiment Using a Subcapsular Tumor Model. Korean J Radiol. 2019. PMID: 30887740. doi:10.3348/kjr.2018.0564
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。