利用下腔静脉超声无创估测右心房压力

Non-invasive Estimation of Right Atrial Pressure Using Inferior Vena Cava Echography.
PubMed 文献 · Ultrasound Med Biol · 2019-02-25

研究纳入49例置管患者,采用半自动方法追踪下腔静脉运动并计算整个静脉段的平均腔静脉指数,结合人体测量学信息建立估测右心房压力的模型;模型估测误差约3.6±2.6 mmHg,仅用操作者手工测量变量时误差约4.0±2.5 mmHg,作者认为需在更大数据集上进一步研究。

要点

  • 半自动追踪下腔静脉运动并平均全段腔静脉指数,可提供更稳定的搏动性指标。
  • 结合下腔静脉指数与人体测量学信息估测右心房压力,误差约3.6±2.6 mmHg。
  • 仅用操作者手工测量变量时估测误差约4.0±2.5 mmHg。

核心结论

  • 半自动下腔静脉追踪方法结合人体测量学信息可估测右心房压力,误差约3.6±2.6 mmHg,作者认为需在更大数据集上进一步研究该方法。

研究设计与样本

  • 研究纳入49例置管患者,以导管记录的右心房压力为参考,采用留一法交叉验证从特征组合中选出估测误差最小的模型。

诊断标准与关键数值

  • 半自动方法模型估测右心房压力的误差约3.6±2.6 mmHg(均值±标准差)。
  • 仅用操作者手工测量变量、不使用软件时估测误差约4.0±2.5 mmHg(均值±标准差)。

局限性

  • 原文指出需在更大数据集上进一步研究该右心房压力估测方法。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

The pulsatility of the inferior vena cava (IVC) reflects the volume status and central venous pressure of patients. The standard clinical indicator of IVC pulsatility is the caval index (CI), measured from ultrasound recordings. However, its estimation is not standardized and is vulnerable to artifacts, mostly because of IVC movements during respiration. Thus, we used a (recently patented) semi-automated method that tracks IVC movements and averages the CI across an entire section of the vein, which provides a more stable indication of pulsatility. This algorithm was used to estimate the CI, pulsatility indicators reflecting either respiratory or cardiac stimulation and the mean diameter of the IVC. These IVC indices, together with anthropometric information, were used as potential features to build an innovative model for the estimation of the right atrial pressure (RAP) recorded from 49 catheterized patients. An exhaustive search was carried out for the best among all possible models that could be obtained by using combinations of these features. The model with minimum estimation error (tested with a leave-one-out approach) was selected. This model estimated RAP with an error of about 3.6 ± 2.6 mm Hg (mean ± standard deviation); the error when using only operator measured variables, without software, was about 4.0 ± 2.5 mm Hg. These promising results underline the need for further study of our RAP estimation method on a larger data set.

来源:Ultrasound Med Biol,查看原文。版权归原作者所有。

引用

  1. Mesin L, Albani S, Sinagra G, et al. Non-invasive Estimation of Right Atrial Pressure Using Inferior Vena Cava Echography. Ultrasound Med Biol. 2019. PMID: 30819412. doi:10.1016/j.ultrasmedbio.2018.12.013