双层层探测器CT在临床常规扫描中的CT协议剂量:初步报告

Dose of CT protocols acquired in clinical routine using a dual-layer detector CT scanner: A preliminary report.
PubMed 文献 · Eur J Radiol · 2019-01-14

本回顾性研究收集六个月期间双层探测器CT(DLCT)连续检查4536份剂量报告,与单层探测器CT(SLCT)连续检查5783份剂量报告比较。所有DLCT检查均以120 kVp采集以支持双能分析,SLCT中79%以120 kVp、21%以100/80 kVp采集。两种设备均使用超过50次的15个适应症协议中,7个协议两者剂量无显著差异(p>0.05),7个协议SLCT剂量高于DLCT(p<0.03),1个协议DLCT剂量显著高于SLCT(p<0.005)。结论认为DLCT可在不增加辐射剂量的情况下获得广谱临床适应症的双能信息。

要点

  • DLCT所有检查均以120 kVp采集,可在常规扫描中实现双能分析。
  • 15个常用协议中,7个协议两者剂量无显著差异,7个协议SLCT剂量高于DLCT,1个协议DLCT剂量更高。
  • 与SLCT相比,DLCT在广谱临床适应症中获取双能信息未增加辐射剂量。

核心结论

  • 双层探测器CT可在不增加辐射剂量的情况下,于广泛临床适应症中获取双能信息。

研究设计与样本

  • 回顾性收集六个月期间双层探测器CT(DLCT;IQon spectral CT,Philips Healthcare)连续检查的剂量长度乘积数据,与单层探测器CT(SLCT;Brilliance iCT,Philips Healthcare)连续检查比较;数据收集前各系统成像协议均已针对诊断图像质量优化;纳入在两种系统上均至少使用50次的CT协议;排除统计离群值后评估报告的剂量水平;共收集DLCT剂量报告4536份、SLCT剂量报告5783份。

诊断标准与关键数值

  • DLCT所有检查均以120 kVp采集,可行双能分析;SLCT检查中79%以120 kVp采集、21%以100/80 kVp采集。
  • 15个适应症协议在两种扫描仪上均使用超过50次。
  • 7个协议两种扫描仪间无显著差异(p>0.05);7个协议SLCT剂量高于DLCT(p<0.03);1个协议DLCT剂量显著高于SLCT(p<0.005)。

临床应用与阅片要点

  • 常规临床使用双层探测器CT时,可在不增加辐射剂量的前提下获得双能信息,但个别协议DLCT剂量可能显著高于SLCT(p<0.005),需按协议评估剂量。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To assess the radiation dose associated with always-on dual-energy acquisitions in clinical practice over a broad range of clinical protocols using a dual-layer detector CT (DLCT; IQon spectral CT, Philips Healthcare) as compared to an otherwise technically equivalent single-layer detector CT (SLCT; Brilliance iCT, Philips healthcare).

MATERIALS AND METHODS: Dose-length-product data for consecutive examinations over a six-month period acquired with DLCT were retrospectively collected and compared to consecutive examinations from an SLCT. Imaging protocols were optimized for diagnostic image quality for each system prior to data collection. Dose reports of CT protocols that were used at least 50 times on both systems were collected. After exclusion of statistical outliers, protocols were evaluated with regard to reported dose levels.

RESULTS: In total, 4536 dose reports for DLCT and 5783 reports for SLCT were collected. All DLCT examinations were acquired at 120 kVp, enabling dual-energy analysis. With SLCT, 79% of examinations were acquired at 120 kVp, and 21% at 100/80 kVp. Protocols for 15 indications were used more than 50 times on both scanners. For seven protocols there was no significant difference between the two scanners (p > 0.05), whereas seven protocols were acquired with higher dose levels on SLCT compared to the DLCT (p < 0.03). For one protocol, the DLCT dose was significantly higher (p < 0.005) compared to the SLCT.

CONCLUSION: Dual-layer detector CT enables acquisition of dual-energy information over a broad range of clinical indications without increasing radiation dose when compared to a conventional single-layer detector CT.