人颈动脉的实时容积评估:手持式多光谱光声断层成像
研究用手持容积多光谱光声断层成像(MSOT)对16名健康志愿者颈动脉分叉行无创实时三维成像,并与临床B超对照,结果显示该技术可实时容积显示颈动脉分叉,对氧合血红蛋白高吸收波长可获得更高对比噪声比,且较常规临床成像方法更不易受运动伪影影响。
要点
- 手持容积MSOT可在单帧容积图像中实时三维显示颈动脉分叉。
- 氧合血红蛋白高吸收波长可获得更高的对比噪声比。
- 该技术较常规临床成像方法更不易受运动伪影影响。
核心结论
- 手持容积多光谱光声断层成像可无创实时显示人颈动脉,较常规临床成像方法更不易受运动伪影影响,有望为颈动脉病变的无创无标记评估提供额外影像生物标志物。
研究设计与样本
- 研究纳入16名健康志愿者,使用自制手持容积MSOT扫描仪行高空间分辨率(约200 µm)、实时(10容积/秒)三维成像,并通过快速调谐激发光波长提供光谱能力,同时以临床B超扫描进行解剖交叉验证。
关键影像表现/征象
- 容积MSOT在单帧容积图像中同时实时三维显示整个颈动脉分叉区域。
- 对氧合血红蛋白高光学吸收对应的波长可获得更高的对比噪声比。
临床应用与阅片要点
- 作者认为该成像方法较常规临床成像方法更不易受运动伪影影响,有望为颈动脉病变的无创无标记评估提供额外影像生物标志物。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Background Multispectral optical imaging has the capability of resolving hemoglobin, lipid, and water. Volumetric multispectral optoacoustic tomography (MSOT) is a hybrid imaging technique that provides a unique combination of functional and molecular contrast with real-time handheld imaging. Purpose To investigate whether volumetric MSOT can provide real-time assessment of the anatomic and functional status of the human carotid artery bifurcation noninvasively. Materials and Methods Imaging of healthy volunteers (n = 16) was performed with a custom-designed handheld volumetric MSOT scanner capable of high-spatial-resolution (approximately 200 µm) and real-time (10 volumes/sec) three-dimensional imaging, while further providing spectroscopic capacity through fast tuning of the excitation light wavelength. For comparison and anatomic cross-validation, volunteers were also scanned with clinical B-mode US. Results Volumetric MSOT achieved real-time imaging and characterization of the entire carotid bifurcation area across three dimensions simultaneously captured in a single volumetric image frame. Analysis of the acquired data further showed that a higher contrast-to-noise ratio can be achieved for wavelengths corresponding to a high optical absorption of oxygenated hemoglobin. Conclusion The human carotid artery was visualized by using handheld volumetric multispectral optoacoustic tomography. This imaging approach is less prone to motion artifacts than are the conventional clinical imaging methods, holding promise for providing additional image-based biomarkers for noninvasive label-free assessment of carotid artery disease. © RSNA, 2019 Online supplemental material is available for this article. See also the editorial by Mezrich in this issue.
引用
- Ivankovic I, Merčep E, Schmedt CG, et al. Real-time Volumetric Assessment of the Human Carotid Artery: Handheld Multispectral Optoacoustic Tomography. Radiology. 2019. PMID: 30747592. doi:10.1148/radiol.2019181325
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。