文献速递骨肌CT

在尸体研究中优化肩关节MDCT关节造影的辐射剂量参数:基本概念阐释

Optimizing radiation dose parameters in MDCT arthrography of the shoulder: illustration of basic concepts in a cadaveric study.
PubMed 文献 · Skeletal Radiol · 2019-02-06

研究采用3具尸体6个肩关节,用256层MDCT在关节腔内注射稀释碘对比剂后扫描,设置6个递减辐射剂量水平(CTDIvol 20、15、10、8、6、4 mGy)和4个递减管电位(140、120、100、80 kVp)。结果显示,CTDIvol高于10 mGy且管电位120 kVp以上时,所有方案的主观噪声和诊断图像质量与20 mGy-140 kVp参考方案等效;10 mGy联合120 kVp可获得诊断图像质量,较标准剂量方案降低达50%,较文献报道方案降低高达500%。

要点

  • 肩关节MDCT关节造影在CTDIvol 10 mGy、管电位120 kVp时可获得诊断图像质量。
  • CTDIvol高于10 mGy且管电位120 kVp以上时,主观噪声与参考方案等效(所有p≥0.22)。
  • 15 mGy以上剂量联合100 kVp以上管电位时,CNR与20 mGy-140 kVp参考方案等效(所有p≥0.054)。

核心结论

  • 肩关节MDCT关节造影在辐射剂量10 mGy、最佳管电位120 kVp时可获得诊断图像质量,较标准剂量方案降低达50%,较文献报道方案降低高达500%。

研究设计与样本

  • 尸体研究,采用3具人类尸体的6个肩关节,使用256层MDCT系统,关节腔内注射稀释碘对比剂后扫描,设置6个递减辐射剂量水平(CTDIvol 20、15、10、8、6、4 mGy)和每个剂量水平下4个递减管电位(140、120、100、80 kVp),测量图像噪声和对比噪声比(CNR),由两名独立盲法观察者按四点量表评估整体诊断图像质量、主观噪声量和伪影严重程度。

诊断标准与关键数值

  • CTDIvol 15 mGy以上联合管电位100 kVp以上时,CNR与参考方案20 mGy-140 kVp等效(所有p≥0.054)。
  • CTDIvol高于10 mGy且管电位120 kVp以上时,所有方案诊断图像质量和主观噪声与参考方案20 mGy-140 kVp等效(所有p≥0.22)。
  • CTDIvol 10 mGy联合管电位120 kVp可获得诊断图像质量。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To determine in a cadaveric study the lowest achievable radiation dose and optimal tube potential generating diagnostic image quality in multidetector computed tomography (MDCT) arthrography of the shoulder.

MATERIALS AND METHODS: Six shoulders from three human cadavers were scanned using a 256-MDCT system after intra-articular injection of diluted iodinated contrast material. Using six decreasing radiation dose levels (CTDIvol: 20, 15, 10, 8, 6, and 4 mGy) and for each dose level, four decreasing tube potentials (140, 120, 100, and 80 kVp), image noise and contrast-to-noise ratio (CNR) were measured. Two independent and blinded observers assessed the overall diagnostic image quality, subjective amount of noise, and severity of artifacts according to a four-point scale. Influence of those MDCT data acquisition parameters on objective and subjective image quality was analyzed using the Kruskal-Wallis and Wilcoxon signed-rank tests, and pairwise comparisons were performed.

RESULTS: Multidetector CT protocols with radiation doses of 15 mGy or higher, combined with tube potentials of 100 kVp or higher, were equivalent in CNR to the reference 20 mGy-140 kVp protocol (all p ≥ 0.054). Above a CTDIvol of 10 mGy and a tube potential of 120 kVp, all protocols generated diagnostic image quality and subjective noise equivalent to the 20 mGy-140 kVp protocol (all p ≥ 0.22).

CONCLUSIONS: Diagnostic image quality in MDCT arthrography of the shoulder can be obtained with a radiation dose of 10 mGy at an optimal tube potential of 120 kVp, corresponding to a reduction of up to 50% compared with standard-dose protocols, and as high as 500% compared with reported protocols in the literature.

来源:Skeletal Radiol,查看原文。版权归原作者所有。

引用

  1. Aguet J, Becce F, Dunet V, et al. Optimizing radiation dose parameters in MDCT arthrography of the shoulder: illustration of basic concepts in a cadaveric study. Skeletal Radiol. 2019. PMID: 30725159. doi:10.1007/s00256-019-3150-6

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。