光谱整形技术用于颈椎CT:能否解决下颈椎区域伪影问题?

Cervical Spine CT Using Spectral Shaping: Can It Be a Solution to Overcome Artifacts in the Lower Cervical Spinal Region?
PubMed 文献 · Korean J Radiol · 2019-01-01

研究比较了锡滤过140 kVp光谱整形颈椎CT(58例)与常规120 kVp CT(49例)的图像质量和辐射剂量,并在75例同时接受CT和MRI的患者中评估椎间盘形态的模态间一致性。结果显示Sn140-kVp扫描主观图像质量更优,C7/T1水平噪声更低、衰减减低更少,有效剂量降低50%(1.0 ± 0.1 mSv vs. 2.0 ± 0.4 mSv,p < 0.001),下颈椎区域模态间一致性倾向于更好。作者认为该协议有助于克服下颈椎CT图像伪影。

要点

  • Sn140-kVp光谱整形颈椎CT较常规120 kVp主观图像质量更优(p < 0.001)。
  • C7/T1水平噪声更低且衰减减低更少,下颈椎伪影改善。
  • 有效剂量降低50%(1.0 ± 0.1 mSv vs. 2.0 ± 0.4 mSv)。

核心结论

  • Sn140-kVp颈椎CT在不增加辐射剂量的情况下改善下颈椎区域图像质量,该协议有助于克服下颈椎CT图像伪影。

研究设计与样本

  • 比较性研究,Sn140-kVp组58例、常规120 kVp组49例;75例同时接受颈椎MRI和CT者用于评估椎间盘形态的模态间一致性;定性图像质量采用5点Likert量表,定量测量C3/4、C6/7、C7/T1水平中央椎管内噪声和衰减。

关键影像表现/征象

  • C7/T1水平Sn140-kVp扫描噪声显著低于常规扫描,衰减减低显著更少(p < 0.001)。
  • 下颈椎区域椎间盘形态的CT与MRI模态间一致性在Sn140-kVp采集倾向于优于常规采集。

诊断标准与关键数值

  • 定性图像质量采用5点Likert量表分析,Sn140-kVp扫描显著优于常规扫描(p < 0.001)。
  • 定量图像质量通过测量C3/4、C6/7、C7/T1水平中央椎管内噪声和衰减评估。
  • 模态间一致性在C3/4、C5/6、C6/7、C7/T1水平评估椎间盘形态。

临床应用与阅片要点

  • Sn140-kVp光谱整形协议可在不增加辐射剂量的情况下改善下颈椎区域图像质量,有助于克服该区域伪影。
  • 下颈椎区域椎间盘形态评估时,Sn140-kVp采集的CT与MRI模态间一致性倾向于更好。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To investigate the image quality, radiation dose, and intermodality agreement of cervical spine CT using spectral shaping at 140 kVp by a tin filter (Sn140-kVp) in comparison with those of conventional CT at 120 kVp.

MATERIALS AND METHODS: Patients who had undergone cervical spine CT with Sn140-kVp (n = 58) and conventional 120 kVp (n = 49) were included. Qualitative image quality was analyzed using a 5-point Likert scale. Quantitative image quality was assessed by measuring the noise and attenuation within the central spinal canals at C3/4, C6/7, and C7/T1 levels. Radiation doses received by patients were estimated. The intermodality agreement for disc morphology between CT and MRI was assessed at C3/4, C5/6, C6/7, and C7/T1 levels in 75 patients who had undergone cervical spine MRI as well as CT.

RESULTS: Qualitative image quality was significantly superior in Sn140-kVp scans than in the conventional scans (p < 0.001). At C7/T1 level, the noise was significantly lower and the decrease in attenuation was significantly less in Sn140-kVp scans, than in the conventional scans (p < 0.001). Radiation doses were significantly reduced in Sn140-kVp scans by 50% (effective dose: 1.0 ± 0.1 mSv vs. 2.0 ± 0.4 mSv; p < 0.001). Intermodality agreement in the lower cervical spine region tended to be better in Sn140-kVp acquisitions than in the conventional acquisitions.

CONCLUSION: Cervical spine CT using Sn140-kVp improves image quality of the lower cervical region without increasing the radiation dose. Thus, this protocol can be helpful to overcome the artifacts in the lower cervical spine CT images.