一种适用于MR神经成像与介入性平板CT融合的拟人脊柱体模的开发
本研究旨在设计一种脊柱体模,用于将MR神经成像(MRN)与介入性平板CT(FPCT)图像融合。体模由琼脂糖凝胶和人工神经构成,并包含与骨CT衰减相等的材料。通过体内MR映射确定目标组织弛豫时间,体外系列稀释确定组织等效成分;用碳酸钙和硫酸钡调节椎体及骶骨放射密度;琼脂糖浸渍聚乙烯纤维模拟神经。获取脂肪抑制T2加权3D SPACE STIR序列用于MRN,并与体模的FPCT扫描融合。结果:脂肪组织体内T1/T2值为394±16 ms和161±16 ms,对应50%蓖麻油浓度;骨髓等效值为822±21 ms和67±6 ms,用40%蓖麻油模拟;人工骨用30% CaCO3和1.5% BaSO4达到皮质骨样放射密度1'115±80 HU;模拟神经在MRN中成功显示并与FPCT融合,实现神经和骨的最佳对比。结论:定制体模在MRN和FPCT中均显示与体内条件相似的组织对比,有助于模拟融合图像引导的脊柱介入。
要点
- 脂肪组织T1/T2为394±16/161±16 ms,对应50%蓖麻油;骨髓等效为822±21/67±6 ms,用40%蓖麻油。
- 人工骨用30% CaCO3和1.5% BaSO4达到皮质骨样放射密度1'115±80 HU。
- 模拟神经在MRN中成功显示并与FPCT融合,神经和骨对比最佳。
核心结论
- 定制体模在MRN和FPCT中均显示与体内条件相似的组织对比,有助于模拟融合图像引导的脊柱介入。
研究设计与样本
- 研究为体模开发研究;体内MR映射测定目标组织弛豫时间(n=5);体外系列稀释确定组织等效成分;体模由琼脂糖凝胶、人工神经及骨等效材料按拟人几何组装;获取脂肪抑制T2加权3D SPACE STIR序列用于MRN,并与体模的FPCT扫描融合。
关键影像表现/征象
- 脂肪组织体内T1/T2值为394±16 ms和161±16 ms,对应50%蓖麻油浓度。
- 骨髓等效T1/T2值为822±21 ms和67±6 ms,用40%蓖麻油模拟。
- 人工骨用30% CaCO3和1.5% BaSO4达到皮质骨样放射密度1'115±80 HU。
- 模拟神经在MRN中成功显示,并与FPCT融合,神经和骨对比最佳。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To design a spine phantom suitable for fusion of MR neurography (MRN) with interventional flat panel computed tomography (FPCT) images from tissue-equivalent agarose gels and artificial nerves in MRI, including material with equal attenuation to bone in computed tomography (CT).
METHODS: T1-/T2-relaxation times of target tissue were determined in vivo (n = 5) using MR mapping-techniques. Serial dilution of castor oil lipogels was performed ex vivo in order to define correct composition for tissue-equivalent relaxation times. Similarly, serial dilution series of calcium carbonate (CaCO3) and barium sulphate (BaSO4) in synthetic resin were used to adjust radiodensity of selected vertebral bodies (L1-L5) and sacrum in CT. Nerve tissue was simulated with agarose-impregnated polyethylene fibers. Spine phantom was assembled using respective components in anthropomorphic geometry. A fat-saturated, T2-weighted 3D SPACE STIR sequence was acquired for MRN and subsequently fused with an on-site FPCT scan of the phantom.
RESULTS: In vivo T1-/T2-values for fat tissue were found to be at 394 ± 16 ms and 161 ± 16 ms, corresponding to a castor oil concentration of 50%. Analogously, bone marrow-equivalent values were measured at 822 ± 21 ms and 67 ± 6 ms, simulated with 40% castor oil. Cortical bone-like radiodensity of 1'115 ± 80 HU was achieved for artificial bone with 30% CaCO3 and 1.5% BaSO4. Simulated nerves were successfully depicted in MRN and fused with FPCT, combining optimal contrasts for nerves and bones on-site.
CONCLUSIONS: The customized phantom showed analogous tissue contrasts to in vivo conditions in both MRN and FPCT, facilitating simulations of fusion-image guided spine interventions.
引用
- Kobe A, Zadory M, Hamie QM, et al. Development of an anthropomorphic spine phantom suitable for fusion of MR neurography with interventional flat-panel CT. Eur J Radiol. 2019. PMID: 30777205. doi:10.1016/j.ejrad.2019.01.019