文献速递骨肌技术与安全CT

利用双层光谱CT定位像进行DXA等效骨密度定量

DXA-equivalent quantification of bone mineral density using dual-layer spectral CT scout scans.
PubMed 文献 · Eur Radiol · 2019-02-13

研究建立基于双层光谱CT定位像的面积骨密度(aBMD)测量方法,在欧标脊柱体模上与DXA比较,标准定位像条件下aBMD与DXA显著相关(r=0.9925,p<0.001),不同管电流下aBMD无显著差异(p≥0.20),并在女性患者中评估可行性,数据与参考年龄匹配人群BMD一致。

要点

  • 双层光谱CT定位像经后处理可半自动计算脊柱aBMD,与DXA显著相关(r=0.9925)。
  • 不同管电流(10至200 mA)下aBMD无显著差异(p≥0.20),方法可适用于不同定位像设置。
  • 该方法处理现有定位像数据,不增加额外辐射,可能实现机会性骨质疏松筛查。

核心结论

  • 基于双层光谱CT定位像的aBMD测量快速可靠,在体模中与DXA测量显著相关,可能实现真正机会性骨质疏松筛查。

研究设计与样本

  • 方法学研究,建立基于一对2D虚拟单能定位像的后处理算法;在欧标脊柱体模上以标准定位像管电流30 mA及其他管电流(10至200 mA)重复测量评估,并使用脂肪等效延伸环模拟不同患者体型,与双能X线吸收测定法(DXA)比较;并在女性患者中评估体内可行性。

诊断标准与关键数值

  • 体模标准定位像所得aBMD与DXA测量显著相关(r=0.9925,p<0.001);平均准确度DXA为4.12%、定位像法为1.60%,精密度DXA为2.64%、定位像法为2.03%;不同管电流下aBMD无显著差异(所有p≥0.20);患者数据与参考年龄匹配人群BMD值一致。

临床应用与阅片要点

  • 该方法处理CT检查中本就采集的定位像数据,不增加辐射暴露,测量耗时小于25秒/患者,可在阅片时一并获得脊柱aBMD的DXA等效指标。

局限性

  • 与DXA的相关性仅在体模中得到验证,体内可行性仅在女性患者样本中评估。
  • 患者数据仅与参考年龄匹配人群BMD值比较,未与DXA直接对照。
  • 方法依赖手动软组织分割与半自动椎体分割。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To develop and evaluate a method for areal bone mineral density (aBMD) measurement based on dual-layer spectral CT scout scans.

METHODS: A post-processing algorithm using a pair of 2D virtual mono-energetic scout images (VMSIs) was established in order to semi-automatically compute the aBMD at the spine similarly to DXA, using manual soft tissue segmentation, semi-automatic segmentation for the vertebrae, and automatic segmentation for the background. The method was assessed based on repetitive measurements of the standardized European Spine Phantom (ESP) using the standard scout scan tube current (30 mA) and other tube currents (10 to 200 mA), as well as using fat-equivalent extension rings simulating different patient habitus, and was compared to dual-energy X-ray absorptiometry (DXA). Moreover, the feasibility of the method was assessed in vivo in female patients.

RESULTS: Derived from standard scout scans, aBMD values measured with the proposed method significantly correlated with DXA measurements (r = 0.9925, p < 0.001), and mean accuracy (DXA, 4.12%; scout, 1.60%) and precision (DXA, 2.64%; scout, 2.03%) were comparable between the two methods. Moreover, aBMD values assessed at different tube currents did not differ significantly (p ≥ 0.20 for all), suggesting that the presented method could be applied to scout scans with different settings. Finally, data derived from sample patients were concordant with BMD values from a reference age-matched population.

CONCLUSIONS: Based on dual-layer spectral scout scans, aBMD measurements were fast and reliable and significantly correlated with the according DXA measurements in phantoms. Considering the number of CT acquisitions performed worldwide, this method could allow truly opportunistic osteoporosis screening.

KEY POINTS: • 2D scout scans (localizer radiographs) from a dual-layer spectral CT scanner, which are mandatory parts of a CT examination, can be used to automatically determine areal bone mineral density (aBMD) at the spine. • The presented method allowed fast (< 25 s/patient), semi-automatic, and reliable DXA-equivalent aBMD measurements for state-of-the-art DXA phantoms at different tube settings and for various patient habitus, as well as for sample patients. • Considering the number of CT scout scan acquisitions performed worldwide on a daily basis, the presented technique could enable truly opportunistic osteoporosis screening with DXA-equivalent metrics, without involving higher radiation exposure since it only processes existing data that is acquired during each CT scan.

来源:Eur Radiol,查看原文。版权归原作者所有。

引用

  1. Laugerette A, Schwaiger BJ, Brown K, et al. DXA-equivalent quantification of bone mineral density using dual-layer spectral CT scout scans. Eur Radiol. 2019. PMID: 30758656. doi:10.1007/s00330-019-6005-6

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。