高加速三维T1加权双回波Dixon肝胆期成像采用压缩感知-敏感度编码:与标准T1加权序列的图像质量和实性病灶可检测性比较
回顾性纳入217例钆塞酸增强3T肝脏MRI患者,比较CS-SENSE mDixon-GRE与标准mDixon-GRE肝胆期图像质量和实性局灶性肝脏病灶检测能力。结果显示两者整体图像质量无显著差异,CS-SENSE mDixon-GRE图像噪声更高但运动伪影更少(均p<0.05),病灶检测JAFROC figure-of-merit分别为0.918和0.953(p=0.142),非劣效性成立,且采集时间缩短。
要点
- CS-SENSE mDixon-GRE肝胆期图像整体质量与标准mDixon-GRE无显著差异,采集时间缩短。
- CS-SENSE mDixon-GRE图像噪声更高但运动伪影更少(均p<0.05)。
- 实性局灶性肝脏病灶检测JAFROC figure-of-merit:标准mDixon-GRE 0.918 vs CS-SENSE mDixon-GRE 0.953(p=0.142),非劣效性成立。
核心结论
- CS-SENSE mDixon-GRE肝胆期序列的整体图像质量和实性局灶性肝脏病灶可检测性非劣于标准mDixon-GRE序列,且采集时间缩短。
研究设计与样本
- 回顾性研究,共217例患者(初步研究54例,主研究163例)接受3T钆塞酸增强肝脏MRI;主研究中117例患者共193个实性结节纳入病灶检测分析;比较标准mDixon-3D-GRE(SENSE,加速因子2.8)与CS-SENSE mDixon-3D-GRE;由两名腹部放射科医师共识评估图像质量,另三名腹部放射科医师独立评估诊断性能,采用JAFROC比较。
关键影像表现/征象
- CS-SENSE mDixon-GRE较标准mDixon-GRE图像噪声更高(p<0.05)。
- CS-SENSE mDixon-GRE较标准mDixon-GRE运动伪影更少(p<0.05)。
诊断标准与关键数值
- 病灶检测读者平均JAFROC figure-of-merit:标准mDixon-GRE为0.918,CS-SENSE mDixon-GRE为0.953(p=0.142)。
- CS-SENSE mDixon-GRE相对标准mDixon-GRE的非劣效性成立,差异为0.064(95%CI:-0.012至0.081)。
临床应用与阅片要点
- 在3T钆塞酸增强肝脏MRI肝胆期,可采用CS-SENSE mDixon-GRE序列替代标准mDixon-GRE序列,以缩短采集时间而不降低实性病灶检出。
- 需注意CS-SENSE mDixon-GRE图像噪声更高但运动伪影更少,阅片时应综合评估。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: To compare a high acceleration three-dimensional (3D) T1-weighted gradient-recalled-echo (GRE) sequence using the combined compressed sensing (CS)-sensitivity encoding (SENSE) method with a conventional 3D GRE sequence using SENSE, with respect to image quality and detectability of solid focal liver lesions (FLLs) in the hepatobiliary phase (HBP) of gadoxetic acid-enhanced liver MRI.
MATERIALS AND METHODS: A total of 217 patients with gadoxetic acid-enhanced liver MRI at 3T (54 in the preliminary study and 163 in the main study) were retrospectively included. In the main study, HBP imaging was done twice using the standard mDixon-3D-GRE technique with SENSE (acceleration factor [AF]: 2.8, standard mDixon-GRE) and the high acceleration mDixon-3D GRE technique using the combined CS-SENSE technique (CS-SENSE mDixon-GRE). Two abdominal radiologists assessed the two MRI data sets for image quality in consensus. Three other abdominal radiologists independently assessed the diagnostic performance of each data set and its ability to detect solid FLLs in 117 patients with 193 solid nodules and compared them using jackknife alternative free-response receiver operating characteristics (JAFROC).
RESULTS: There was no significant difference in the overall image quality. CS-SENSE mDixon-GRE showed higher image noise, but lesser motion artifact levels compared with the standard mDixon-GRE (all p < 0.05). In terms of lesion detection, reader-averaged figures-of-merit estimated with JAFROC was 0.918 for standard mDixon-GRE, and 0.953 for CS-SENSE mDixon-GRE (p = 0.142). The non-inferiority of CS-SENSE mDixon-GRE over standard mDixon-GRE was confirmed (difference: 0.064 [-0.012, 0.081]).
CONCLUSION: The CS-SENSE mDixon-GRE HBP sequence provided comparable overall image quality and non-inferior solid FFL detectability compared with the standard mDixon-GRE sequence, with reduced acquisition time.
引用
- Nam JG, Lee JM, Lee SM, et al. High Acceleration Three-Dimensional T1-Weighted Dual Echo Dixon Hepatobiliary Phase Imaging Using Compressed Sensing-Sensitivity Encoding: Comparison of Image Quality and Solid Lesion Detectability with the Standard T1-Weighted Sequence. Korean J Radiol. 2019. PMID: 30799575. doi:10.3348/kjr.2018.0310
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。