文献速递腹部CT

体重自适应超低剂量胰腺灌注CT:辐射剂量、图像质量与灌注参数

Weight-adapted ultra-low-dose pancreatic perfusion CT: radiation dose, image quality, and perfusion parameters.
PubMed 文献 · Abdom Radiol (NY) · 2019-01-01

前瞻性纳入100例患者,按体重分为五组采用不同管电压与管电流组合行胰腺灌注CT,记录辐射剂量并主观及客观评价图像质量、测量灌注参数。结果辐射剂量在50 kg以下患者为8.7 mSv、80 kg以上为18.9 mSv,主观图像质量评分均值高于4.45分(5分制),A-D组客观图像质量等效(P>0.05)而E组更优(P<0.05),五组间正常胰腺实质灌注参数(BF、BV、PMB)无显著差异。结论认为体重自适应超低剂量胰腺灌注CT可有效降低辐射剂量而不损害图像质量,正常实质灌注参数准确可靠。

要点

  • 体重自适应超低剂量胰腺灌注CT在50 kg以下患者辐射剂量为8.7 mSv,80 kg以上为18.9 mSv。
  • 主观图像质量评分均值高于4.45分(5分制),两名放射科医师间一致性良好。
  • 五组间正常胰腺实质BF、BV、PMB无显著差异,提示灌注参数不受剂量方案影响。

核心结论

  • 体重自适应超低剂量胰腺灌注CT可有效降低辐射剂量而不损害图像质量,正常胰腺实质灌注参数准确可靠。

研究设计与样本

  • 前瞻性研究纳入100例患者(男47例、女53例),按体重分为A、B、C、D、E五组,各组采用不同管电压与管电流组合行胰腺灌注CT;记录CTDI与DLP,主观及客观评价图像质量(噪声、信噪比、对比噪声比),测量BF、BV、PMB;五组间采用单因素方差分析比较。

诊断标准与关键数值

  • 辐射剂量:50 kg以下患者达8.7 mSv,80 kg以上患者为18.9 mSv;主观图像质量评分均值高于4.45分(5分制);A-D组客观图像质量等效(P>0.05),E组更优(P<0.05);五组间正常胰腺实质BF、BV、PMB无显著差异。

临床应用与阅片要点

  • 按体重选择管电压与管电流组合可在降低辐射剂量的同时维持胰腺灌注CT的图像质量与正常实质灌注参数测量。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: We evaluate the reliability and feasibility of weight-adapted ultra-low-dose pancreatic perfusion CT.

METHODS: A total of 100 (47 men, 53 women) patients were enrolled prospectively and were assigned to five groups (A, B, C, D, and E) with different combination of tube voltage and tube current according to their body weight. Radiation dose parameters including volume CT dose index (CTDI) and dose-length product (DLP) were recorded. Image quality was evaluated both subjectively and objectively (noise, signal-to-noise ratio, contrast-to-noise ratio). Perfusion parameters including blood flow (BF), blood volume (BV), and permeability (PMB) were measured. The dose, image quality measurements, and perfusion parameters were compared between the five groups using one-way analysis of variance (ANOVA).

RESULTS: Radiation dose reached 8.7 mSv in patients under 50 kg and was 18.9 mSv in patients above 80 kg. The mean subjective image quality score was above 4.45 on a 5-point scale with good agreement between two radiologists. Groups A-D had equivalent performance on objective image quality (P > 0.05), while Group E performed even better (P < 0.05). No significant differences emerged in comparison with perfusion parameters (BF, BV, PMB) of normal pancreas parenchyma between the five groups.

CONCLUSION: Weight-adapted ultra-low-dose pancreatic perfusion CT can effectively reduce radiation dose without prejudice to image quality, and the perfusion parameters of normal parenchyma are accurate and reliable.

来源:Abdom Radiol (NY),查看原文。版权归原作者所有。