低剂量四维心肌CT灌注中噪声降低与运动消除:ASTRA4D算法的初步临床评估

Noise reduction and motion elimination in low-dose 4D myocardial computed tomography perfusion (CTP): preliminary clinical evaluation of the ASTRA4D algorithm.
PubMed 文献 · Eur Radiol · 2019-02-04

前瞻性纳入30例疑似或确诊冠心病患者行动态增强320排CTP,评估联合运动消除与时空降噪的ASTRA4D算法。与PCA基准相比,ASTRA4D显著降低时间噪声(-83%,178.5 vs 29.9)和空间噪声(-34%,21.4 vs 14.1),提高SNR(+47%,3.6 vs 5.3)及主观图像质量;在15例以MR-MPI为参考标准中,ASTRA4D每节段敏感度91%(58/64)显著优于PCA的52%(33/64)(p=0.011),特异度相似(96%,429/446 vs 98%,435/446)。

要点

  • ASTRA4D联合局部时间回归与形变图像配准,30例患者配准均成功。
  • 与PCA基准相比,ASTRA4D时间噪声降低83%、空间噪声降低34%,SNR提高47%。
  • 以MR-MPI为参考,ASTRA4D每节段敏感度91%(58/64)显著高于PCA的52%(33/64)(p=0.011)。

核心结论

  • ASTRA4D配准算法改善了时空噪声分布和CTP序列图像质量,使四维CTP检测心肌缺血的敏感度显著提高。

研究设计与样本

  • 前瞻性纳入30例疑似或确诊冠状动脉疾病患者,行动态增强320排CTP;以MR-MPI为参考标准在15例患者中评估诊断准确性。

关键影像表现/征象

  • 与PCA基准相比,ASTRA4D时间噪声降低83%(178.5 vs 29.9),空间噪声降低34%(21.4 vs 14.1)。
  • 与PCA基准相比,ASTRA4D的SNR提高47%(3.6 vs 5.3)。
  • ASTRA4D主观图像质量评分较PCA提高:运动+1.0、对比度+1.0、轮廓锐利度+0.5(1为差,5为优)。

诊断标准与关键数值

  • ASTRA4D每节段敏感度91%(58/64),PCA为52%(33/64),p=0.011。
  • ASTRA4D每节段特异度96%(429/446),PCA为98%(435/446),两者相似。
  • 参考标准为MR-MPI,评估对象为15例患者的每节段灌注缺损检测。

临床应用与阅片要点

  • ASTRA4D可改善低剂量四维CTP序列的时空噪声分布与主观图像质量,有助于提高灌注缺损检出敏感度。
  • 与PCA相比,ASTRA4D在保持特异度相似的同时显著提高敏感度,阅片时需注意算法对图像噪声和运动伪影的改善效果。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To propose and evaluate a four-dimensional (4D) algorithm for joint motion elimination and spatiotemporal noise reduction in low-dose dynamic myocardial computed tomography perfusion (CTP).

METHODS: Thirty patients with suspected or confirmed coronary artery disease were prospectively included and underwent dynamic contrast-enhanced 320-row CTP. A novel deformable image registration method based on the principal component analysis (PCA) of the ante hoc temporally smoothed voxel-wise time-attenuation curves (ASTRA4D) is presented. Quantitative (standard deviation, signal-to-noise ratio (SNR), temporal variation, volumetric deformation) and qualitative (motion, contrast, contour sharpness [1, poor; 5, excellent]) measures of CTP quality were assessed for the original and motion-compensated sequences (without and with temporal filtering, PCA/ASTRA4D). Following myocardial perfusion deficit detection by two readers, diagnostic accuracy was evaluated using magnetic resonance myocardial perfusion imaging (MR-MPI) as the reference standard in 15 patients.

RESULTS: Registration using ASTRA4D was successful in all 30 patients and resulted in comparison with the benchmark PCA in significantly (p < 0.001) reduced noise over time (- 83%, 178.5 vs 29.9) and spatially (- 34%, 21.4 vs 14.1) as well as improved SNR (+ 47%, 3.6 vs 5.3) and subjective image quality (motion, contrast, contour sharpness [+ 1.0, + 1.0, + 0.5]). ASTRA4D had significantly improved per-segment sensitivity of 91% (58/64) and similar specificity of 96% (429/446) compared with PCA (52%, 33/64; 98%, 435/446; p = 0.011) in the visual detection of perfusion deficits.

CONCLUSIONS: The ASTRA4D registration algorithm improved the spatiotemporal noise profile and CTP sequence image quality, resulting in significantly improved sensitivity of 4D CTP in the detection of myocardial ischemia.

KEY POINTS: • ASTRA4D combines local temporal regression and deformable image registration. • Quantitative and qualitative measures of CTP quality are improved compared to PCA. • Improved spatiotemporal differentiation of ischemic regions leads to an excellent perfusion deficit concordance of ASTRA4D with MRI.

来源:Eur Radiol,查看原文。版权归原作者所有。

引用

  1. Lukas S, Feger S, Rief M, et al. Noise reduction and motion elimination in low-dose 4D myocardial computed tomography perfusion (CTP): preliminary clinical evaluation of the ASTRA4D algorithm. Eur Radiol. 2019. PMID: 30715584. doi:10.1007/s00330-018-5899-8

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。