柔性12通道丝网印刷儿科MRI线圈的评估
本研究设计了一种儿科尺寸的12通道丝网印刷柔性MRI线圈,在体模中评估单元耦合与信噪比,并对20名儿童(2天至12岁)进行扫描,与3名年龄性别匹配的历史对照(商用32通道心脏阵列)比较。结果显示印刷线圈信噪比略低但差异无统计学意义(P=0.26),诊断质量评分4.5±0.6,受试者及照护者偏好评分4.1±0.6。
要点
- 柔性丝网印刷儿科线圈在体模中噪声协方差矩阵与商用线圈相似,信噪比仅轻微下降。
- 20例儿童扫描显示图像信噪比与商用线圈差异无统计学意义(P=0.26),诊断质量评分4.5±0.6。
- 受试者及照护者偏好印刷线圈,平均评分4.1±0.6(1偏好商用,5偏好印刷)。
核心结论
- 柔性丝网印刷儿科MRI接收线圈在体模和儿童受试者中产生足够的信噪比,图像质量与常规线圈相似,但受试者及其照护者更偏好该印刷线圈。
研究设计与样本
- 通过丝网印刷工艺设计儿科尺寸12通道线圈阵列,评估单元耦合与体模信噪比;2017年9月至11月扫描20名受试者(年龄2天至12岁,男11例女9例),与3名年龄性别匹配的历史对照(商用32通道心脏阵列,3T)比较;照护者接受度通过护士、技师、麻醉医师及受试者或家长评分;图像诊断质量采用Likert量表(5=高图像质量,1=不可诊断);评估并比较图像信噪比。
诊断标准与关键数值
- 诊断质量采用Likert量表评分,5分为高图像质量,1分为不可诊断,本研究诊断质量评分为4.5±0.6。
- 照护者接受度评分量表:1分偏好商用线圈,5分偏好印刷线圈,平均得分4.1±0.6。
- 图像信噪比均值:印刷线圈椎旁肌54±49、腹壁肌78±51、腰大肌59±35;商用线圈分别为64±55、65±48、57±43;差异无统计学意义(P=0.26)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Background Screen-printed MRI coil technology may reduce the need for bulky and heavy housing of coil electronics and may provide a better fit to patient anatomy to improve coil performance. Purpose To assess the performance and caregiver and clinician acceptance of a pediatric-sized screen-printed flexible MRI coil array as compared with conventional coil technology. Materials and Methods A pediatric-sized 12-channel coil array was designed by using a screen-printing process. Element coupling and phantom signal-to-noise ratio (SNR) were assessed. Subjects were scanned by using the pediatric printed array between September and November 2017; results were compared with three age- and sex-matched historical control subjects by using a commercial 32-channel cardiac array at 3 T. Caregiver acceptance was assessed by asking nurses, technologists, anesthesiologists, and subjects or parents to rate their coil preference. Diagnostic quality of the images was evaluated by using a Likert scale (5 = high image quality, 1 = nondiagnostic). Image SNR was evaluated and compared. Results Twenty study participants were evaluated with the screen-printed coil (age range, 2 days to 12 years; 11 male and nine female subjects). Loaded pediatric phantom testing yielded similar noise covariance matrices and only slightly degraded SNR for the printed coil as compared with the commercial coil. The caregiver acceptance survey yielded a mean score of 4.1 ± 0.6 (scale: 1, preferred the commercial coil; 5, preferred the printed coil). Diagnostic quality score was 4.5 ± 0.6. Mean image SNR was 54 ± 49 (paraspinal muscle), 78 ± 51 (abdominal wall muscle), and 59 ± 35 (psoas) for the printed coil, as compared with 64 ± 55, 65 ± 48, and 57 ± 43, respectively, for the commercial coil; these SNR differences were not statistically significant (P = .26). Conclusion A flexible screen-printed pediatric MRI receive coil yields adequate signal-to-noise ratio in phantoms and pediatric study participants, with similar image quality but higher preference by subjects and their caregivers when compared with a conventional MRI coil. © RSNA, 2019 Online supplemental material is available for this article. See also the editorial by Lamb in this issue.
引用
- Winkler SA, Corea J, Lechêne B, et al. Evaluation of a Flexible 12-Channel Screen-printed Pediatric MRI Coil. Radiology. 2019. PMID: 30806599. doi:10.1148/radiol.2019181883
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。