基于模型的迭代重建算法在降低对比剂浓度的儿童胸部CT中改善图像质量
研究纳入56例接受低剂量增强胸部CT的儿童(中位年龄4岁)作为研究组,与56例对照组比较。研究组采用100 kVp、270 mgI/ml对比剂及MBIR重建,对照组采用120 kV、320 mgI/ml对比剂及ASIR重建。结果显示两组辐射剂量无差异,研究组对比剂剂量降低19.7%,MBIR较对照组ASIR图像噪声降低58.6%,SNR和CNR分别提高143.6%和165.7%。
要点
- MBIR可使儿童低剂量胸部CT在对比剂剂量降低19.7%时图像噪声降低58.6%。
- 与对照组ASIR图像相比,MBIR使SNR和CNR分别提高143.6%和165.7%。
- 研究组与对照组辐射剂量无差异(CTDIvol 1.79±1.45 mGy vs 1.68±0.92 mGy,p=0.65)。
核心结论
- MBIR在对比剂剂量降低19.3%的情况下改善了儿童低辐射剂量胸部CT的图像质量。
研究设计与样本
- 研究纳入56例接受低剂量增强胸部CT的儿童(中位年龄4岁)作为研究组,并与56例对照组儿童比较;研究组使用100 kVp和270 mgI/ml对比剂,图像采用50% ASIR和MBIR重建,对照组使用120 kV和320 mgI/ml对比剂,图像仅采用ASIR重建。
关键影像表现/征象
- 研究组与对照组辐射剂量无差异(CTDIvol 1.79±1.45 mGy vs 1.68±0.92 mGy,p=0.65)。
- 研究组对比剂剂量较对照组降低19.7%(5.84±2.69 gI vs 7.27±3.80 gI)。
- MBIR较对照组ASIR图像噪声降低58.6%,SNR和CNR分别提高143.6%和165.7%。
- 两组ASIR图像质量相似。
- 所有图像的增强程度均满足诊断要求。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: To evaluate model-based iterative reconstruction (MBIR) in improving the image quality of chest CT in children with reduced concentration contrast medium (CM).
METHODS: Fifty-six children (median age of 4 years) who received low-dose enhanced chest CT were enrolled as the study group and compared with the control group of 56 children. Both groups used the automatic tube current modulation to achieve age-based noise index values of 11-15 HU. The study group used 100 kVp and reduced CM concentration of 270 mgI/ml, and the images in this group were reconstructed with 50% adaptive statistical iterative reconstruction (ASIR) and MBIR. The control group used 120 kV and standard CM of 320 mgI/ml, and the images in this group were reconstructed with ASIR only. Subjective image quality and objective image quality of the three image sets were evaluated. The subjective quality included overall image noise, enhancement degree, lesion (including mediastinum mass, pulmonary space-occupying lesions, and parenchymal infiltrative lesions) conspicuity, and beam-hardening artifacts. The objective quality included the measurement of noise in the left ventricle and back muscle to calculate signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) of ventricle.
RESULTS: There was no difference in radiation dose between the study (CTDIvol of 1.79 ± 1.45 mGy) and control (1.68 ± 0.92 mGy) groups (p = 0.65). However, the study group used 19.7% lower CM dose than the control group (5.84 ± 2.69 vs. 7.27 ± 3.80 gI), and the enhancement in all images met the diagnostic requirements. MBIR reduced image noise by 58.6% and increased SNR and CNR by 143.6% and 165.7%, respectively, compared to ASIR images in the control group. The two ASIR image sets had similar image quality.
CONCLUSION: MBIR improved the image quality of low-radiation-dose chest CT in children at 19.3% reduced CM dose.