文献速递胸部DR

胸部平板探测器系统的最佳射线质:真实体模研究

Optimal beam quality for chest flat panel detector system: realistic phantom study.
PubMed 文献 · Eur Radiol · 2019-02-08

研究用模拟肺结节和正常结构的真实肺体模,在入射皮肤剂量固定为0.1 mGy条件下比较90–140 kV管电压加或不加铜滤过时的胸部平板探测器影像质量;三位胸部放射科医师独立评分显示,90 kV加铜滤过的影像在椎体、肺血管及与膈肌和心脏重叠的结节可见性上优于传统120 kV无滤过影像,其余模拟肺结节和蜂窝状磨玻璃影的可见性无显著差异。

要点

  • 在入射皮肤剂量固定为0.1 mGy时,90 kV加铜滤过的影像在椎体、肺血管及与膈肌和心脏重叠结节的可见性上优于传统120 kV无滤过影像。
  • 除与膈肌和心脏重叠的结节及蜂窝状磨玻璃影外,其余模拟肺结节在各管电压加或不加铜滤过与120 kV无滤过间的可见性无显著差异。
  • 作者认为90 kV加铜滤过可能适用于胸部平板探测器摄影,并认为有必要开展胸部平板探测器射线质优化的临床研究。

核心结论

  • 在入射皮肤剂量相同(0.1 mGy)条件下,90 kV加铜滤过的平板探测器影像质量优于传统120 kV无滤过影像。

研究设计与样本

  • 采用真实肺体模模拟肺结节和正常结构,行胸部平板探测器摄影,管电压90–140 kV加或不加铜滤过,入射皮肤剂量固定为0.1 mGy,以管电压120 kV无滤过的传统平板探测器影像为标准;三位胸部放射科医师在不知曝光设置的情况下独立用5级评分(+2至-2)评价影像质量,采用Wilcoxon秩和检验并Bonferroni校正进行统计比较。

关键影像表现/征象

  • 90 kV加铜滤过影像在椎体、肺血管及与膈肌和心脏重叠的结节可见性上优于传统120 kV无滤过影像(p < 0.05)。
  • 其余模拟肺结节(与膈肌和心脏重叠的结节及蜂窝状磨玻璃影除外)在各管电压加或不加铜滤过与120 kV无滤过间的可见性无显著差异。

诊断标准与关键数值

  • 影像质量采用5级评分:+2为明显优于标准,+1为略优于标准,0与标准相等,-1为略差于标准,-2为明显差于标准。
  • 以管电压120 kV无铜滤过的传统平板探测器影像为标准。
  • 统计采用Wilcoxon秩和检验并Bonferroni校正。

临床应用与阅片要点

  • 在入射皮肤剂量固定为0.1 mGy时,90 kV加铜滤过可能适用于胸部平板探测器摄影。
  • 作者认为有必要开展胸部平板探测器射线质优化的临床研究。

局限性

  • 作者认为有必要开展胸部平板探测器射线质优化的临床研究(原文结论中提出需进一步临床研究)。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To investigate optimal beam quality for chest flat panel detector (FPD) system by semi-quantitatively assessment using a realistic lung phantom.

MATERIALS AND METHODS: Chest FPD radiographs were obtained on a realistic lung phantom with simulated lung opacities using various X-ray tube voltage levels (90-140 kV) with/without copper filter. Entrance skin dose was set to maintain identical for all images (0.1 mGy). Three chest radiologists unaware of the exposure settings independently evaluated the image quality of each simulated opacity and normal structure using a 5-point scale (+ 2: clearly superior to the standard; + 1: slightly superior to the standard; 0: equal to the standard; - 1: slightly inferior to the standard; - 2: clearly inferior to the standard). The traditional FPD image obtained at a tube voltage of 120 kV was used as the standard. The scores of image quality were statistically compared using the Wilcoxon rank test with Bonferroni correction.

RESULTS: FPD images using 90-kV shot with copper filter were superior to the traditional 120-kV shot without filter with respect to the visibility of vertebra, pulmonary vessels, and nodules overlapping diaphragm and heart (p < 0.05). There was no significant difference with respect to the visibility of all other simulated lung opacities (lung nodules except for overlying diaphragm/heart and honeycomb opacity) between each tube voltage level with/without copper filter and the traditional 120-kV shot without filter.

CONCLUSION: Image quality of FPD images using 90 kV with copper filtration is superior to that using standard tube voltage when dose is identical.

KEY POINTS: • FPD image quality using 90 kV with filter is superior to that using traditional beam. • Ninety-kilovolt shot with copper filter may be suitable for chest FPD image. • Clinical study dealing with chest FPD beam optimization would be warranted.

来源:Eur Radiol,查看原文。版权归原作者所有。

引用

  1. Kuwahara C, Aoki T, Oda N, et al. Optimal beam quality for chest flat panel detector system: realistic phantom study. Eur Radiol. 2019. PMID: 30737566. doi:10.1007/s00330-019-5998-1

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。