3T场强下儿童脑部2D BLADE涡轮梯度自旋回波与2D自旋回波回波平面扩散加权MRI的比较:初步经验
在53例儿童(10.4±7.9岁)中比较3T下2D TGSE BLADE与常规SE-EPI扩散加权脑成像,评估伪影与诊断质量。43.4%病例首选TGSE BLADE,49.1%两者相当;诊断质量评分SE-EPI 2.2±0.8 vs TGSE 1.2±0.4(p<0.05)。16例41个ROI的表观扩散系数值两组统计相似,线性回归斜率0.92(r²=0.97)。
要点
- TGSE BLADE在脑部几何变形和高磁敏感区信号堆积方面少于常规SE-EPI
- 诊断质量评分TGSE BLADE优于SE-EPI(1.2±0.4 vs 2.2±0.8,p<0.05)
- 两组表观扩散系数值统计相似,线性回归斜率0.92(r²=0.97)
核心结论
- TGSE BLADE扩散加权成像在脑部几何变形和高磁敏感区信号堆积方面少于常规SE-EPI。
研究设计与样本
- 在53例儿科患者(10.4±7.9岁)中于3T采集扩散加权数据,均于增强前采集,先采集SE-EPI;采用四点量表评分(1最佳,4最差),由一名神经放射学家对两种序列评分并记录首选序列。
关键影像表现/征象
- TGSE BLADE在脑部几何变形方面少于常规SE-EPI。
- TGSE BLADE在高磁敏感区信号堆积方面少于常规SE-EPI。
- 92.5%的病例两种序列运动伪影均极少。
- TGSE BLADE评分中无病例因显著伪影致图像质量勉强可接受而评为4分。
诊断标准与关键数值
- 诊断质量平均评分:SE-EPI 2.2±0.8 vs TGSE BLADE 1.2±0.4(p<0.05)。
- 43.4%的病例首选TGSE BLADE,49.1%的病例两种序列相当。
- 16例患者41个感兴趣区(包括尾状核、壳核、苍白球、丘脑和病变区)的表观扩散系数值两组统计相似,线性回归斜率0.92(r²=0.97)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
RATIONALE AND OBJECTIVES: We describe our preliminary experience using a 2D turbo gradient- and spin-echo (TGSE) diffusion-weighted (DW) pulse sequence with non-Cartesian BLADE trajectory at 3 T in pediatric patients. We compared the TGSE BLADE to conventional DW spin-echo echo-planar imaging (SE-EPI) in pediatric brain imaging, assessing the presence of artifacts from signal pile-ups, geometric distortion, motion, susceptibility from air-tissue interface, shunts and orthodontia, and diagnostic image quality.
MATERIALS AND METHODS: Data were acquired in 53 patients (10.4 ± 7.9 years). All DW imaging data were acquired precontrast, with SE-EPI first. A four-point scale for rating was used-1 (best) and 4 (worst). A neuroradiologist scored the two sequences and further noted whether the TGSE BLADE approach or SE-EPI was preferred in each case. Apparent diffusion coefficients were compared quantitatively between the two sequences in a subset of 16 patients, in 41 separate regions of interests including caudate nucleus, putamen, globus pallidus, thalamus, and pathological areas.
RESULTS: In 43.4% of the cases, TGSE BLADE was preferred; in 49.1% of the cases, both sequences were preferred equally. Average scores for SE-EPI were 2.2 ± 0.8 versus TGSE's 1.2 ± 0.4 in assessing diagnostic quality (p < 0.05). Motion artifacts were minimal on both sequences in 92.5% of the cases. In the TGSE BLADE scores, no case received a "4" for significant artifacts with marginally acceptable image quality. Apparent diffusion coefficients values between the two sequences were statistically similar, with a linear regression slope of 0.92 (r2 = 0.97).
CONCLUSION: TGSE BLADE DW imaging exhibited less geometric distortion in the brain and reduced signal pile-ups in areas of high susceptibility than conventional SE-EPI.
引用
- Hu HH, McAllister AS, Jin N, et al. Comparison of 2D BLADE Turbo Gradient- and Spin-Echo and 2D Spin-Echo Echo-Planar Diffusion-Weighted Brain MRI at 3 T: Preliminary Experience in Children. Acad Radiol. 2019. PMID: 30777649. doi:10.1016/j.acra.2019.02.002
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。