压缩感知高分辨率磁共振成像用于颅内及颅外动脉:与传统并行成像的比较
在14名健康志愿者中比较压缩感知联合SENSE(CS)与传统SENSE用于颅内及颅外动脉高分辨率磁共振成像。CS在T1WI加速因子5.5、6.8、9.7及PD加速因子5.8时图像质量和血管显示优于SENSE;CS T1WI加速因子5.5及CS PD加速因子3.2和4.0与原始图像无显著差异。CS信噪比和对比噪声比高于SENSE但低于原始图像。作者认为CS优于SENSE,可能是可靠的血管高分辨率磁共振加速方法。
要点
- CS在T1WI加速因子5.5、6.8、9.7及PD加速因子5.8时图像质量和血管显示优于SENSE(p<0.05)。
- CS T1WI加速因子5.5和CS PD加速因子3.2、4.0与原始图像质量无显著差异(p>0.05)。
- CS可接受成像比例高于SENSE(CS T1WI加速因子6.8和PD加速因子5.8;p<0.0167)。
核心结论
- CS优于SENSE,可能是使用T1WI加速因子5.5和PD加速因子3.2、4.0进行血管高分辨率磁共振成像的可靠加速方法。
研究设计与样本
- 在14名健康志愿者中,使用3特斯拉系统进行高分辨率磁共振成像,采集三维T1加权成像和质子密度加权成像,分别采用CS或SENSE在相同总加速因子下(T1WI为5.5、6.8、9.7;PD为3.2、4.0、5.8)并与原始序列(SENSE T1WI加速因子3.5;SENSE PD加速因子2.0)比较。
诊断标准与关键数值
- 两名神经放射科医生测量信噪比和对比噪声比,并使用视觉评分系统评估图像质量;可接受成像定义为视觉评分≥2分。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: To compare conventional sensitivity encoding (SENSE) to compressed sensing plus SENSE (CS) for high-resolution magnetic resonance imaging (HR-MRI) of intracranial and extracranial arteries.
MATERIALS AND METHODS: HR-MRI was performed in 14 healthy volunteers. Three-dimensional T1-weighted imaging (T1WI) and proton density-weighted imaging (PD) were acquired using CS or SENSE under the same total acceleration factors (AFt)-5.5, 6.8, and 9.7 for T1WI and 3.2, 4.0, and 5.8 for PD-to achieve reduced scanning times in comparison with the original imaging sequence (SENSE T1WI, AFt 3.5; SENSE PD, AFt 2.0) using the 3-tesla system. Two neuroradiologists measured signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR), and used visual scoring systems to assess image quality. Acceptable imaging was defined as a visual score ≥ 2. Repeated measures analysis of variance and Cochran's Q test were performed.
RESULTS: CS yielded better image quality and vessel delineation than SENSE in T1WI with AFt of 5.5, 6.8, and 9.7, and in PD with AFt of 5.8 (p < 0.05). CS T1WI with AFt of 5.5 and CS PD with AFt of 3.2 and 4.0 did not differ significantly from original imaging (p > 0.05). SNR and CNR in CS were higher than they were in SENSE, but lower than they were in the original images (p < 0.05). CS yielded higher proportions of acceptable imaging than SENSE (CS T1WI with AFt of 6.8 and PD with AFt of 5.8; p < 0.0167).
CONCLUSION: CS is superior to SENSE, and may be a reliable acceleration method for vessel HR-MRI using AFt of 5.5 for T1WI, and 3.2 and 4.0 for PD.
引用
- Suh CH, Jung SC, Lee HB, et al. High-Resolution Magnetic Resonance Imaging Using Compressed Sensing for Intracranial and Extracranial Arteries: Comparison with Conventional Parallel Imaging. Korean J Radiol. 2019. PMID: 30799580. doi:10.3348/kjr.2018.0424
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。