儿童计算机断层扫描中的自动管电流调制与管电压选择:关于辐射剂量与图像质量的体模研究

Automatic Tube Current Modulation and Tube Voltage Selection in Pediatric Computed Tomography: A Phantom Study on Radiation Dose and Image Quality.
PubMed 文献 · Invest Radiol · 2019-01-01

该体模研究使用新生儿、1岁、5岁和10岁儿童拟人模型,评估自动管电流调制(ATCM)和自动管电压选择(ATVS)对儿童头部和躯干CT辐射剂量与图像质量的影响。结果显示ATCM使头部CTDIvol降低8%–24%、胸部16%–39%、腹部/盆腔25%–41%,并维持图像噪声;ATVS在CT血管成像中进一步降低剂量(头部达70%、胸部77%、腹部/盆腔34%),在平扫中头部剂量增加达21%、胸部降低达34%、腹部/盆腔降低达48%,同时维持对比噪声比。

要点

  • ATCM使儿童头部CTDIvol降低8%–24%、胸部16%–39%、腹部/盆腔25%–41%,并维持图像噪声。
  • 在CT血管成像中,ATVS使头部剂量降低达70%、胸部达77%、腹部/盆腔达34%。
  • 在平扫中,ATVS使头部剂量增加达21%,胸部降低达34%、腹部/盆腔降低达48%。

核心结论

  • 在儿童CT中,使用ATCM可显著降低辐射剂量并维持图像噪声。
  • 额外使用ATVS可进一步降低胸部及腹部/盆腔的辐射剂量,并针对特定临床诊断任务维持对比噪声比。

研究设计与样本

  • 体模研究,使用4个代表新生儿、1岁、5岁和10岁儿童平均个体的拟人物理体模。
  • 对头部、胸部及腹部/盆腔进行标准采集,分别采用固定管电流、ATCM和ATVS三种模式,并在不同辐射剂量水平下采集以生成不同质量水平的图像。
  • 测定参考容积CT剂量指数(CTDIvol)、参考图像噪声和参考对比噪声比,评估ATCM与ATVS的潜在剂量降低。

关键影像表现/征象

  • ATCM使头部CTDIvol降低8%–24%、胸部降低16%–39%、腹部/盆腔降低25%–41%,并维持图像噪声。
  • 在CT血管成像中,ATVS使头部剂量降低达70%、胸部达77%、腹部/盆腔达34%。
  • 在平扫中,ATVS使头部剂量增加达21%,胸部降低达34%、腹部/盆腔降低达48%。

诊断标准与关键数值

  • 评估指标包括参考CTDIvol、参考图像噪声和参考对比噪声比,用以比较固定管电流、ATCM和ATVS三种模式。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: The aim of this study was to investigate the effects of a modern automatic tube current modulation (ATCM) and automatic tube voltage selection (ATVS) system on radiation dose and image quality in pediatric head, and torso computed tomography (CT) examinations for various clinical indications.

MATERIALS AND METHODS: Four physical anthropomorphic phantoms that represent the average individual as neonate, 1-year-old, 5-year-old, and 10-year-old child were used. Standard head, thorax, and abdomen/pelvis acquisitions were performed with (1) fixed tube current, (2) ATCM, and (3) ATVS. Acquisitions were performed at various radiation dose levels to generate images at different levels of quality. Reference volume CT dose index (CTDIvol), reference image noise, and reference contrast-to-noise ratios were determined. The potential dose reductions with ATCM and ATVS were assessed.

RESULTS: The percent reduction of CTDIvol with ATCM ranged from 8% to 24% for head, 16% to 39% for thorax, and 25% to 41% for abdomen/pelvis. The percent reduction of CTDIvol with ATVS varied on the clinical indication. In CT angiography, ATVS resulted to the highest dose reduction, which was up to 70% for head, 77% for thorax, and 34% for abdomen/pelvis. In noncontrast examinations, ATVS increased dose by up to 21% for head, whereas reduced dose by up to 34% for thorax and 48% for abdomen/pelvis.

CONCLUSIONS: In pediatric CT, the use of ATCM significantly reduces radiation dose and maintains image noise. The additional use of ATVS reduces further the radiation dose for thorax and abdomen/pelvis, and maintains contrast-to-noise ratio for the specified clinical diagnostic task.