评估颈动脉狭窄的超声与磁共振成像技术:一种新型体模方法
本研究提出一种可用于超声和磁共振扫描仪的新型专用体模,用于评估颈动脉粥样硬化病变。体模由聚甲基丙烯酸甲酯制成,内含模拟血管的中空圆柱及模拟狭窄的塑料管。在2台超声和2台磁共振扫描仪上进行基于准确性的定量图像分析,结果显示3.0 T磁共振扫描仪在表征血管狭窄方面有优势,但超声与磁共振技术间Fisher精确检验及评分者间一致性无显著差异,部分方法间一致性相关系数显示中等一致,3.0 T与其他方法间一致性较差,可能与3.0 T分辨率优于超声和1.5 T磁共振有关。这些方法在评估血管狭窄方面似乎具有相似效能,支持所开发体模作为通用且可重复的工具用于质量控制程序。
要点
- 3.0 T磁共振在表征血管狭窄的准确性测量中显示出优势。
- 超声与磁共振技术间Fisher精确检验及评分者间一致性无显著差异。
- 部分方法间一致性相关系数显示中等一致,3.0 T与其他方法间一致性较差。
核心结论
- 这些方法在评估血管狭窄方面似乎具有相似效能,支持所开发体模作为通用且可重复的工具用于质量控制程序。
研究设计与样本
- 本研究提出一种可用于超声和磁共振扫描仪的新型专用体模,由聚甲基丙烯酸甲酯制成,内含模拟血管的中空圆柱及模拟狭窄的塑料管,在2台超声和2台磁共振扫描仪上进行基于准确性的定量图像分析。
诊断标准与关键数值
- 准确性测量显示3.0 T磁共振扫描仪用于表征血管狭窄。
- Fisher精确检验和评分者间一致性显示超声与磁共振技术间无显著差异。
- 一致性相关系数显示部分方法间中等一致,3.0 T与其他方法间一致性较差。
局限性
- 3.0 T与其他方法间一致性较差,原文认为这可能由于3.0 T分辨率优于超声和1.5 T磁共振。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Carotid atherosclerosis is very important in the pathogenesis of cerebral ischemia. Ultrasonography (US) and magnetic resonance imaging (MRI) are the predominant noninvasive techniques capable to identify the presence and stage of intra-plaque hemorrage. In this work, we propose a novel dedicated phantom that can be used for both US and MRI scanners to evaluate carotid atherosclerotic lesions. The phantom consists of a polymethyl metacrylate (PMMA) diagonally crossed by a PMMA hollow cylinder simulating a blood vessel. To simulate a stenosis, we inserted a plastic hollow tube inside the cylinder. Quantitative image analysis, based on accuracy measurements, was performed on two US and two MRI scanners. The accuracy measurements have highlighted the use of the 3.0 T MRI scanner to characterize the vessel stenosis. However, no significant difference between US and MRI techniques was found in Fisher exact test and inter-rater agreement. The concordance correlation coefficient showed a moderate agreement between some methods. Agreement between 3.0 T and other methods results poor, and this could be due to the fact that the 3.0 T has a better resolution compared to a US and MR 1.5 T. These methods seem to have similar efficacies for the evaluation of vessel stenosis, legitimizing the use of the developed phantom as a versatile and reproducible instrument that could be used during quality controls programs.
引用
- Bonanno L, Marino S, Morabito R, et al. Evaluation of US and MRI techniques for carotid stenosis: a novel phantom approach. Radiol Med. 2018. PMID: 30542910. doi:10.1007/s11547-018-0971-7
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。