光子计数探测器CT联合锡滤过减少金属伪影并提高剂量效率
在3D打印脊柱模型和20例临床EID-CT患者中,评估光子计数探测器CT(PCD-CT)高能量阈值(TH)图像联合锡滤过对金属伪影和剂量效率的影响。体模TH图像联合锡滤过减少金属伪影且噪声与TL图像相当;患者PCD-CT TH图像对皮质、骨小梁及植入物-骨小梁界面的显示评分显著高于EID-CT,噪声显著低于EID-CT。
要点
- PCD-CT高能量阈值图像联合锡滤过可减少金属伪影并提高剂量效率。
- 患者PCD-CT TH图像对皮质、骨小梁及植入物-骨小梁界面的显示评分显著高于EID-CT。
- PCD-CT TH图像噪声显著低于EID-CT,放射科医师强烈偏好PCD-CT TH图像。
核心结论
- PCD-CT高能量阈值图像联合锡滤过与EID-CT相比,金属伪影显著减少且剂量效率提高。
研究设计与样本
- 体模实验加前瞻性、机构审查委员会批准、符合健康保险流通与责任法案的研究,20例临床EID-CT患者接受PCD-CT扫描,使用锡滤过,由3名放射科医师阅片。
关键影像表现/征象
- 体模TH图像联合锡滤过减少金属伪影,噪声(32 HU)与未使用锡滤过的TL图像(29 HU)相当。
- 患者PCD-CT TH图像对皮质、骨小梁和植入物-骨小梁界面的显示评分分别为4.4 ± 0.9、4.4 ± 1.0和4.4 ± 1.0,显著高于EID-CT图像的3.3 ± 1.3、3.3 ± 1.2和3.3 ± 1.6(P < 0.0001)。
- PCD-CT TH图像噪声(93 ± 41 HU)显著低于EID-CT图像(133 ± 92 HU,P < 0.05)。
- 放射科医师因诊断信心提高和伪影严重程度降低而强烈偏好PCD-CT TH图像。
临床应用与阅片要点
- PCD-CT高能量阈值图像联合锡滤过可用于减少金属伪影并提高剂量效率。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: The aim of this study was to investigate the impact on metal artifacts and dose efficiency of using a tin filter in combination with high-energy threshold (TH) images of a photon-counting detector (PCD) computed tomography (CT) system.
MATERIALS AND METHODS: A 3D-printed spine with pedicle screws was scanned on a PCD-CT system with and without tin filtration. Image noise and severity of artifacts were measured for low-energy threshold (TL) and TH images. In a prospective, institutional review board-approved, Health Insurance Portability and Accountability Act-compliant study, 20 patients having a clinical energy-integrating detector (EID) CT were scanned on a PCD-CT system using tin filtration. Images were reviewed by 3 radiologists to evaluate visualization of anatomic structures, diagnostic confidence, and image preference. Artifact severity and image noise were measured. Wilcoxon signed rank was used to test differences between PCD-CT TH and EID-CT images.
RESULTS: Phantom TH images with tin filtration reduced metal artifacts and had comparable noise (32 HU) to TL images (29 HU) acquired without tin filtration. Visualization scores for the cortex, trabeculae, and implant-trabecular interface from PCD-CT TH images (4.4 ± 0.9, 4.4 ± 1.0, and 4.4 ± 1.0) were significantly higher (P < 0.0001) than EID-CT images (3.3 ± 1.3, 3.3 ± 1.2, and 3.3 ± 1.6). A strong preference was shown for PCD-CT TH images due to improved diagnostic confidence and decreased artifact severity. Noise in PCD-CT TH images (93 ± 41 HU) was significantly lower than that in EID-CT images (133 ± 92 HU, P < 0.05).
CONCLUSIONS: Threshold high images acquired with tin filtration on PCD-CT demonstrated a substantial decrease in metal artifacts and an increase in dose efficiency compared with EID-CT.