在16厘米宽探测器CT上采用70 kVp和ASiR-V算法降低瘦体型患者冠状动脉CT血管成像的辐射与对比剂剂量:与常规100 kVp方案的比较
前瞻性纳入30例BMI≤23 kg/m2患者行70 kVp前瞻性心电触发CCTA(A组),与数据库中30例BMI匹配的常规100 kVp CCTA患者(B组)比较。两组平均CT值、对比噪声比及主观图像质量无显著差异,但A组有效剂量较B组降低75.3%(0.43±0.20 mSv vs 1.74±1.01 mSv),对比剂剂量减少42.4%(22.46±2.94 ml vs 38.99±5.10 ml)。
要点
- 70 kVp联合80% ASiR-V在16厘米宽探测器CT上可获亚毫希沃特级辐射剂量下的诊断级图像。
- A组有效剂量0.43±0.20 mSv,较B组1.74±1.01 mSv降低75.3%(p<0.001)。
- A组对比剂剂量22.46±2.94 ml,较B组38.99±5.10 ml减少42.4%(p<0.001)。
核心结论
- 对于BMI≤23 kg/m2患者,采用70 kVp、80% ASiR-V及16厘米宽探测器CT的前瞻性心电触发CCTA,与常规100 kVp方案相比,在提供诊断级图像的同时大幅降低辐射剂量和对比剂剂量。
研究设计与样本
- 前瞻性研究:A组30例BMI≤23 kg/m2患者行70 kVp CCTA(噪声指数36 HU,对比剂剂量率16 mg I/kg/s、注射9秒,80% ASiR-V重建);B组为数据库中BMI匹配的30例接受常规100 kVp CCTA者(噪声指数25 HU,对比剂剂量率25 mg I/kg/s、注射10秒,60% ASiR-V重建),比较辐射剂量、对比剂剂量和图像质量。
关键影像表现/征象
- 两组平均CT值、对比噪声比(CNR)及主观图像质量均无显著差异(所有p>0.05)。
- A组有效剂量0.43±0.20 mSv,B组为1.74±1.01 mSv,A组较B组降低75.3%(p<0.001)。
- A组对比剂剂量22.46±2.94 ml,B组为38.99±5.10 ml,A组较B组减少42.4%(p<0.001)。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: To investigate the value of prospectively ECG-triggered coronary CT angiography (CCTA) for lean patients with body mass index (BMI) ≤ 23 kg/m2 using 70 kVp and high-level volume-based adaptive statistical iterative reconstruction (ASiR-V) algorithm on a 16-cm wide-detector CT system for reducing both radiation and contrast doses in comparison with the conventional 100-kVp protocol.
MATERIALS AND METHODS: Thirty patients (group A) were prospectively enrolled to undergo 70-kVp CCTA on a 16-cm wide-detector CT scanner with noise index (NI) of 36 HU and at weight-dependent contrast dose rate of 16 mg I/kg/s for 9-s injection. Images were reconstructed with 80% ASiR-V. Radiation dose, contrast dose, and image quality were statistically compared with 30 patients (group B) in database with matching BMI who underwent conventional 100-kVp CCTA with NI of 25 HU, and at 25 mg I/kg/s rate for 10-s injection and reconstructed with 60% ASiR-V.
RESULTS: There was no significant difference in patient demographics between the two groups (all p > 0.05). The two groups also had similar mean CT values and contrast-noise ratio (CNR) and subjective image quality (all p > 0.05). However, group A with 70 kVp reduced the effective dose by 75.3% compared with group B (0.43 ± 0.20 mSv vs. 1.74 ± 1.01 mSv, p < 0.001), and required 42.4% less contrast dose than group B (22.46 ± 2.94 ml vs. 38.99 ± 5.10 ml, p < 0.001).
CONCLUSIONS: Prospectively ECG-triggered CCTA using 70 kVp and high-level ASiR-V on a 16-cm wide-detector CT system provides diagnostic images with substantial reduction in both radiation and contrast doses for patients with BMI ≤ 23 kg/m2 compared to the conventional 100-kVp protocol.
KEY POINTS: • 70-kVp CCTA produces excellent images at sub-millisievert radiation. • 70-kVp CCTA reduces both radiation and contrast doses over conventional protocol. • Achieving low-dose CCTA with combined low kVp and high-level ASIR-V.