有效探测器像素与CT体素大小对增强微血管血容量准确估计的影响
本研究在离体猪心肌标本中,以微CT 20 µm体素测值为参考,评估探测器像素大小(20–260 µm)和图像体素大小对微血管密度估计的影响,并比较全身EID CT与PCD CT的测量准确性;微CT体素至260 µm时微血管密度最多低估11%,全身EID CT低估9.5%,PCD CT高估0.1%。
要点
- 以微CT 20 µm体素测值为参考,微CT体素至260 µm时微血管密度最多低估11%。
- 全身EID CT重建600 µm层厚时微血管密度低估9.5%,PCD CT重建250 µm层厚时高估0.1%。
- 通过测量增强微血管所致CT衰减增加,可在临床CT较大探测器像素条件下较准确计算微血管密度。
核心结论
- 数据表明,通过测量这些增强微血管引起的CT衰减增加,可以用临床CT扫描仪中较大的探测器像素准确计算微血管密度。
研究设计与样本
- 离体猪心肌标本研究,使用微CT扫描仪以20至260 µm的探测器像素大小和图像体素大小生成数据,同一标本在全身EID CT和PCD CT系统扫描并分别重建600 µm和250 µm层厚。
诊断标准与关键数值
- 以20 µm体素大小测得的微血管密度值归一化为参考真值。
- 微CT模拟体素至260 µm时,微血管密度最多低估11%。
- 全身EID CT(600 µm层厚)微血管密度低估9.5%,PCD CT(250 µm层厚)高估0.1%。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
RATIONALE AND OBJECTIVES: The purpose of this study was to determine the impact of effective detector-pixel-size and image voxel size on the accurate estimation of microvessel density (ratio of microvascular lumen volume/tissue volume) in an excised porcine myocardium specimen using microcomputed tomography (CT), and the ability of whole-body energy-integrating-detector (EID) CT and photon-counting-detector (PCD) CT to measure microvessel density in the same ex vivo specimen.
MATERIALS AND METHODS: Porcine myocardial tissue in which the microvessels contained radio-opaque material was scanned using a micro-CT scanner and data were generated with a range of detector pixel sizes and image voxel sizes from 20 to 260 microns, to determine the impact of these parameters on the accuracy of microvessel density estimates. The same specimen was scanned in a whole-body EID CT and PCD CT system and images reconstructed with 600 and 250 micron slice thicknesses, respectively. Fraction of tissue volume that is filled with opacified microvessels was determined by first subtracting the mean background attenuation value from all voxels, and then by summing the remaining attenuation.
RESULTS: Microvessel density data were normalized to the value measured at 20 µm voxel size, which was considered reference truth for this study. For emulated micro-CT voxels up to 260 µm, the microvessel density was underestimated by at most 11%. For whole-body EID CT and PCD CT, microvessel density was underestimated by 9.5% and overestimated by 0.1%, respectively.
CONCLUSION: Our data indicate that microvessel density can be accurately calculated from the larger detector pixels used in clinical CT scanners by measuring the increase of CT attenuation caused by these opacified microvessels.