文献速递心血管US

应用三维斑点追踪超声心动图评估永久性选择性希氏束起搏对左心室同步性的影响

Assessment of Permanent Selective His Bundle Pacing in Left Ventricular Synchronization Using 3-D Speckle Tracking Echocardiography.
PubMed 文献 · Ultrasound Med Biol · 2018-12-01

研究将62例患者随机分为选择性希氏束起搏组(32例)和右心室心尖部起搏组(30例),术后及6个月随访时用三维斑点追踪超声心动图评估左心室收缩机械同步性。结果显示两组在术后1周和6个月时均检出显著左心室不同步,希氏束起搏组起搏阈值更高、R波振幅更低,但起搏参数在中期随访中与传统右心室心尖部起搏同样稳定。作者认为三维斑点追踪超声心动图评估左心室同步性可行且更便捷。

要点

  • 选择性希氏束起搏组起搏阈值在随访期间显著高于右心室心尖部起搏组。
  • 选择性希氏束起搏组R波振幅显著低于右心室心尖部起搏组。
  • 术后1周和6个月时,两组均检出显著左心室不同步。

核心结论

  • 三维斑点追踪超声心动图评估左心室同步性可行,且为评估左心室同步性提供了更便捷的方法。

研究设计与样本

  • 随机分组研究,共62例患者,选择性希氏束起搏组32例、右心室心尖部起搏组30例,采用三维超声心动图标准心尖四腔心切面转化为全容积模式,离线分析三维斑点追踪超声心动图,主要终点为术后及6个月随访时的左心室机械同步性。

诊断标准与关键数值

  • 通过评估16节段收缩期达峰时间三维应变的最大时间差和标准差,在术后1周和6个月时检出显著左心室不同步。
  • 选择性希氏束起搏组起搏阈值在整个随访期间显著高于右心室心尖部起搏组。
  • 选择性希氏束起搏组R波振幅显著低于右心室心尖部起搏组。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

The study described here aimed to evaluate left ventricular (LV) systolic mechanical synchronization during permanent selective His bundle pacing (SHBP) using 3-D speckle-tracking echocardiography post-operatively and 6 mo after pacemaker implantation in 62 patients randomly assigned to SHBP (n = 32) or right ventricular apical pacing (RVAP, n = 30). A standard apex four-chamber view was exposed and was transformed into full-volume mode under 3-D echocardiography. Three-dimensional speckle-tracking echocardiography was analyzed offline. The primary endpoint was LV mechanical synchronization post-operatively and during the 6-mo follow-up. Significant LV dyssynchrony was detected while evaluating the maximum time difference and standard deviation of 16-segment systolic time to peak 3-D strain at 1 wk and 6 mo. The pacing thresholds were significantly higher in the SHBP than in the RVAP group throughout follow-up. The R-wave amplitude was significantly lower in the SHBP group than with RVAP. The pacing parameters during SHBP were as stable as during conventional RVAP during the mid-term follow-up. In conclusion, 3-D speckle-tracking echocardiography is feasible and provides a more convenient method for evaluating LV synchrony.

来源:Ultrasound Med Biol,查看原文。版权归原作者所有。

引用

  1. Tang J, Chen S, Liu L, et al. Assessment of Permanent Selective His Bundle Pacing in Left Ventricular Synchronization Using 3-D Speckle Tracking Echocardiography. Ultrasound Med Biol. 2018. PMID: 30509784. doi:10.1016/j.ultrasmedbio.2018.10.006

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。