验证一种超声方案测量孤立功能性脊柱单元在功能性扭转运动中椎间轴向扭转的作用

Validation of an Ultrasound Protocol to Measure Intervertebral Axial Twist during Functional Twisting Movements in Isolated Functional Spinal Units.
PubMed 文献 · Ultrasound Med Biol · 2018-12-11

本研究在猪离体功能性脊柱单元模型上,用超声成像方案测量椎体节段旋转,并与光学运动捕捉系统进行验证。第一部分确认超声探头角度对声像图点数字化无影响;第二部分对12个猪功能性脊柱单元施加压缩、屈伸和轴向扭转,两种测量系统间相关性>0.903,在所有屈伸姿势下绝对系统误差为0.01°。

要点

  • 超声与光学运动捕捉系统测量轴向扭转的相关性>0.903,绝对系统误差0.01°。
  • 超声探头角度对声像图点数字化无影响,支持该超声方案的可行性。
  • 超声可在体外脊柱运动节段追踪轴向扭转运动,并具有在体内评估绝对椎间轴向扭转运动的潜力。

核心结论

  • 超声可用于追踪体外脊柱运动节段的轴向扭转运动,并具有在体内评估绝对椎间轴向扭转运动的潜力。

研究设计与样本

  • 研究论文;第一部分验证超声探头角度对声像图点数字化的影响;第二部分采用12个猪功能性脊柱单元固定于机械测试系统,施加压缩(压缩耐受度的15%)、屈伸及轴向扭转(0°、2°、4°或6°),以光学运动捕捉系统和后表面超声追踪轴向扭转运动。

诊断标准与关键数值

  • 超声与光学运动捕捉系统测量轴向扭转的相关性>0.903。
  • 在所有屈伸姿势下,两种测量系统的绝对系统误差为0.01°。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Ultrasound has potential for use in evaluation of bone and joint movement during axial twist of the lumbar spine both in vivo and in vitro. Such segmental rotations could then be measured under controlled external thoracic axial twist conditions and in response to mechanical loading. The purpose of this study was to measure vertebral segmental rotations in a porcine model of the human lumbar spine using an ultrasound imaging protocol and to validate use of this imaging technique with an optical motion capture system. In part 1, ultrasound transducer angle was confirmed to have no effect on sonogram point digitization. In part 2, 12 porcine functional spinal units were fixed to a mechanical testing system, and compression (15% of compressive tolerance), flexion-extension and axial twist (0°, 2°, 4° or 6°) were applied. Axial twist motion was tracked using an optical motion capture system and posterior surface ultrasound. Correlation between the two measurement systems was >0.903, and absolute system error was 0.01° across all flexion-extension postures. These findings indicate that ultrasound can be used to track axial twist motion in an in vitro spine motion segment and has the potential for use in vivo to evaluate absolute intervertebral axial twist motion.

来源:Ultrasound Med Biol,查看原文。版权归原作者所有。

引用

  1. McKinnon CD, Callaghan JP. Validation of an Ultrasound Protocol to Measure Intervertebral Axial Twist during Functional Twisting Movements in Isolated Functional Spinal Units. Ultrasound Med Biol. 2018. PMID: 30545788. doi:10.1016/j.ultrasmedbio.2018.10.023

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。