文献速递神经MR

压缩感知-敏感度编码加速脑成像:扫描时间缩短而图像质量未降低

Compressed Sensing-Sensitivity Encoding (CS-SENSE) Accelerated Brain Imaging: Reduced Scan Time without Reduced Image Quality.
PubMed 文献 · AJNR Am J Neuroradiol · 2018-12-06

本研究纳入66例已知、治疗后或疑似脑肿瘤患者,比较压缩感知-敏感度编码加速3D T1-扰相梯度回波与3D T2-FLAIR序列同常规采集的图像质量。两位神经放射学家盲法评估显示,加速序列图像质量与常规采集相似,SNR与对比噪声比无显著差异,FLAIR和扰相梯度回波序列扫描时间分别缩短25%和35%。

要点

  • 压缩感知-敏感度编码加速3D T1-扰相梯度回波与T2-FLAIR序列图像质量与常规采集相似。
  • FLAIR和扰相梯度回波序列扫描时间分别缩短25%和35%。
  • SNR与对比噪声比在加速与常规序列间无显著差异。

核心结论

  • 压缩感知-敏感度编码加速的3D T1-扰相梯度回波和T2-FLAIR脑序列图像质量与标准采集相似,且扫描时间缩短;该技术可能在不牺牲图像质量的情况下缩短扫描时间。

研究设计与样本

  • 前瞻性临床研究,纳入66例已知、治疗后或疑似脑肿瘤患者,在其临床脑MR检查中加做压缩感知-敏感度编码加速3D T1-扰相梯度回波或3D T2-FLAIR序列及对应的常规采集;由两位对序列和患者信息盲法的神经放射学家独立评估。

诊断标准与关键数值

  • 采用4分或5分Likert量表评估整体图像质量、SNR、伪影范围/严重程度、灰白质交界和病灶边界锐利度。
  • T1-扰相梯度回波序列中,压缩感知-敏感度编码在全部5项指标上的图像质量平均略优于常规图像,但无统计学显著差异,95%置信区间下限提示其图像质量在常规成像10%以内。
  • T2-FLAIR序列中,压缩感知-敏感度编码图像在5项指标中的3项平均在常规图像10%以内。
  • 压缩感知-敏感度编码图像伪影稍多(P = .068)、灰白质锐利度稍低(P = .36),但差异均无统计学意义。
  • SNR和对比噪声比无显著差异。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND AND PURPOSE: Compressed sensing-sensitivity encoding is a promising MR imaging acceleration technique. This study compares the image quality of compressed sensing-sensitivity encoding accelerated imaging with conventional MR imaging sequences.

MATERIALS AND METHODS: Patients with known, treated, or suspected brain tumors underwent compressed sensing-sensitivity encoding accelerated 3D T1-echo-spoiled gradient echo or 3D T2-FLAIR sequences in addition to the corresponding conventional acquisition as part of their clinical brain MR imaging. Two neuroradiologists blinded to sequence and patient information independently evaluated both the accelerated and corresponding conventional acquisitions. The sequences were evaluated on 4- or 5-point Likert scales for overall image quality, SNR, extent/severity of artifacts, and gray-white junction and lesion boundary sharpness. SNR and contrast-to-noise ratio values were compared.

RESULTS: Sixty-six patients were included in the study. For T1-echo-spoiled gradient echo, image quality in all 5 metrics was slightly better for compressed sensing-sensitivity encoding than conventional images on average, though it was not statistically significant, and the lower bounds of the 95% confidence intervals indicated that compressed sensing-sensitivity encoding image quality was within 10% of conventional imaging. For T2-FLAIR, image quality of the compressed sensing-sensitivity encoding images was within 10% of the conventional images on average for 3 of 5 metrics. The compressed sensing-sensitivity encoding images had somewhat more artifacts (P = .068) and less gray-white matter sharpness (P = .36) than the conventional images, though neither difference was significant. There was no significant difference in the SNR and contrast-to-noise ratio. There was 25% and 35% scan-time reduction with compressed sensing-sensitivity encoding for FLAIR and echo-spoiled gradient echo sequences, respectively.

CONCLUSIONS: Compressed sensing-sensitivity encoding accelerated 3D T1-echo-spoiled gradient echo and T2-FLAIR sequences of the brain show image quality similar to that of standard acquisitions with reduced scan time. Compressed sensing-sensitivity encoding may reduce scan time without sacrificing image quality.

来源:AJNR Am J Neuroradiol,查看原文。版权归原作者所有。

引用

  1. Vranic JE, Cross NM, Wang Y, et al. Compressed Sensing-Sensitivity Encoding (CS-SENSE) Accelerated Brain Imaging: Reduced Scan Time without Reduced Image Quality. AJNR Am J Neuroradiol. 2018. PMID: 30523142. doi:10.3174/ajnr.A5905

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。