文献速递技术与安全CT

双源CT虚拟单源CT的图像质量表现

Image Quality Performance of Virtual Single-Source CT Using Dual-Source Computed Tomography.
PubMed 文献 · Acad Radiol · 2018-10-25

该体模研究在同一辐射剂量下比较双源与单源CT数据集的图像质量,以评估双源扫描协议的剂量分割方法。结果显示双源模式噪声降低、CNR改善,SSIM和空间分辨率与单源模式差异可忽略(空间分辨率损失<5%),提示双源CT可在不损失图像质量的情况下分割辐射剂量。

要点

  • 双源模式较单源模式图像噪声降低,CNR改善。
  • 双源模式SSIM与单源模式几乎恒定,空间分辨率下降<5%。
  • 双源CT可将辐射剂量分配给两个X射线管而不损失图像质量。

核心结论

  • 体模研究表明,双源CT协议的SSIM和空间分辨率性能与单源CT相比变化可忽略,但噪声和CNR在双源CT显示改善,因此双源CT允许在X射线管之间分割辐射剂量并比较不同辐射剂量水平的数据集而不损失图像质量。

研究设计与样本

  • 体模研究,使用192层双源第三代CT(Somatom Force,Siemens Healthcare,Forchheim,Germany)对Catphan体模(Catphan503;The Phantom Laboratory,Salem,New York)成像,管电压70–120 kVp,采用基于器官的管电流调制技术(X-CARE),管电流时间乘积80–300 mAs,分别以单源和双源模式采集,双源模式管电流比值为50%|50%(DS 0.5)、60%|40%(DS 0.6)、70%|30%(DS 0.7)、80%|20%(DS 0.8)、90%|10%(DS 0.9),重建采用滤波反投影(Br40)和高级模型迭代重建算法(ADMIRE)强度1–5。

诊断标准与关键数值

  • 图像质量评估指标包括噪声、对比噪声比(CNR)、低对比度可检测性(以结构相似性指数SSIM表示)和空间分辨率(以线扩散函数半高全宽量化);双源模式空间分辨率性能损失低于5%。

临床应用与阅片要点

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

RATIONALE AND OBJECTIVES: The aim of this study is to analyze the image quality provided by a dual-source (DS) data set and a single-source (SS) data set at the same radiation exposure, in order to evaluate a dose splitting method for dual-source scanning protocols.

MATERIALS AND METHODS: A 192-slice dual-source third generation CT (Somatom Force; Siemens Healthcare, Forchheim, Germany) was used to image a Catphan phantom (Catphan503; The Phantom Laboratory, Salem, New York) utilizing different X-ray tube voltages from 70 to 120 kVp with an organ-based tube-current modulation technique (X-CARE; Siemens Healthcare, Forchheim, Germany). In order to keep the radiation dose (given by volume computed tomography dose index) in a clinically relevant range, different X-ray tube time-current products ranging from 80 to 300 mAs were selected. The data sets by each X-ray tube voltage were collected using a single-source as well as a dual-source mode. The measurements in the dual-source mode were performed with five different tube currents of the X-ray tube A and B. Thereby, the tube current ratios were 50%|50% (DS 0.5), 60%|40% (DS 0.6), 70%|30% (DS 0.7), 80%|20% (DS 0.8), and 90%|10% (DS 0.9). The images were reconstructed by the use of a filter-back projection (Br40) and an advanced mode led iterative reconstruction algorithms (advanced modeled iterative reconstruction algorithms [ADMIRE]; Siemens Healthcare, Forchheim, Germany) with a strength range of 1-5. The image quality was evaluated in terms of noise, contrast-to-noise ratio (CNR), low-contrast detectability expressed as the structural similarity index (SSIM) and spatial resolution quantified by the full width at half maximum of the line-spread function.

RESULTS: Image noise decreased by the use of the dual-source mode, which led to improvement of their CNR compared to the single-source mode. SSIM showed an almost constant behavior by both modes. The spatial resolution indicated a lower trend by the dual-source mode in comparison to the single-source mode. However, the loss of the spatial resolution performance was lower than 5% for the dual-source modes.

CONCLUSION: The presented phantom study demonstrated that SSIM and spatial resolution performance obtained by dual-source CT protocols showed a negligible variation to those by the single-source CT. However, the noise and CNR displayed an improvement for the dual-source CT. Therefore, the use of the dual-source CT enables to split the radiation dose between X-ray tubes and to compare the data sets with different radiation dose levels without loss in the image quality.

来源:Acad Radiol,查看原文。版权归原作者所有。

引用

  1. Alikhani B, Werner M, Jamali L, et al. Image Quality Performance of Virtual Single-Source CT Using Dual-Source Computed Tomography. Acad Radiol. 2018. PMID: 30482627. doi:10.1016/j.acra.2018.09.021

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。