文献速递儿科影像技术与安全CTDR

低成本儿童骨盆模体的构建与验证

Construction and validation of a low cost paediatric pelvis phantom.
PubMed 文献 · Eur J Radiol · 2018-09-14

研究以5岁儿童解剖为基础,用巴黎石膏(PoP)模拟儿童骨、聚甲基丙烯酸甲酯(PMMA)模拟软组织构建低成本儿童骨盆模体,并通过CT密度比较、信噪比(SNR)分析和心理测量图像质量量表进行验证。结果显示皮质骨CT密度差异为-4.7%至-4.1%、软组织为-23.4%,SNR在全部kVp下两模体强正相关(r>0.95),心理测量量表Cronbach's Alpha接近0.8。作者认为该低成本模体具有适用于X线成像的解剖特征,可能用于剂量与图像质量优化研究。

要点

  • 皮质骨替代物CT密度与真实儿童相差-4.7%至-4.1%,软组织替代物相差-23.4%。
  • 全部kVp下模体与商用模体SNR强正相关(r>0.95)。
  • 心理测量图像质量量表Cronbach's Alpha接近0.8。

核心结论

  • 物理与视觉验证结果提示该低成本模体具有适用于X线成像的解剖特征,可能用于剂量与图像质量优化研究。

研究设计与样本

  • 基于5岁儿童解剖构建模体,用巴黎石膏(PoP)代表儿童骨、聚甲基丙烯酸甲酯(PMMA)代表软组织,以CT扫描匹配骨性解剖并预留灌注PoP的腔隙;验证包括CT密度比较、前后位X线图像SNR对比分析(对照商用拟人模体)及心理测量图像质量量表(表面效度)。

诊断标准与关键数值

  • CT密度百分比差异:皮质骨为-4.7%至-4.1%,软组织为-23.4%。
  • SNR(mAs响应):两模体在全部kVp下强正相关(r>0.95)。
  • SNR(kVp响应):1至8 mAs区间强正相关(r=0.85),随mAs增加相关性下降(20 mAs时r=-0.21)。
  • 心理测量图像质量量表Cronbach's Alpha接近0.8。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: Imaging phantoms can be cost prohibitive, therefore a need exists to produce low cost alternatives which are fit for purpose. This paper describes the development and validation of a low cost paediatric pelvis phantom based on the anatomy of a 5-year-old child.

METHODS: Tissue equivalent materials representing paediatric bone (Plaster of Paris; PoP) and soft tissue (Poly methyl methacrylate; PMMA) were used. PMMA was machined to match the bony anatomy identified from a CT scan of a 5-year-old child and cavities were created for infusing the PoP. Phantom validation comprised physical and visual measures. Physical included CT density comparison between a CT scan of a 5-year old child and the phantom and Signal to Noise Ratio (SNR) comparative analysis of anteroposterior phantom X-ray images against a commercial anthropomorphic phantom. Visual analysis using a psychometric image quality scale (face validity).

RESULTS: CT density, the percentage difference between cortical bone, soft tissue and their equivalent tissue substitutes were -4.7 to -4.1% and -23.4%, respectively. For SNR, (mAs response) there was a strong positive correlation between the two phantoms (r > 0.95 for all kVps). For kVp response, there was a strong positive correlation between 1 and 8 mAs (r = 0.85), this then decreased as mAs increased (r = -0.21 at 20 mAs). Psychometric scale results produced a Cronbach's Alpha of almost 0.8.

CONCLUSIONS: Physical and visual measures suggest our low-cost phantom has suitable anatomical characteristics for X-ray imaging. Our phantom could have utility in dose and image quality optimisation studies.

来源:Eur J Radiol,查看原文。版权归原作者所有。