下肢CT血管成像新方案中的辐射剂量与图像质量
前瞻性研究将60例下肢CTA患者随机分入单源CT头足方向采集组(P1)与双源CT Flash足头方向采集组(P2),比较辐射剂量与图像质量。P2组血管内衰减、信噪比、对比噪声比及膝下小动脉图像质量评分均高于P1组,CTDIvol降低40.9%、DLP降低42.8%。
要点
- 双源CT Flash足头方向采集较单源CT头足方向采集CTDIvol降低40.9%、DLP降低42.8%。
- 双源CT Flash方案血管内衰减586.8 HU高于单源方案496.1 HU(p=0.011)。
- 双源CT Flash方案膝下小动脉图像质量评分3.8分高于单源方案3.1分(p=0.001)。
核心结论
- 下肢CTA采用双源CT Flash足头方向采集较单源CT头足方向采集辐射剂量更低,血管内衰减、信噪比、对比噪声比更高,膝下小动脉图像质量更好。
研究设计与样本
- 前瞻性研究,60例因外周闭塞性动脉疾病行下肢CTA的患者随机分为A组(对照组,P1方案:单源CT头足方向采集)与B组(实验组,P2方案:双源CT Flash足头方向采集)。
诊断标准与关键数值
- P2组血管内衰减586.8±140.3 HU vs P1组496.1±129.3 HU(p=0.011)。
- P2组信噪比33.0±11.3 vs P1组27.4±12.3(p=0.042)。
- P2组对比噪声比30.1±13.3 vs P1组24.2±10.3(p=0.029)。
- P2组膝下小动脉图像质量评分3.8±0.2 vs P1组3.1±0.2(p=0.001)。
- P2组CTDIvol 1.3±0.1 mGy vs P1组2.2±0.3 mGy,CTDIvol降低40.9%(p=0.006)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To compare radiation dose and image quality of lower extremity computed tomography angiography (CTA) between cranio-caudal acquisition with single-source CT (SSCT) and flash caudo-cranial acquisition with dual-source CT (DSCT).
MATERIALS AND METHODS: In this prospective study, 60 patients were randomly assigned to Group A (control) or Group B (experimental) to undergo lower extremity CTA for peripheral obliterative arterial disease. Group A received protocol 1 (P1) with SSCT cranio-caudal acquisition. Group B received protocol (P2) with DSCT flash caudo-cranial acquisition. Intravascular attenuation (IVA), signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR) and image noise were compared. Two radiologists assessed the image quality. Computed tomography volume dose index (CTDIvol) and dose-length product (DLP) were also compared.
RESULTS: IVA with P2 was higher than with P1 (586.8 ± 140.3 vs. 496.1 ± 129.3 HU, p = 0.011), as was SNR (33.0 ± 11.3 vs. 27.4 ± 12.3; p = 0.042), CNR (30.1 ± 13.3 vs. 24.2 ± 10.3; p = 0.029) and image quality score of small arteries below the knee (3.8 ± 0.2 vs. 3.1 ± 0.2; p = 0.001). Radiation dose was significantly lower in P2 than in P1 with CTDIvol reduction of 40.9% (1.3 ± 0.1 vs. 2.2 ± 0.3 mGy; p = 0.006) and DLP reduction of 42.8% (148.7 ± 21.9 vs 260.2 ± 59.1 mGy * cm; p = 0.018).
CONCLUSION: Lower extremity CTA with DSCT flash caudo-cranial acquisition allows lower radiation dose with higher IVA, SNR, CNR and better image quality for small arteries below the knee than SSCT cranio-caudal acquisition.
引用
- Schicchi N, Fogante M, Oliva M, et al. Radiation dose and image quality with new protocol in lower extremity computed tomography angiography. Radiol Med. 2018. PMID: 30478814. doi:10.1007/s11547-018-0963-7
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。