三维容积MR关节造影评估原发性粘连性肩关节囊炎患者肩关节容量
研究对28例临床及影像证实的原发性粘连性肩关节囊炎患者和25名对照行肩关节薄层三维容积MR关节造影,测量关节总容量、关节外对比剂外渗及肩袖间隙、腋隐窝、肱二头肌腱鞘容量,发现患者肩袖间隙和肱二头肌腱鞘容量显著降低、关节外对比剂外渗显著增多、关节总容量显著降低。
要点
- 原发性粘连性肩关节囊炎患者肩袖间隙和肱二头肌腱鞘容量较对照显著降低。
- 患者关节外对比剂外渗量显著高于对照,关节总容量显著低于对照。
- 肩袖间隙和肱二头肌腱鞘闭塞及关节容量降低可能是有用的诊断影像标准。
核心结论
- 三维容积MR关节造影上肱二头肌腱鞘和肩袖间隙闭塞及关节容量降低可能是诊断原发性粘连性肩关节囊炎的有用影像标准。
研究设计与样本
- 研究纳入28例临床和影像学证实的原发性粘连性肩关节囊炎患者及25名对照,行肩关节薄层三维容积MR关节造影序列,测量关节总容量、关节外对比剂外渗量及肩袖间隙、腋隐窝、肱二头肌腱鞘容量。
关键影像表现/征象
- 患者组肩袖间隙平均容量为7.67 ± 2.6 cm³,显著低于对照组12.31 ± 2.5 cm³(p < 0.0001)。
- 患者组肱二头肌沟平均容量为1.67 ± 0.9 cm³,显著低于对照组2.88 ± 0.9 cm³(p ≤ 0.0001)。
- 患者组关节外对比剂外渗平均容量为9.93 ± 1.7 cm³,显著高于对照组5.1 ± 1.4 cm³(p = 0.002)。
- 患者组关节总容量平均为22.52 ± 1.1 cm³,对照组为26.01 ± 1.2 cm³(p = 0.003)。
- MR关节造影显示71%的患者出现对比剂外渗,48%的对照者出现关节外对比剂外渗。
诊断标准与关键数值
- 肩袖间隙容量阈值未报告,患者组7.67 ± 2.6 cm³ vs 对照组12.31 ± 2.5 cm³(p < 0.0001)。
- 肱二头肌沟容量阈值未报告,患者组1.67 ± 0.9 cm³ vs 对照组2.88 ± 0.9 cm³(p ≤ 0.0001)。
- 关节外对比剂外渗量阈值未报告,患者组9.93 ± 1.7 cm³ vs 对照组5.1 ± 1.4 cm³(p = 0.002)。
- 关节总容量阈值未报告,患者组22.52 ± 1.1 cm³ vs 对照组26.01 ± 1.2 cm³(p = 0.003)。
临床应用与阅片要点
- 在原发性粘连性肩关节囊炎患者中,薄层三维容积MR关节造影是评估关节容量和关节囊结构的有用影像方法。
- 阅片时可关注肱二头肌腱鞘和肩袖间隙的闭塞以及关节容量降低,这些可能是有用的诊断影像标准。
- 关节外对比剂外渗在患者中占71%、对照中占48%,判读时需结合其他征象。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: To evaluate the three-dimensional (3D) volumetric MR arthrographic findings of shoulder joint capacity in patients with primary adhesive capsulitis.
METHODS: Thin-section 3D volumetric MR arthrography sequences of the shoulder joint were obtained in 28 patients with clinically and radiologically proven primary adhesive capsulitis and in 25 controls. Volumetric measurements of the total glenohumeral joint capacity, extra-articular contrast material leakage, and the rotator interval, axillary recess, and biceps tendon sheath capacities were performed for the study and control groups.
RESULTS: Mean volume of the rotator interval was 7.67 ± 2.6 cm3 in the study group, which was significantly lower than in the control group (12.31 ± 2.5 cm3) (p < 0.0001). Mean volume of the bicipital groove was significantly decreased in the patient group compared to the control group (1.67 ± 0.9 cm3 vs 2.88 ± 0.9 cm3) (p ≤ 0.0001). Mean volume of extra articular contrast material extravasation was 9.93 ± 1.7 cm3 in the patient group, which was significantly higher than in the control group (5.1 ± 1.4 cm3) (p = 0.002). Mean total glenohumeral joint volume was 22.52 ± 1.1 cm3 in the patient group and 26.01 ± 1.2 cm3 in the control group (p = 0.003).
CONCLUSION: On 3D volumetric MR arthrographic examination, obliterations in the biceps tendon sheath and rotator interval as well as decreased joint capacity may be useful imaging criteria for diagnosing primary adhesive capsulitis.
ADVANCES IN KNOWLEDGE: In the present study, MR arthrography showed contrast material extravasation in 71% of the patients and showed extra articular contrast material extravasation in 48% of the control subjects. On 3D volumetric MR arthrographic examination, obliterations in the biceps tendon sheath and rotator interval as well as decreased joint capacity may be useful imaging criteria for diagnosing primary adhesive capsulitis.In patients with primary adhesive capsulitis, thin-section 3D volumetric MR arthrography is a useful imaging modality to evaluate both joint capacity and capsular structure.