以低于50英镑的成本自制骨科CT体模
本研究用膝关节置换假体、腻子、聚苯乙烯等材料自制模拟股骨与胫骨及全膝关节置换的CT体模,将塑料条和圆柱置于金属植入物与骨替代物之间模拟骨病灶,总制作时间不足3小时、总成本低于50英镑,体模可运输、性能稳定,可用于比较不同CT设备显示金属植入物周围骨病灶的能力。
要点
- 体模由膝关节置换假体部件、腻子和聚苯乙烯制成,总成本低于50英镑、制作时间不足3小时。
- 体模为空时便于运输,可现场注水以获得可重复的衰减值,且随时间保持稳定。
- 作者认为该体模可用于设备采购比较、质量保证、扫描参数优化和研究。
核心结论
- 作者认为,用少量资源即可制作简单、可运输的CT体模,用于比较不同CT设备显示金属植入物周围骨病灶的能力。
研究设计与样本
- 本研究为体模研制研究,用膝关节置换假体部件、腻子和聚苯乙烯构建模拟股骨、胫骨及全膝关节置换的CT体模,塑料条和圆柱在制作时置于金属植入物与骨替代物之间作为「骨病灶」,体模固定于用环氧树脂密封的防水亚克力盒中。
关键影像表现/征象
- 体模模拟股骨与胫骨及全膝关节置换,塑料条和圆柱置于金属植入物与骨替代物之间充当「骨病灶」,未报告具体影像征象。
临床应用与阅片要点
- 作者认为该体模可用于设备采购过程中的直接比较,比较不同设备的技术因素和软件性能。
- 作者认为该体模可用于质量保证、扫描参数优化和研究。
- 作者认为该体模可客观评估临床实践中的成像方案。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE:: In procuring a CT scanner for orthopaedic imaging, the ability of the scanner to cope with metal artefacts for visualising bone and bone lesions around orthopaedic implants is an important feature. A durable and easily transportable CT phantom would help to compare this feature between CT scanners.The aims of this study were to develop a CT phantom that is easy to build, easily transportable, stable over time, cheap and challenging CT scanner performance.
METHODS:: A CT phantom resembling a femur and tibia with a total knee replacement was constructed from spare components of a knee replacement, wall filler and polystyrene. A number of plastic strips and cylinders were placed between metal implant and bone substitute during construction to act as "bone lesions". The phantom was fixed in a watertight acrylic box with epoxy resin.
RESULTS:: The total manufacturing time was below 3 h staggered over several days and the total cost was below £50. When empty, the phantom is easily transportable. The box can be filled with water on site visits ensuring a reproducible attenuation. This phantom is stable (i.e. not affected by decay of biological tissue).
CONCLUSION:: The phantom was easy to construct and is well transportable and stable in time. The phantom can be used in a procurement process allowing direct comparison of different scanners regarding technical factors and software performance. It can further be used for quality assurance, scan parameter optimisation and research. We conclude that a simple and transportable CT phantom can be built using few resources that allows to compare CT scanners with respect to their ability to visualise bone lesions around metal implants.
ADVANCES IN KNOWLEDGE:: It is possible to build a CT knee replacement phantom in a few hours and for less than £50. Other than the total knee implant, this CT phantom can be built with material available from any DIY store and simple tools. This CT phantoms allows objective comparisons in CT procurement. This CT phantom allows objective assessment of imaging protocols for clinical practice.
引用
- Tins B, Kuiper JH. Building an orthopaedic CT phantom for under £50. Br J Radiol. 2018. PMID: 30379580. doi:10.1259/bjr.20180279
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。