对比增强前后与仅对比增强锥形束乳腺CT:能否在保持诊断准确性的同时降低辐射暴露
本研究纳入49例(61侧乳腺)乳腺密度c/d型、乳腺X线或超声诊断为BI-RADS 4/5病变的女性,比较仅对比增强锥形束乳腺CT与增强前后联合采集(含减影图像)的诊断准确性及辐射剂量。结果显示两者AUC、敏感度、特异度无显著差异,而仅对比增强方案辐射剂量显著更低(中位平均腺体剂量5.9 mGy vs 11.7 mGy,p<0.001)。
要点
- 仅对比增强CBBCT与增强前后联合采集的诊断准确性无显著差异(AUC 0.84 vs 0.83,p=0.643)。
- 仅对比增强CBBCT的中位平均腺体辐射剂量为5.9 mGy,显著低于联合采集的11.7 mGy(p<0.001)。
- 减影图像中重度配准不良常见,尤其双侧检查时4级配准不良占58.3%。
核心结论
- 在c/d型乳腺中,仅对比增强CBBCT的诊断准确性与增强前后联合CBBCT相当,同时辐射剂量显著降低约两倍。
研究设计与样本
- 纳入49例女性(61侧乳腺),中位年龄57.9岁,乳腺X线或超声诊断为BI-RADS 4/5病变且乳腺密度为c/d型;两名放射科医师分别以BI-RADS量表独立评分仅对比增强CBBCT与增强前后联合CBBCT(含减影图像),计算观察者间和观察者内一致性,并以共识评估诊断准确性。
关键影像表现/征象
- 减影图像配准不良程度按1级(无伪影)至4级(伪影宽度>4 mm)评估,总体1/2/3/4级分别占16.9%/27.1%/16.9%/39%。
- 双侧检查中配准不良1/2/3/4级分别占3.2%/29.2%/8.3%/58.3%。
诊断标准与关键数值
- 仅对比增强CBBCT vs 增强前后联合CBBCT:AUC 0.84 vs 0.83(p=0.643),敏感度0.89 vs 0.85(p=0.158),特异度0.73 vs 0.76(p=0.655)。
- 观察者间和观察者内一致性均为优(组内相关系数ICC分别为0.76和0.83)。
- 中位平均腺体辐射剂量:仅对比增强CBBCT 5.9 mGy vs 增强前后联合CBBCT 11.7 mGy(p<0.001)。
临床应用与阅片要点
- 在c/d型乳腺、BI-RADS 4/5病变评估中,仅对比增强CBBCT可作为增强前后联合采集的替代方案,诊断准确性相当且辐射剂量更低。
- 减影图像配准不良常见,尤其双侧检查时4级配准不良占58.3%,可能限制减影图像判读。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: To evaluate whether post-contrast cone-beam breast CT (CBBCT) alone is comparable to the current standard of combined pre- and post-contrast CBBCT regarding diagnostic accuracy and superior regarding radiation exposure.
MATERIAL AND METHODS: This study included 49 women (61 breasts) with median age 57.9 years and BI-RADS 4/5 lesions diagnosed on mammography/ultrasound in density type c/d breasts. Two radiologists rated post-contrast CBBCT and pre- and post-contrast CBBCT with subtraction images on the BI-RADS scale separately for calculation of inter- and intra-observer agreement and in consensus for diagnostic accuracy assessment. Sensitivity, specificity, and area under the curve (AUC) were compared via McNemar test and DeLong method, respectively. Subtraction imaging misregistration were measured from 1 (no artifacts) to 4 (artifacts with width > 4 mm).
RESULTS: A total of 100 lesion (51 malignant; 6 high risk; 43 benign) were included. AUC, sensitivity, and specificity showed no significant differences comparing post-contrast CBBCT alone versus pre- and post-contrast CBBCT (AUC 0.84 vs. 0.83, p = 0.643; sensitivity 0.89 vs. 0.85, p = 0.158; specificity 0.73 vs. 0.76, p = 0.655). Inter- and intra-observer agreement was excellent (intra-class correlation coefficient ICC = 0.76, ICC = 0.83, respectively). Radiation dose was significantly lower for post-contrast CBBCT alone versus pre- and post-contrast CBBCT (median average glandular radiation dose 5.9 mGy vs. 11.7 mGy, p < 0.001). High-degree misregistrations were evident in the majority of subtraction images (level 1/2/3/4 16.9%/27.1%/16.9%/39%), in particular for bilateral exams (3.2%/29.2%/8.3%/58.3%).
CONCLUSION: Diagnostic accuracy of post-contrast CBBCT alone is comparable to pre- and post-contrast CBBCT in type c/d breasts, while yielding a significant twofold radiation dose reduction.
KEY POINTS: • The diagnostic accuracy of post-contrast CBBCT alone is comparable to dual acquisition of pre- and post-contrast CBBCT. • Acquisition of the post-contrast CBBCT scan alone reduces radiation exposure compared to pre- and post-contrast CBBCT, thus countering one of the main limitations of CBBCT. • High-degree misregistration artifacts limit the interpretation of subtraction images from pre- and post-contrast CBBCT studies.