平面合成乳腺摄影与旋转合成乳腺摄影对乳腺癌检出及显示的对比:基于乳腺断层合成的研究
该研究为全交叉多阅片者多病例研究,3名放射科医生回顾性判读190例(27例恶性、31例良性、132例正常),分别独立使用平面合成乳腺摄影(SM)和旋转合成乳腺摄影(RM),不参考DBT断层图像。结果显示SM与RM的AUC和敏感度无显著差异,但RM特异度显著高于SM(0.72 vs 0.66,P=0.028),且阅片者显著偏好RM对各类病灶的显示,但RM平均阅片时间更长(30秒 vs 23秒)。
要点
- RM与SM的乳腺癌检出AUC和敏感度无显著差异
- RM特异度显著高于SM(0.72 vs 0.66,P=0.028)
- 阅片者显著偏好RM对钙化和软组织病灶的显示,但RM阅片时间更长
核心结论
- 放射科医生使用RM与SM可获得相似的乳腺癌检出效果,且更偏好RM的类3D旋转显示,可能提高特异度,但代价是阅片时间延长。
研究设计与样本
- 全交叉多阅片者多病例(MRMC)研究,回顾性纳入190例(27例恶性、31例良性、132例正常),3名放射科医生分别独立使用SM和RM判读,不参考DBT断层图像,以BI-RADS和可疑度评分(1-10)评估,采用iMRMC软件进行MRMC方差分析。
关键影像表现/征象
- RM对钙化和软组织病灶的可见性视觉评分显著优于SM(P<0.05)
诊断标准与关键数值
- SM的AUC为0.66,RM的AUC为0.67,两者差异无统计学意义(P=0.818)
- SM敏感度为0.62,RM敏感度为0.60,两者差异无统计学意义(P=0.794)
- SM特异度为0.66,RM特异度为0.72,RM显著高于SM(P=0.028)
临床应用与阅片要点
- RM与SM均可用于乳腺癌检出,RM可能提高特异度,但需注意阅片时间较长
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To compare breast cancer detection and depiction between planar synthetic mammography (SM) and rotating synthetic mammography (RM) generated from digital breast tomosynthesis (DBT).
MATERIALS AND METHODS: In a fully-crossed multi-reader multi-case (MRMC) study, three radiologists retrospectively reviewed 190 cases (27 malignant, 31 benign, 132 normal), once with SM alone and once with RM alone, the DBT stack of slices was not reviewed. Lesions were scored using BI-RADS® and level of suspiciousness (1-10). Sensitivity, specificity, and area under the receiver operating characteristic curve (AUC) were computed using MRMC Analysis of Variance using the open-access software iMRMC. Additionally, readers were asked to make a visual grading analysis (VGA) on visibility of calcifications and soft tissue lesions (1-5 scale with 5 = Excellent visualization). The VGA scores were analyzed using the visual grading characteristics (VGC) method.
RESULTS: On average, the AUC was similar between SM and RM (0.66 versus 0.67, P = 0.818). The sensitivity was equivalent (0.62 versus 0.60, P = 0.794), while specificity was significantly lower in SM than in RM (0.66 versus 0.72, P = 0.028). Radiologists significantly (P < 0.05) preferred the display of all types of lesions in RM over SM. The average reading time per case was higher for RM than for SM (30 s versus 23 s, P < 0.05).
CONCLUSION: Radiologists achieve similar cancer detection with RM as with SM. They prefer the 3D-like rotating representation of soft tissue lesions and calcifications in comparison to the 2D visualization, which might improve their specificity, but at the expense of longer reading time.
引用
- Rodriguez-Ruiz A, Lardenoije S, Wanders AJT, et al. Comparison of breast cancer detection and depiction between planar and rotating synthetic mammography generated from breast tomosynthesis. Eur J Radiol. 2018. PMID: 30396675. doi:10.1016/j.ejrad.2018.09.022
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。