基于弹性成像点定量与瞬时弹性成像参考的肝脏硬度测量中幅度依赖性不一致
研究纳入185例受试者(20名健康志愿者和165例慢性肝病患者,其中128例为非酒精性脂肪性肝病),以瞬时弹性成像(TE)为参考标准,评估弹性成像点定量(ElastPQ)诊断肝纤维化的效能。ElastPQ诊断F≥2、F≥3和F4的最佳截断值分别为5.5 kPa(AUC=0.96)、8.1 kPa(AUC=0.98)和9.9 kPa(AUC=0.98),但两种方法的肝脏硬度测量值不可互换,尤其在TE值>10 kPa时差异增大且显著。
要点
- ElastPQ诊断F≥2、F≥3和F4的最佳截断值分别为5.5 kPa、8.1 kPa和9.9 kPa,对应AUC分别为0.96、0.98和0.98。
- TE与ElastPQ测量的平均差值为0.98 kPa(95%CI 0.51-1.45),两种方法相关性显著(r=0.86,p<0.001)。
- Bland-Altman图显示两种方法测量值存在差异,尤其在TE值>10 kPa时差异增大且显著,测量值不可互换。
核心结论
- ElastPQ是评估肝纤维化的可靠方法,但其肝脏硬度测量值与TE不可互换,尤其在TE值>10 kPa时;ElastPQ对代偿性晚期慢性肝病患者亚分类的诊断效能应进一步研究。
研究设计与样本
- 研究纳入185例受试者(20名健康志愿者和165例慢性肝病患者,其中128例为非酒精性脂肪性肝病,83例(44.9%)为女性,中位年龄53岁,BMI 27.3 kg/m2),以TE定义的肝纤维化分期为参考标准,采用相关分析、AUC分析、Bland-Altman图和Passing-Bablok回归比较TE与ElastPQ两种方法。
诊断标准与关键数值
- ElastPQ诊断F≥2的最佳截断值为5.5 kPa(AUC=0.96),诊断F≥3的最佳截断值为8.1 kPa(AUC=0.98),诊断F4的最佳截断值为9.9 kPa(AUC=0.98);研究人群中≥F2、≥F3和F4期患者分别占24.3%、13.5%和11.4%。
临床应用与阅片要点
- ElastPQ可用于评估肝纤维化,但其肝脏硬度测量值不能与TE互换,尤其在TE值>10 kPa时差异增大且显著。
局限性
- ElastPQ对代偿性晚期慢性肝病患者亚分类的诊断效能应进一步研究。
- 本文数据主要适用于非酒精性脂肪性肝病患者。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: To investigate diagnostic performance of point shear wave elastography by elastography point quantification (ElastPQ) for non-invasive assessment of liver fibrosis in patients with chronic liver diseases (CLD).
METHODS: Liver stiffness measurement (LSM) by transient elastography (TE) and ElastPQ was performed in patients with CLD and healthy volunteers. The stage of liver fibrosis was defined by TE which served as the reference. We compared two methods by using correlation, area under the receiver operating characteristics curve (AUC) analysis, Bland and Altman plot and Passing-Bablok regression.
RESULTS: A total of 185 subjects (20 healthy volunteers and 165 patients with CLD (128 non-alcoholic fatty liver disease), 83 (44.9%) females, median age 53 years, BMI 27.3 kg/m2) were evaluated. There were 24.3%, 13.5% and 11.4% patients in ≥ F2, ≥ F3 and F4 stage, respectively. The best performing cutoff LSM values by ElastPQ were 5.5 kPa for F ≥ 2 (AUC = 0.96), 8.1 kPa for F ≥ 3 (AUC = 0.98) and 9.9 kPa for F4 (AUC = 0.98). Mean (SD) difference between TE and ElastPQ measurements was 0.98 (3.27) kPa (95% CI 0.51-1.45, range 4.99-21.60 kPa). Two methods correlated significantly (r = 0.86; p < 0.001), yet Bland and Altman plot demonstrated difference between measurements, especially with TE values > 10 kPa. Passing and Bablok regression analysis yielded significant constant and proportional difference between ElastPQ and TE.
CONCLUSION: ElastPQ is reliable method for assessment of liver fibrosis but LSM values are not interchangeable with TE, especially above 10 kPa. Diagnostic performance of ElastPQ for sub-classification of patients with compensated advanced chronic liver disease should therefore be furtherly investigated.
KEY POINTS: • ElastPQ appears to be reliable method for assessment of liver fibrosis, with data presented here mostly applicable to NAFLD. • LSM values produced by TE and ElastPQ are NOT interchangeable-in values < 10 kPa, they are similar, but in values > 10 kPa, they appear to be increasingly and significantly different. • Diagnostic performance of ElastPQ for sub-classification of patients with compensated advanced chronic liver disease should be furtherly investigated.