1.5T与3T负荷心肌灌注心血管磁共振检测显著冠状动脉疾病的诊断准确性比较
前瞻性纳入281例疑似或已知冠状动脉疾病患者,分别接受1.5T(n=135)或3T(n=146)心血管磁共振心肌灌注检查,以侵入性冠状动脉造影为参考标准。1.5T与3T在检测显著冠状动脉疾病方面的诊断性能相似,仅在无陈旧性心肌梗死的多支血管病变患者中,3T的血管水平诊断性能更高。
要点
- 1.5T与3T负荷心肌灌注心血管磁共振检测显著冠状动脉疾病的诊断性能总体相似。
- 无陈旧性心肌梗死的多支血管病变患者中,3T血管水平敏感度、特异度及AUC均高于1.5T。
- 显著冠状动脉疾病定义为左主干直径狭窄≥50%或其他心外膜动脉直径狭窄≥70%。
核心结论
- 3T心血管磁共振检测显著冠状动脉疾病的诊断性能与1.5T相似,但在无陈旧性心肌梗死的多支血管病变患者中诊断性能更高。
研究设计与样本
- 前瞻性研究,纳入281例疑似或已知冠状动脉疾病患者(年龄62.4±8.3岁,193例男性),其中1.5T组135例、3T组146例,以侵入性冠状动脉造影为参考标准,由两名独立放射科医师判读灌注缺损。
诊断标准与关键数值
- 显著冠状动脉疾病定义为左主干直径狭窄≥50%及其他心外膜动脉直径狭窄≥70%;1.5T患者水平敏感度84%、特异度64%、阳性预测值74%、阴性预测值76%、AUC 0.75,血管水平分别为68%、83%、66%、84%、AUC 0.76;3T患者水平分别为80%、71%、71%、80%、AUC 0.76,血管水平分别为75%、86%、64%、91%、AUC 0.81;无陈旧性心肌梗死的多支血管病变患者血管水平,3T敏感度71% vs 1.5T 36%,特异度92% vs 69%,AUC 0.82 vs 0.53。
临床应用与阅片要点
- 在无陈旧性心肌梗死的多支血管病变患者中,3T血管水平诊断性能高于1.5T(敏感度71% vs 36%,特异度92% vs 69%,AUC 0.82 vs 0.53)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: To compare the diagnostic performance of cardiovascular magnetic resonance (CMR) myocardial perfusion at 1.5- and 3-tesla (T) for detecting significant coronary artery disease (CAD), with invasive coronary angiography (ICA) as the reference method.
MATERIALS AND METHODS: We prospectively enrolled 281 patients (age 62.4 ± 8.3 years, 193 men) with suspected or known CAD who had undergone 1.5T or 3T CMR and ICA. Two independent radiologists interpreted perfusion defects. With ICA as the reference standard, the diagnostic performance of 1.5T and 3T CMR for identifying significant CAD (≥ 50% diameter reduction of the left main and ≥ 70% diameter reduction of other epicardial arteries) was determined.
RESULTS: No differences were observed in baseline characteristics or prevalence of CAD and old myocardial infarction (MI) using 1.5T (n = 135) or 3T (n = 146) systems. Sensitivity, specificity, positive and negative predictive values, and area under the receiver operating characteristic curve (AUC) for detecting significant CAD were similar between the 1.5T (84%, 64%, 74%, 76%, and 0.75 per patient and 68%, 83%, 66%, 84%, and 0.76 per vessel) and 3T (80%, 71%, 71%, 80%, and 0.76 per patient and 75%, 86%, 64%, 91%, and 0.81 per vessel) systems. In patients with multi-vessel CAD without old MI, the sensitivity, specificity, and AUC with 3T were greater than those with 1.5T on a per-vessel basis (71% vs. 36%, 92% vs. 69%, and 0.82 vs. 0.53, respectively).
CONCLUSION: 3T CMR has similar diagnostic performance to 1.5T CMR in detecting significant CAD, except for higher diagnostic performance in patients with multi-vessel CAD without old MI.
引用
- Min JY, Ko SM, Song IY, et al. Comparison of the Diagnostic Accuracies of 1.5T and 3T Stress Myocardial Perfusion Cardiovascular Magnetic Resonance for Detecting Significant Coronary Artery Disease. Korean J Radiol. 2018. PMID: 30386133. doi:10.3348/kjr.2018.19.6.1007
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。