文献速递心血管技术与安全CT

第一至第三代双源CT扫描中无诊断相关伪影的无导线心脏起搏器检查

Leadless Cardiac Pacemaker (LCP) without Diagnostic Relevant Artifacts in DualSource and DualEnergy-CT Examinations in First- to Third-Generation DSCT Scanner.
PubMed 文献 · Acad Radiol · 2018-11-07

研究在200例第一至第三代双源CT(DSCT)扫描中评估无导线心脏起搏器(LCP)的伪影负荷,采用双能量(DECT)与DSCT技术并改变电流-时间乘积,发现总体平均Likert值为1.93±0.61,无图像超过3分;第二代为1.68±0.47,显著优于其他两代;DSCT图像质量显著优于DECT;起搏器长轴朝向扫描仪z轴可显著降低伪影。

要点

  • LCP在第一至第三代DSCT现有协议下均可获得诊断可接受的图像,总体平均Likert值1.93±0.61,无图像超过3分。
  • 起搏器角度越小伪影越轻:0°为1.4±0.5,90°为2.55±0.51,朝向扫描仪z轴可显著降低伪影。
  • 第二代DSCT伪影负荷1.68±0.47,显著优于第一代和第三代;DSCT图像质量显著优于DECT。

核心结论

  • LCP可在第一至第三代DSCT扫描仪上使用所评估且目前已在用的协议和技术安全检查,伪影负荷可通过将LCP长轴对齐或接近扫描仪z轴而显著降低。

研究设计与样本

  • 研究共分析200例检查,LCP在第一至第三代DSCT扫描仪上分别采用双能量(DECT)和DSCT技术并改变电流-时间乘积进行成像。
  • DECT检查在专用工作站上手动计算虚拟单能图像,虚拟电压由主观评估最低伪影负荷后手动选择。
  • 系统改变起搏器与机架的角度,从0°至90°以10°递增;伪影负荷由两名委员会认证放射科医师共识采用五分Likert量表评分(1无伪影、2少量、3中等、4较多、5大量),1–3分视为可诊断。

关键影像表现/征象

  • 总体平均Likert值为1.93±0.61,无图像被评估为Likert值大于3。
  • 角度评估显示伪影负荷随角度降低而明显且显著减轻:0°为1.4±0.5,90°为2.55±0.51。
  • 扫描仪层面,第二代DSCT为1.68±0.47,显著优于其他两台扫描仪。
  • 技术比较显示DSCT检查的图像质量显著优于DECT检查。

诊断标准与关键数值

  • 伪影负荷采用五分Likert量表评分:1无伪影、2少量伪影、3中等伪影、4较多伪影、5大量伪影;Likert值1–3视为可诊断。

临床应用与阅片要点

  • 使用所评估且目前已在用的协议和技术,LCP可在第一至第三代DSCT扫描仪上安全检查。
  • 将LCP长轴对齐或接近扫描仪z轴可显著降低伪影负荷。
  • 第二代DSCT的伪影负荷显著低于其他两代,DSCT图像质量显著优于DECT。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

RATIONALE AND OBJECTIVES: To identify the influence and artifact burden in cardiac CT imaging of a leadless cardiac pacemaker (LCP) performed with all three generations of DualSource CT (DSCT) Scanners.

MATERIALS AND METHODS: The LCP was examined in DSCT scanners of the first to third generation using DualEnergy (DECT) and DSCT as well as alterations of the current-time product. For DECT examinations, virtual monoenergetic images were computed manually on a dedicated workstation. Virtual voltage was manually selected by subjective assessment of the lowest artifact burden. Systematic variations of the pacemaker angle to the gantry were assessed, too. The angle was successively increased by 10°, ranging from 0° to 90°. Artifact burden was quantified on a five-point Likert scale (1- no artifacts, 2- few artifacts, 3- moderate artifacts, 4- many artifacts, and 5- massive artifacts). Likert values of 1-3 were considered diagnostic and assessed by two board-certified radiologists in consensus.

RESULTS: In total, 200 examinations were analyzed, a mean Likert value of 1.93 ± 0.61 was found overall. None of the images were assessed Likert value >3. The positioning evaluation showed a clear and significant reduction of artifact burden toward lower angles, (0°: 1.4 ± 0.5 vs. 90° 2.55 ± 0.51). At scanner level, second-generation DSCT performed significantly better (1.68 ± 0.47) than both other scanners. Comparison of technique (DECT vs. DSCT) revealed a significantly improved image quality in DSCT examinations.

CONCLUSION: LCP can be safely examined in DSCT scanner of the first to third generation with the evaluated protocols and techniques, which are currently in use. Artifact burden can be significantly reduced by aligning or approaching the LCP's longitudinal axis toward the scanner's z-axis.

来源:Acad Radiol,查看原文。版权归原作者所有。

引用

  1. Reinartz SD, Gohmann RF, Hardt F, et al. Leadless Cardiac Pacemaker (LCP) without Diagnostic Relevant Artifacts in DualSource and DualEnergy-CT Examinations in First- to Third-Generation DSCT Scanner. Acad Radiol. 2018. PMID: 30413293. doi:10.1016/j.acra.2018.10.003

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。